Introduction

This review is intended to provide an educational tool for physicians who want to recommend an insulin pump to their patients and need to learn about the current technologies available to them. The review focuses on the OmniPod® Insulin Management System (Insulet Corp., Bedford, MA, USA), comparing it to traditional insulin pumps and explaining its value within the range of products.

Traditional Insulin Pump Therapy

Insulin pumps were first developed for commercial use in the late 1970s. Over the past 30 years, the components have changed little. The components of traditional pump systems used for diabetes management include:

  • A device about the size of a deck of cards that contains an insulin reservoir, a pumping mechanism, batteries, and a microchip that carries the user’s daily insulin delivery programs and delivery history. The device has a screen and buttons to aid the user in delivering the proper amount of insulin. Current pumps come with color screens, sleek outlines, and much-improved user interface. The patient usually wears the device in a pocket or hooked on a belt or waistband.

  • An infusion set that attaches to the patient’s skin with adhesive and includes a short plastic cannula that is inserted just beneath the skin to allow access for the insulin. The patient either uses a mechanical inserter to attach the set and insert the cannula, or inserts the infusion set by hand.

  • A length of tubing that delivers the insulin from the pump to the infusion set and cannula. Tubing ranges from 30 to 110 cm (11.8 to 43.3 inches).

  • A separate blood glucose (BG) monitor. The patient uses the result on the monitor to calculate the dosage of insulin s/he should take and then inputs that instruction to the pump. Most pumps also include a bolus calculator to assist with these calculations. (In recent years, more pumps have added linked glucose meters that convey the BG value to the pump for correction bolus calculations.)

Table 1 lists insulin pump products currently available, alphabetically by manufacturer.

Table 1 Insulin pump (continuous subcutaneous insulin infusion) products currently available.

Benefits of Insulin Pump Therapy

Patients using multiple daily injections (MDI) typically inject four times a day or more, usually using two different types of insulin. The appeal of insulin pumps has been that they eliminate the need for these injections while allowing for greater customization and tailoring of insulin delivery. Insurers recognize the benefits to certain patients; for example, to get reimbursement for insulin pump therapy in the EU, the patient must have one or more of these indications:

  • Dawn phenomenon

  • Severe hypoglycemia unawareness

  • An irregular daily schedule, such as people who work various shifts or have unforeseen physical activities requiring a reduction of basal insulin

  • Pregnancy

  • Diabetes complications that require very tight glucose control

Continuous subcutaneous insulin infusion (CSII) therapy is not limited to those populations. It can benefit any patient with type 1 diabetes. Use of an insulin pump frees the patient from the strict eating schedule required when using long- and intermediate-acting insulins, and allows increased flexibility in daily activities. While the basal (or background) dosage is preprogrammed, it can be changed as needed. Meal and correction boluses are given as needed during the course of the day. So, for example, if the patient has no appetite for breakfast, s/he can replace it with a small snack or nothing, checking periodically that the BG remains steady. Further, if the patient wants to exercise at any given time, s/he can reduce or halt the basal delivery for a specified time, add food intake, reduce the bolus prior to exercise, or all three, as needed—all while checking the BG periodically.

Limitations of Traditional Insulin Pump Therapy

While pumping is freeing in some respects, it is also an added responsibility. Because pumps use rapid-acting insulin analogs, any interruption in insulin delivery may result in hyperglycemia and the rapid onset of diabetic ketoacidosis. In addition, traditionally designed systems have a number of limitations that could be improved upon to allow more patients with diabetes to benefit from CSII therapy. These include:

  • Long tubing connects the insulin reservoir in a traditional pump to the infusion set. The tubing has to be managed by patients in all aspects of daily living (dressing, exercising, etc). This tubing can catch on things during daily use, resulting in disruption of insulin delivery.

  • The process of inserting the cannula with mechanical inserters or by hand is awkward and can result in variable cannula insertion angles and depths. This mechanical or manual cannula insertion process can be intimidating to some patients.

  • Managing the large number of components required during setup and infusion set changes is complex.

  • Locations to insert infusion sets are relatively limited.

These limitations left room for innovation and improvement, which the designers of the OmniPod Insulin Management System implemented in their design and development, described in detail in the next section.

The Omnipod System: What is it and How Does it Work?

