Introduction

One of the earliest and most common applications of clinical echocardiography is evaluation of left ventricular (LV) function and size. Accurate, reproducible and quantitative evaluation of LV function and size is vital for diagnosis, treatment and prediction of prognosis of heart disease. Early three-dimensional (3D) echocardiographic techniques showed better reproducibility than two-dimensional (2D) echocardiography and narrower limits of agreement for assessment of LV function and size in comparison with reference methods, mostly cardiac magnetic resonance (CMR) imaging, but acquisition methods were cumbersome and a lack of user-friendly analysis software initially precluded widespread use [13]. Through the advent of matrix transducers enabling real-time 3D echocardiography (3DE) and impressive improvements in analysis software featuring semi-automated volumetric analysis, 3DE evolved into a simple and fast imaging modality for everyday clinical use. 3DE provides the possibility to evaluate the entire LV in three spatial dimensions during the complete cardiac cycle, offering a more accurate and complete quantitative evaluation of the LV [47]. Improved efficiency in acquisition and analysis may provide clinicians with important diagnostic information within minutes.

The current article reviews the methodology and application of 3DE for quantitative evaluation of the LV, provides the scientific evidence for its current clinical use, and discusses its current limitations and potential future directions.

Methodology

Technology

3DE has been made possible in particular by the development of matrix transducers. A matrix transducer contains thousands of piezoelectric elements in a 2D array, which can be electronically steered in multiple directions very rapidly, enabling the acquisition of a real-time 3D volume dataset. Such a volume dataset has a pyramidal shape with a curved base and is approximately 30° × 60°, depending on which manufacturer’s hardware is used. This is more than adequate for visualisation of different structures within the heart such as valves and masses. Until recently, inclusion of the entire LV within a dataset necessitated an automated ECG-triggered capture of multiple consecutive real-time datasets (usually four to seven) during briefly held respiration and electronically stitching these datasets together. The latest generation of 3D scanners provide acquisition of a ‘full volume’ at a frame rate of 40–50 Hz. Either method results in a ‘full-volume’ dataset of up to 90° × 110°, which in general is sufficient to allow for accurate analysis of the LV. The use of more sub-volumes results in higher line density and higher resolution. However, it also increases the chance of stitching artifacts due to potentially erroneous ECG triggering causing unsynchronised sub-volumes. Until recently, this precluded 3D analysis in patients with irregular heart rhythms such as atrial fibrillation. Fortunately, one of the more recent developments in 3DE is the possibility of acquiring a full-volume dataset within a single heartbeat, obviating this limitation by providing instantaneous real-time volumetric imaging of the entire LV and decreasing acquisition time. A recent comparison between multiple consecutive real-time acquisitions, two-beat and single-beat acquisitions demonstrated a significantly lower frame rate in single-beat acquisitions with, as a consequence, underestimation of ejection fraction (EF) [8]. The two-beat modality provided similar accuracy in LV volume and EF measurements and may be preferred due to fewer stitching artifacts. In atrial fibrillation, however, single-beat acquisition may be superior because absence of stitching artifacts may be more important than image quality deterioration [9].

Acquisition

After positioning the transducer accurately to include the whole LV into the 3D volume, a single 3D acquisition generally takes less than 10 s. Although all standard acquisition windows for echocardiography are available and useful for 3D acquisitions, the ideal and generally preferred approach for acquisition of the LV is from the apical window. Depending on the shape of the heart and its position within the chest, a more off-axis position may be appropriate to ensure the acquisition of the entire LV. To guarantee optimal image quality, transducer frequency and overall gain should be adjusted accordingly. Typically, the gain setting should be somewhat higher than used for 2D echocardiography to get the most favourable result. In addition, the smallest possible dataset encompassing the whole LV should be acquired to ensure maximal temporal and spatial resolution. Finally, it is important to ascertain that the acquisition is made during breath hold to minimise the risk of breathing artifacts. In multi-beat acquisitions, stitching artifacts can easily be detected in the transversal plane.

Visualisation

Concurrently with the acquisition of a 3D dataset, it is volume rendered within the echocardiography system, resulting in a 3D volume dataset which can be rotated and viewed from any angle. In combination with the ability to crop part of the dataset away, it is possible to visualise the anatomy and function of the LV in great detail. Cropping a 3D volume can be done by utilising a ‘crop-box’ and by use of an ‘anyplane’ mode. With the first method the 3D dataset is displayed within a box, the sides of which can be selected and cropped away. The more conventional 2D long-axis and short-axis views can be inspected easily and rapidly this way but ultimately any long-axis or short-axis view can be chosen. The ‘anyplane’ mode enables cropping of the 3D volume from every conceivable orientation, providing unlimited sections of the LV. Although the interpretation of such unconventional cut-planes can initially be difficult, the learning curve is relatively steep due to the 3D nature of the acquired images.

