Abstract
Slipped Capital Femoral Epiphysis (SCFE) is a common pathologic hip condition in adolescents. In the standard treatment, a surgeon relies on multiple intra-operative fluoroscopic X-ray images to plan the screw placement and to guide a drill along the intended trajectory. More complex cases could require more images, and thereby, higher radiation dose to both patient and surgeon. We introduce a novel technique using an Inertial Measurement Unit (IMU) for recovering and visualizing the orthopedic tool trajectory in two orthogonal Xray images in real-time. The proposed technique improves screw placement accuracy and reduces the number of required fluoroscopic X-ray images without changing the current workflow. We present results from a phantom study using 20 bones to perform drilling and screw placement tasks. While dramatically reducing the number of required fluoroscopic images from 20 to 4, the results also show improvement in accuracy compared to the manual SCFE approach.
Chapter PDF
Similar content being viewed by others
Keywords
References
Sugano, N.: Computer-assisted orthopedic surgery. Journal of Orthopaedic Science 8(3), 442–448 (2003)
Musahl, V., Plakseychuk, A., Fu, F.H.: Current opinion on computer-aided surgical navigation and robotics: role in the treatment of sports-related injuries. Sports Med. 32(13), 809–818 (2002)
Rohilla, R., Singh, R., Magu, N., Devgan, A., Siwach, R., Sangwan, S.: Simultaneous use of cannulated reamer and schanz screw for closed intramedullary femoral nailing. ISRN Surg. (2011) (published online)
Liao, H., Ishihara, H., Tran, H.H., Masamune, K., Sakuma, I., Dohi, T.: Fusion of Laser Guidance and 3-D Autostereoscopic Image Overlay for Precision-Guided Surgery. In: Dohi, T., Sakuma, I., Liao, H. (eds.) MIAR 2008. LNCS, vol. 5128, pp. 367–376. Springer, Heidelberg (2008)
Marmurek, J., Wedlake, C., Pardasani, U., Eagleson, R., Peters, T.: Image-Guided Laser Projection for Port Placement in Minimally Invasive Surgery. Stud. Health Technol. Inform. 119, 367–372 (2006)
Fuchs, H., State, A., Pisano, E.D., Garrett, W.F., Hirota, G., Livingston, M., Whitton, M.C., Pizer, S.: Towards Performing Ultrasound-Guided Needle Biopsies from within a Head-Mounted Display. In: Höhne, K.H., Kikinis, R. (eds.) VBC 1996. LNCS, vol. 1131, pp. 591–600. Springer, Heidelberg (1996)
Walti, J., Jost, G.F., Cattin, P.C.: A New Cost-Effective Approach to Pedicular Screw Placement. In: Linte, C.A. (ed.) AE-CAI 2014. LNCS, vol. 8678, pp. 90–97. Springer, Heidelberg (2014)
Volonte, F., Pugin, F., Bucher, P., Sugimoto, M., Ratib, O., Morel, P.: Augmented Reality and Image Overlary Navigation with OsiriX in Laparoscopic and Robotic Surgery: Not Only a Matter of Fashion. J. Hepatobiliary Pancreat Sci. 18, 506–509 (2011)
Diotte, B., Fallavollita, P., Wang, L., Weidert, S., Thaller, P.-H., Euler, E., Navab, N.: Radiation-free drill guidance in interlocking of intramedullary nails. In: Ayache, N., Delingette, H., Golland, P., Mori, K. (eds.) MICCAI 2012, Part I. LNCS, vol. 7510, pp. 18–25. Springer, Heidelberg (2012)
Hoffmann, M., et al.: Next generation distal locking for intramedullary nails using an electromagnetic X-ray-radiation-free real-time navigation system. Journal of Trauma and Acute Care Surgery 73, Jg., Nr. 1, 243–248 (2012)
Stathopoulos, I., et al.: Radiation-free distal locking of intramedullary nails: Evaluation of a new electromagnetic computer-assisted guidance system. Injury, 44. Jg., Nr. 6, 872–875 (2013)
Arlettaz, Y., et al.: Distal locking of femoral nails: evaluation of a new radiation independent targeting system. Journal of Orthopaedic Trauma 26. Jg., Nr. 11, S. 633–637 (2012)
Madgwick, S.O.H., Harrison, A.J.L., Vaidyanathan, R.: Estimation of IMU and MARG orientation using a gradient descent algorithm. In: 2011 IEEE International Conference Rehabilitation Robotics (ICORR), pp. 1–7 (2011), doi:10.1109/ICORR.2011.5975346
Pring, E.M., Adamczyk, M., Hosalkar, H.S., Bastrom, T.P., Wallace, C.D., Newton, P.O.: In situ screw fixation of slipped capital femoral epiphysis with a novel approach: a double-cohort controlled study. Journal of Children’s Orthopaedics 4(3), 239–244 (2010)
Azizi Koutenaei, B., Guler, O., Wilson, E., et al.: Improved Screw Placement for Slipped Capital Femoral Epiphysis (SCFE) using Robotically-Assisted Drill Guidance. In: International Conference on Medical Image Computing and Computer-Assisted Intervention, vol. 17(01), pp. 488–495 (2014)
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2015 Springer International Publishing Switzerland
About this paper
Cite this paper
Koutenaei, B.A. et al. (2015). Inertial Measurement Unit for Radiation-Free Navigated Screw Placement in Slipped Capital Femoral Epiphysis Surgery. In: Navab, N., Hornegger, J., Wells, W., Frangi, A. (eds) Medical Image Computing and Computer-Assisted Intervention -- MICCAI 2015. MICCAI 2015. Lecture Notes in Computer Science(), vol 9349. Springer, Cham. https://doi.org/10.1007/978-3-319-24553-9_44
Download citation
DOI: https://doi.org/10.1007/978-3-319-24553-9_44
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-24552-2
Online ISBN: 978-3-319-24553-9
eBook Packages: Computer ScienceComputer Science (R0)