Insulin pumps are not all the same. Physicians and their patients need to choose a pump to fit the individual’s needs and lifestyle. The OmniPod System is quite innovative in resolving the limitations of traditional pumps listed above.

The OmniPod Insulin Management System is a novel insulin pump system designed by Insulet Corporation. The company was founded in 2000. In mid-2003, Insulet received its first 501(k) clearance from the Food and Drug Administration (FDA) for the OmniPod System. In early 2005, the commercial version of the OmniPod System was cleared by the FDA, and the product was shipped to the first patient later that year. By the end of 2008, the System was available and supported in all 50 US states. As of July 2010, the System is available in the United Kingdom and Germany under the brand mylife™ OmniPod®, and will soon be launched in other EU countries along with Australia and China.

The OmniPod System has won three awards for innovation in both industrial design and medical technology design (see Table 2). These honors demonstrate that those in the larger medical area (but not directly concerned with diabetes) are also convinced that this device is a real improvement in therapy.

Table 2 Design awards won by the OmniPod Insulin Management System

System Components

The Insulet OmniPod Insulin Management System is comprised of two components (Figure 1): the OmniPod Disposable Infusion Pump (Pod) and the OmniPod Personal Diabetes Manager (PDM), which incorporates the FreeStyle® (Abbott Diabetes Care, Alameda, CA, USA) BG meter.

Figure 1
figure 1

OmniPod Insulin Management System

The PDM is a hand-held, battery-powered remote controller for the Pod (insulin pump). It has seven functional buttons, a back-lit liquid-crystal colored display, and a BG meter. The PDM provides audio alarms, alerts, and reminders related to insulin delivery, reservoir level, Pod functioning, and battery life.

The Pod is activated and controlled exclusively through use of the PDM. The Pod and PDM interact wirelessly using secure, bidirectional radio frequency. Wireless interaction is only required to set a bolus delivery, change system settings, perform BG measurements, or change the Pod; in operation mode and during basal delivery, no wireless connection is needed, so the PDM does not need to be near the Pod. The Pod sounds an alarm if any attention is needed.

The Pod is a microprocessor-controlled device worn directly on the body. It delivers insulin based on the user’s custom-programmed basal rates and bolus doses for up to 3 days according to patient’s individual insulin needs.

Contraindications

Similar to traditional insulin pumps, the OmniPod is not recommended for people who are:

  • Unable to perform at least four BG tests per day

  • Unable to maintain contact with their healthcare provider

  • Unable to use the system according to instructions.

The integrated BG meter is not to be used for:

  • Testing newborns

  • Testing arterial blood

  • Diagnosing diabetes

Detailed Product Description

The Pod

The Pod is a single-use disposable device. It is small, lightweight, and worn directly on the body. The adhesive on the base of the Pod keeps it securely in place for up to 3 days. The Pod incorporates a piezo ceramic resonator to provide audio alarms, alerts, and reminders.

Insulin is delivered through an integrated soft cannula in the Pod that is inserted automatically into the subcutaneous tissue. The cannula is inserted in five 1000ths of a second. The Pod is intended for use with rapid-acting U-100 insulin: NovoLog® (Novo Nordisk A/S, Copenhagen, Denmark), NovoRapid® (Novo Nordisk A/S), Humalog® (Eli Lilly, Indianapolis, IA, USA), or Apidra® (Sanofi-Aventis, Paris, France).

The pad on the base of the Pod uses a standard, commercially available, medical-grade adhesive, designed for one-time use. Once removed, a Pod cannot be reapplied.

The Pod is watertight at a depth of 25 feet for up to 60 minutes (IPX8). Its housing includes a viewing window at the infusion site, which allows the patient to inspect the insertion site for proper cannula placement and integrity. The Pod also includes a vent to ensure pressure equilibration with the surrounding environment. The housing does not have any elements that allow user access to the pump mechanism or electronics and it contains no user-serviceable parts.

Pod Motor Mechanism

The motor mechanism of the Pod incorporates microprocessor control, shape memory alloy (SMA) wire assembly (Figure 2), step sensor, and a motion sensor. The SMA wire assembly rotates the driveshaft/leadscrew, advances the piston within the reservoir, and delivers insulin through the cannula. Each actuation of the motor mechanism delivers 0.05 units of insulin.

Figure 2
figure 2

Schematic of the OmniPod motor mechanism, featuring shape memory alloy (SMA) wire

Pod Insulin Reservoir

The Pod incorporates a fully integrated insulin reservoir that holds up to 200 deliverable units of U-100 insulin (Figure 3). It requires no separate administration sets or reservoirs.