Analysis

Most vendors offer software packages for both online and offline quantitative analysis of the LV. Most of these commercially available software packages provide the possibility to check for stitching artifacts in a transversal plane using the slicing (particularly multislice) function that can also be used to analyse wall motion and distribution of hypertrophy. Furthermore, certain anatomical landmarks such as the mitral annulus and apex are identified in multiple conventional 2D planes. Thereafter, the software semi-automatically detects and traces the endocardial border of the LV during the cardiac cycle, with a possibility to perform manual corrections as needed [10]. Subsequently the software creates a cast of the LV, allowing comprehensive volumetric analysis of LV function and morphology. Not only LV volumes and ejection fraction (EF) can automatically be calculated, but time-volume curves can be generated to evaluate segmental LV function and the presence of dyssynchrony. Figure 1 is a screen shot of the work flow of one of the commercially available software packages. Finally, the segmental contraction pattern can be displayed graphically through parametric imaging in a dynamic ‘bulls eye’ display (an example is shown in Fig. 2), providing information on segmental volume changes and timing of segmental contraction, which may be of particular interest in guiding optimal LV lead placement for cardiac resynchronisation therapy (CRT).

Fig. 1
figure 1

Example of the work flow of one of the commercially available software packages. Top, left and right: two perpendicular long-axis images of the LV. The endocardial contour is detected semi-automatically and is projected on the image. Centre left: a third perpendicular short-axis image, again with the endocardial contour. Note that this short-axis image was made with the ultrasound transducer in the apical position. Centre right: The LV ‘cast’ is shown with colour-coding of individual LV segments. Bottom: For each LV segment, a time-volume curve is generated. Regional wall motion and dyssynchrony can be appreciated easily

Fig. 2
figure 2

Parametric images derived from three-dimensional datasets. Images of a patient are shown with CRT turned off (left) and on (right), respectively. The LV is divided into 16 segments as described by the American Society of Echocardiography. Colour coding is used to represent regional time-to-minimum systolic volume, with red indicating late activation. In the left hand map, the lateral wall is identified as the latest activated LV region without CRT (left). As can be appreciated in the right hand map by the overall homogenous blue colour indicating the absence of large regions with delayed contraction, synchronicity is re-established when CRT is applied (right)

Feasibility

An adequate echocardiographic image quality is a prerequisite for meaningful 3DE and most authors use image quality as an inclusion criterion. Four articles provide information about image quality in an unselected population [9, 1113]. In a total of 270 patients, 162 had good or optimal image quality, 51 had average image quality, and 41 had poor image quality.

Furthermore we would like to comment that in our experience, semi-automated endocardial contour detection is influenced negatively by the presence of chordae tendiniae, papillary muscles and sub-valvular mitral calcification, even in patients with a good image quality necessitating doing some manual corrections of the automated endocardial contour in the majority of patients.

Recently, a new matrix transducer (X Matrix X5-1, Philips, Andover, MA) was introduced that has a smaller footprint and allows acquisition of both 2D images and 3D datasets without the need to change the probe providing a higher temporal and spatial resolution and improving the feasibility and quality of acquisitions. In addition to the need to improve image quality, wider angle acquisition is also indispensable for adequate visualisation of the entire heart in one volume in severely enlarged ventricles. Further improvement in border detection and speckle tracking algorithms is necessary to obviate the need for offline analysis. Moreover, further development of DICOM algorithms, which allow to display and analyse 3D datasets independently on vendor specific software, is also needed.

Clinical applications

LV volumes and function

In evaluation of LV function, 2D echocardiography has its own limitations, mainly that quantitative measurements such as Simpson’s biplane summation-of-disks method for LV volume measurement are based on assumptions of symmetrical LV geometry that are inherently unreliable. These assumptions as well as the unintentional use of foreshortened views account for the relative inaccuracy and poor reproducibility of these techniques. 3DE obviates both these limitations, and therefore measurements of LV volumes and EF are significantly more accurate and reproducible than conventional 2D echocardiography when compared with CMR as the reference technique (Figs. 3 and 4) [630]. Unsurprisingly, studies that only included patients with good echocardiographic image quality [6, 7, 21, 22, 30] showed better agreement than those who did not [13, 19, 23]. Furthermore, studies using semi-automated endocardial contour analysis [6, 13, 26, 28, 29] in 3DE generally showed better agreement than those using manual contour tracing [19, 23, 27]. An exception to the rule is an early study by Kühl et al. [4] published in 2004, but it is likely that semi-automatic contour detection software has substantially improved since then. Finally, two studies focusing on potential differences between semi-automated contour tracing algorithms found some of the differences to be statistically significant, but clinically irrelevant [22, 30].