Figure 3
figure 3

A look under the hood, featuring core components of the OnmiPod

Fill Process

Each Pod is supplied with a fill syringe and a fill needle that are used to introduce insulin into the Pod. The fill syringe is sized to match the maximum fluid volume of the Pod with a self-stopping plunger rod that limits the amount of insulin it can contain. The syringe is clearly marked with the minimum fill volume (85 units) of the Pod and in units of insulin.

After the patient fills the syringe with insulin, s/he injects it into the Pod reservoir through the fill port on the bottom of the Pod. The self-sealing fill septum is designed for single use only. It is located on the bottom (skin contacting) surface of the Pod to eliminate the possibility of injecting insulin into the port for a second use. Typical pod wear times vary from 48 to 72 hours. The pod has the capability of lasting up to 80 hours to allow for flexibility in changing the pod due to unforeseen circumstances.

Cannula Insertion Process

When the patient is ready to attach the Pod, s/he removes the adhesive and needle cover from the back and sticks the Pod onto the prepared site. By following the PDM’s screens and pressing specific buttons, the patient tells the Pod to activate its insertion mechanism, which quickly drives both the insertion needle and the soft cannula into the subcutaneous tissue. The insertion needle punctures the skin and stiffens the cannula during the insertion process. When it reaches its forward limit of travel, the insertion mechanism retracts and withdraws only the needle, leaving the cannula in place at the infusion site. The actual insertion takes only five 1000ths of a second. The Pod places the soft cannula into the patient’s tissue at a 45-degree angle, to a depth of 6.5 mm (Figure 4). The adhesive pad on the base holds the Pod and the cannula in place.

Figure 4
figure 4

Catheter view demonstrating 45-degree angle and 6.5 mm vertical depth

Insulin Delivery

The PDM handles all processes for the operation of the system. The display uses full text to prompt the user through the set-up and use processes. The patient must use the PDM to program the Pod with the his/her custom basal profile, check the Pod status, initiate a bolus, or make changes to the insulin delivery profile. The PDM needs to be within 60 cm to communicate with the Pod to perform these tasks. After set-up and initiation, the Pod runs at the programmed basal rate independently of the PDM. If the PDM requires the patient’s attention, the Pod sounds an alarm beeper.

Incorporated BG Measurement

The PDM incorporates a FreeStyle (Abbott Diabetes Care) BG monitor. The BG data is stored and its history can be displayed on the PDM. For example, a patient’s BG variations can be seen in both graphical and list form for the past 1, 7, 14, 30, 60, and 90 days. The list form shows, for each time period, the average number of readings taken, the average BG, the minimum and maximum readings, the patient’s BG goal range, and the percentage of readings within, above, and below that goal. Along with insulin delivery records, this information can be quite valuable to the physician or nurse when reviewing the patient’s status.

Suggested Bolus Calculation

The PDM provides the ability to calculate a suggested bolus dosage, which has traditionally been done manually by MDI users. Similar convenience features have since been incorporated into traditional CSII pumps. The suggested bolus calculation feature is provided as a convenience to the user to aid in determining the bolus dosage needed based on carbohydrates ingested, most recent BG reading, user-settable correction factor, insulin-to-carbohydrate ratio, target BG value, and duration of insulin action.

Does the Omnipod System Promote Better Type 1 Diabetes Management?

CSII delivery, in general, allows for greater flexibility in insulin delivery options leading to more lifestyle choices related to meal planning and exercise. More specifically, I believe the use of the OmniPod System contributes to improved diabetes management for a variety of reasons, including ease of training, setup and insertion, uninterrupted and untethered insulin delivery, and inclusion of a glucose meter in the PDM. An additional advantage of the System may be that the Pod, once on the body, does not move around much. This results in less fluctuation in hydrostatic pressures, ideally leading to more consistent insulin delivery. In the near future, Insulet Corporation will release in the US a modified PDM that receives continuous glucose sensing data from the iDex platform (DexCom, Inc., San Diego, CA, USA).