Fig. 3
figure 3

LV volumes and EF measurements by 3DE: comparison with CMR. In total, we identified 21 studies including 1040 patients (747 males, 60 ± 9 years) in which LV volumes and ejection fraction as measured by 3DE were compared with CMR as the reference method [47, 1117, 1923, 2529]. Bland-Altman statistics are shown for EDV, ESV, and EF by 3DE in comparison with CMR as the reference method. Mean difference ±2 standard deviations are shown as a percentage of the mean value, as measured by 3DE. Studies are shown in chronological order, with the most recent publications to the right. Jenkins et al. [12]: a TomTec offline, b QLAB online; Soliman et al. [21]: a multiplane interpolation, b full-volume reconstruction; Soliman et al. [22]: a TomTec 4D LV-Analysis ver. 2.0, b QLAB V4.2; Mor-Avi et al. [25]: a CMR short-axis vs. 3DE long-axis image analysis, b CMR and 3DE long-axis image analysis, c trabeculae included in LV volume in CMR analysis, d trabeculae excluded from LV volume in CMR analysis; Chukwu et al. [29]: a normal volunteers, b myocardial infarction patients

Fig. 4
figure 4

LV volumes and function measurements by 3DE : a intra-observer variability; b inter- observer variability. Legend: To study the reproducibility of LV volumes and EF measurements, we identified 16 studies including 407 patients in which reproducibility of these measurements by 3DE was assessed [4, 5, 1317, 20, 2225, 28, 30]. Studies are shown in chronological order, with the most recent publications to the right. Kühl et al. [4]: a TomTec CardioView RT semi-automated analysis, b TomTec CardioView RT manual analysis; Soliman et al. [22]: a TomTec 4D LV-Analysis ver. 2.0, b Philips QLAB V4.2; Hansegård et al. [30]: a TomTec 4D LV-Analysis ver. 2.2, b GE 4DLVQ software, EchoPAC ver. 108.1.0

Despite the high correlation with CMR as a reference technique, most studies showed significant underestimation of 3DE-derived LV volumes, irrespective of the method of analysis [19, 23, 26]. The potential reasons for this systematic underestimation of LV volumes by 3DE as compared with CMR were identified in a recent multicentre study by Mor-Avi et al [25]. They concluded that 3DE-derived LV volumes are underestimated in most patients because, unlike CMR, 3DE imaging cannot differentiate between the myocardium and trabeculae. Therefore, in most patients the endocardial contour is traced along the innermost endocardial border by 3DE, i.e. excluding the trabeculae from the LV cavity. By convention, however, the endocardial contour is traced along the outermost endocardial border by CMR, i.e. including the trabeculae in the LV cavity. To minimise the difference between the two modalities, tracing the endocardium to include trabeculae in the LV cavity is recommended for 3DE. In addition to tracing the endocardium beyond the visible interface, sufficient experience [25] and good image quality [31] are the most important factors to achieve accurate and reproducible 3DE quantification of LV volumes. Finally, if image quality is poor, multiple studies have shown that echo contrast agents can be of incremental value to improve the accuracy and reproducibility of LV volume and function measurements [13, 23, 32].

The superiority of 3DE over 2DE in clinical decision-making was proven in a study comparing the clinical impact of 3DE versus 2DE in different clinical scenarios. This study concluded that measurement of LV volumes and EF by 3DE is clinically feasible and has the potential to significantly alter clinical decision-making [33]. Further studies and more robust outcome data on defining 3DE cut-offs for clinical decision are still needed to confirm such clinical superiority on definite and clinically meaningful endpoints to be part of the routine clinical use.

LV dyssynchrony

Cardiac resynchronisation therapy has been shown to improve LV function, survival and quality of life. However, accurate prediction of response to pacing remains difficult as approximately 30% of patients selected with current selection criteria do not respond to CRT [34]. A low correlation exists between LV mechanical dyssynchrony and QRS duration, and mechanical dyssynchrony might be a better predictor of response to CRT than QRS duration. At present, tissue Doppler imaging (TDI) is the mostly used echocardiographic technique for quantification of mechanical dyssynchrony [35, 36]. However, TDI can only provide information on the systolic myocardial motion of the basal and mid-segments, in the longitudinal direction. Because its reproducibility is relatively low and no TDI-based dyssynchrony parameter has proven to be useful for the prediction of response to CRT in a large clinical trial [37], the need for reliable alternative echocardiographic parameters of dyssynchrony has been recognised. 3DE offers information on systolic contraction based on inward endocardial motion, providing a complete evaluation of intraventricular mechanical dyssynchrony of all LV segments simultaneously. The site of maximal mechanical delay can be visualised in a polar map. In addition, quantitative dyssynchrony analysis is possible by measuring time needed for each individual segment to reach minimum systolic volume. The standard deviation of the dispersion in time from the onset of QRS complex to minimal regional volume of all segments can be used as a parameter of LV dyssynchrony.