In 2005, Dr. Lois Jovanoviç and I performed the first clinical evaluation of the OmniPod System in 20 subjects with type 1 diabetes.1 (See also Table 3 for brief summaries of studies.110) This 30-day study, supported by a grant from Insulet, assessed the comfort, function, and use of the System compared with conventional insulin pumps. The trial was designed to see how the subjects liked the new pump compared to their current, traditional CSII pump. We found that the subjects much preferred wearing the OmniPod System. Ninety percent of subjects (18 of 20) preferred using the OmniPod automated cannula insertion system versus inserting with their current infusion sets. What we didn’t expect from a month-long pilot study was that the vast majority of the subjects had A1C values at the end that were significantly lower, dropping from a mean of 7.1% (range 5.5–8.1) to a mean of 6.8% (range 5.4–7.6; P<0.002). This significant change was even more surprising because I had not changed the subjects’ pump settings when starting them on the OmniPod System.

Table 3 Results of clinical trials and reviews involving the OmniPod Insulin Management System

We expected some “study effect” on the subjects’ A1C values but we did not expect that large a decrease in such a short time. What could be responsible for this improved control despite staying on the same pump settings as their traditional pump? In a later study, I discovered the likely answer to that question.2 One obvious difference was that the subjects wearing the OmniPod System never disconnected from their pump for any reason other than putting on a new Pod when the insulin ran out. CSII therapy allows for uninterrupted, around-the-clock delivery of insulin, but in practice, many patients disconnect their insulin pumps from their infusion sets when showering, changing clothes, swimming, exercising, or during intimate moments. Ranging in time from a few minutes to several hours, these periods of interrupted insulin infusion result in the cessation of basal insulin delivery. Interrupting insulin delivery may be appropriate during periods of exercise when insulin sensitivity increases or during hypoglycemia; however, cessation of insulin delivery may be detrimental even for short periods, especially if it occurs when insulin resistance is at its peak. A prospective, open-label study was designed to measure the impact of short-term infusion-set disconnects and infusion-set changes on glucose levels.

Nineteen subjects with type 1 diabetes were studied on two separate in-clinic days.2 One hour after arriving at the clinic in a fasting state, subjects either temporarily disconnected their infusion sets from their pumps, interrupting basal insulin infusion for 30 minutes, or, on a separate day, changed their infusion sets. Glucose levels were monitored for an additional 4 hours on both occasions. Changing infusion sets did not affect short-term glucose control. However, the 30-minute interruption of basal insulin infusion resulted in significant glucose elevation: 1 mg/dL for each minute basal insulin infusion was interrupted.

A second, less obvious difference between a traditional CSII pump and the OmniPod System is that a traditional pump is tethered to the body and therefore moves around in relation to the insulin infusion site. Could this movement result in changes in insulin delivery? A bench-top study was designed to quantify these hydrostatic effects on CSII pumps during basal and bolus insulin delivery.3 We tested CSII pumps from Medtronic Diabetes (MiniMed 512 and 515), Smiths Medical (Deltec Cozmo 1700), and Insulet (OmniPod) using insulin aspart (NovoLog, Novo Nordisk).The pumps were filled and primed per manufacturer’s instructions. The fluid level change was measured using an inline graduated glass pipette (100 microliters) when the pipette was moved in relation to the pump (80 cm Cosmo and 110 cm Medtronics) and when level. Pumps were compared during 1 and 5 U boluses and basal insulin delivery of 1.0 and 1.5 U/hour. We found pronounced differences that were seen during basal delivery in pumps using 80–100 cm tubing. For the 1 U/hour rate, differences ranged from 74.5% of the expected delivery when the pumps were below the pipettes and pumping upward, to 123.3% when the pumps were above the pipettes and pumping downward. For the 1.5 U/hour rate, differences ranged from 86.7% to 117.0% when the pumps were below or above the pipettes, respectively. Compared to pumps with tubing, the OmniPod System performed with significantly less variation in insulin delivery. Changing position of a conventional CSII pump in relation to its tubing results in significant changes in insulin delivery. The siphon effect in the tubing may affect the accuracy of insulin delivery, especially during low basal rates. This effect has been reported when syringe pumps were moved in relation to infusion sites but had not been reported with CSII pumps. Please note that this was a bench-top laboratory evaluation without clinical validation. (While the specific MiniMed and Deltec pumps used in this study are no longer being sold, presumably the same siphon effect applies with newer pumps that use similar tubing.)

Additional clinical differences may exist than those given above. More research may be needed to quantify those clinical differences between traditionally designed insulin pumps and this new category of pump.

Review of Clinical Data

Clinical Trials and Reviews

Clinical trials like those referred to above have demonstrated the effectiveness of Insulet’s OmniPod Insulin Management System in managing diabetes. Table 3 lists these and a number of other trials and published research papers related to the OmniPod System.110

Ongoing Research and Development

The OmniPod System is currently in use as part of the Juvenile Diabetes Research Foundation’s (JDRF) Artificial Pancreas Project (APP). This is a global initiative to develop a first-generation artificial-pancreas therapy model. It integrates a continuous BG monitor and an insulin pump, with a customized algorithm, to deliver the appropriate dose of insulin. Starting approximately 3 years ago, the OmniPod System continues to be used at the majority of the JDRF-funded clinical research sites worldwide. The OmniPod System is also scheduled to be used at a number of EU sites as part of the AP@home Artificial Pancreas Project.11

Comparison With Other Insulin Pumps

Figure 5 provides a comparison of the Insulet OmniPod System to four traditionally designed insulin pumps. Notice especially the list of additional features at the end of the table. These include the OmniPod System’s tubing-free design, innovative safety features, integrated bolus calculator, and many personalized settings.

Figure 5
figure 5

Insulin pump therapy technology reference guide.1

Figure 5
figure 6

Insulin pump therapy technology reference guide.1

The OmniPod System may have limitations for some users:

  • Occasional Pod failure does occur. This may be due to inserting the Pod in a site with scar tissue, system failure, occlusion, or some other cause.

  • In extreme heat, the Pod’s adhesive may not hold securely. As with traditional infusion sets, the patient may have to add extra adhesive.

  • Some patients are especially self-conscious and may not want a “bump” on their bodies. They may feel they can hide a conventional pump more easily, even with tubing and the bulk of the pump itself.

  • Patients who require more than 100 units of insulin per day may do better choosing a pump with a larger, 300-unit reservoir. Some may be able to change the Pod every 2 days instead of every 3 days.

Patient Concerns

Patients with diabetes will have concerns about using a new device. Table 4 shows the answers to some of the most common questions patients have. A broader range of questions and answers is available online.12

Table 4 Common patient questions about the OmniPod Insulin Management System

Conclusion

For many patients with diabetes, CSII is the only way to reach acceptable glucose control levels. The OmniPod Insulin Management System improves on the traditional method of delivering insulin via CSII for the following reasons:

  • It frees patients with diabetes of being tethered to their pumps.

  • It provides consistent, round-the-clock, uninterrupted insulin delivery.

  • It allows for easier set up and automated cannula insertion.

  • Greater convenience often leads to better compliance.

The OmniPod System offers the comfort and safety of feedback and alarms, the ability to follow daily, monthly, and quarterly progress, and the feeling of accomplishment in managing one’s own diabetes. Additionally, patients do not need to be technologically “savvy,” because the device’s user interface is very clear and straightforward. Almost all patients with type 1 diabetes can benefit from using the OmniPod System.

Several next-generation systems are rumored to be in development at Insulet Corporation. These systems include a PDM that receives continuous glucose monitor (CGM) data from the DEXCOM 7 Plus continuous glucose sensor, a Pod that is over a third smaller in size with the same technical features, a Pod that incorporates a CGM sensor, and a Pod that has the ability to deliver multiple compounds.

The OmniPod platform lends itself to many other uses. It is presently being used in Europe by Ferring Corporation as a delivery mechanism for an infertility drug. Other potential applications include: pain management; chemotherapeutic agents; AIDS, Parkinson’s, and multiple sclerosis drugs; biologics; thyroid, growth hormones, and related peptides; as well as in-hospital uses.

Acknowledgments

The author thanks Insulet Corporation for providing technical support for this article and for permission to use the photographs.

The author has the following disclosures: Consultant Advisor: Animas, Insulet, MannKind, and Roche. Research Support: DexCom, Eli Lilly, GluMetrics, Insulet, LifeScan, Medtronic, and Roche.

Cynthia M. Shaler provided editorial assistance in preparing this manuscript. Ms. Shaler is a technical and medical designer, editor, and writer.

Prior to peer review, the manufacturers of the OmniPod Insulin Management System, Insulet and Ypsomed, were offered the opportunity to review the article solely to ensure accuracy of the data. Minor changes were made to the content as a result.

Open Access. This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.