There are conflicting reports on the correlation between LV dyssynchrony measured by TDI and 3DE methods, ranging from fair to poor agreement [3843]. In large part this has been attributed to the fact that 3DE and TDI provide distinctly different measures of ventricular motion. When both techniques were directly compared for prediction of volumetric and clinical response to CRT, results clearly showed that 3DE has better sensitivity and specificity to predict response to CRT than TDI [43]. Three other studies demonstrated that 3DE-derived SDI was highly predictive of both the acute and chronic response [24, 42, 44]. However, currently there is no agreement on cut-off values for the prediction of response to CRT and a larger multicentre study assessing reproducibility and predictive value for clinical endpoints is warranted. Reference values and reproducibility of LV dyssynchrony in normal volunteers and in various groups of patients are given in Tables 1 and 2. These values are highly variable and are based on relatively small studies in which varying software was used [24, 38, 4151].

Table 1 Reference values for 3DE-derived systolic dyssynchrony index (SDI)
Table 2 Intra- and inter-observer variability of SDI measurement by 3DE

Finally, the importance of optimal LV pacing lead position was emphasised in a 3DE study [52] comparing response to CRT, defined as increase in oxygen uptake and decrease in LV volume, in patients with the LV pacing lead at the segment with the maximum mechanical delay to patients with the LV pacing lead at other segments. Response was optimal in patients with optimal LV pacing lead position as assessed by 3DE, and decreased significantly with increasing distance between optimal pacing site and achieved pacing site. In this regard, 3DE may prove to be useful for the optimisation of LV lead placement [39, 52]. In contrast, a smaller study [53] with a similar design found that implantation of the LV lead in the most delayed LV segment determined by 3DE did not result in additional improvement in symptoms or LV function.

Future perspectives

Semi-automated endocardial contour detection provides information of systolic contraction based on 3D endocardial excursion only. In contrast, speckle tracking is an angle-independent technique that can quantitatively evaluate myocardial deformation or strain in multiple directions by tracking the movement of natural acoustic markers of the myocardium known as speckles which appear on grey-scale images. When applied to 2D echocardiography images, however, speckle tracking is significantly hampered by the intrinsic nature of its 2D methodology. Not only foreshortened views and geometric assumptions limit the applicability of 2D speckle tracking, but out-of-plane motion of speckles during the cardiac cycle also result in suboptimal tracking and considerable noise, reducing the accuracy and reproducibility of its quantitative evaluation of strain.

Recently, 3D speckle tracking has been introduced to overcome these limitations. With 3D speckle tracking, the complex LV wall motion can be assessed in all three spatial dimensions and there is no loss of information because the speckles do not move in and out of the volume. Measurements are based on regions of interest rather than (semi-automated) contour detection. 3D speckle tracking measurements of LV volumes shows higher correlation with CMR, smaller biases, and narrower limits of agreement compared with 2D speckle tracking [54]. LV twist has been reliably measured, but as yet in healthy volunteers only [55]. In LV mechanical dyssynchrony, semi-automatic assessment is reasonably reproducible and a clear difference was demonstrated between RV apical and biventricular pacing [56]. The ability to identify the site of latest mechanical activation may aid in optimisation of biventricular lead placement [57]. Additional studies will have to validate the use of this novel technology for different clinical purposes and establish its incremental value compared with volumetric endocardial contour tracing.

Conclusions

In the last two decades the technology behind 3DE has greatly evolved, continually expanding its clinical applications. Ample evidence suggests that LV volumes and function measurements by 3DE have closer limits of agreement with CMR measurements as a reference method and better reproducibility than 2D echocardiography. 3DE assessment of LV dyssynchrony shows great potential. Future advancements in hardware will facilitate the acquisition of wide-angle, single-cycle pyramidal datasets with higher spatial and temporal resolution. Automatic analysis of endocardial contours and speckle tracking will enable fast online measurements that are accurate and reproducible and will ensure its integration into the routine echocardiographic examination.