Introduction

For joint replacements, orthopedic implants are frequently used and expected to rise for the next 15–20 years.1 As reported previously, 18% of implants fail due to aseptic loosening while 20% fail because of infection.2 , 3 Hence, strong osseointegration and antibacterial properties both need to be taken into consideration while developing an implant material.4

Internal fixation devices are still being made of austenitic stainless steels (largely SS 316L) in spite of their lower corrosion resistance compared with titanium. This is a result of their excellent mechanical properties and low cost.5 , 6 Lately, a newer generation of steel UNS S31254 SS (254SS) has been studied. As reported in the literature, despite 254SS not being bioactive and lacking strong osseointegration, the absence of toxic effects and high nitrogen content of 254SS have made it a candidate for scientific investigation, in the hope of developing a futuristic orthopedic implant material.7 , 8 Nevertheless, the literature confirms that these problems can be mitigated by depositing hydroxyapatite (HAP) coatings on metallic implants.9,10,11,12,13,14,15,16, 17

Hydroxyapatite (HAP) holds a significant position as an inorganic biomaterial, but bulk HAP ceramics possess poor mechanical properties restricting its use for non-load-bearing implant applications. To overcome this limitation, HAP is being coated on metals and their alloys to derive the advantages of both the bioactivity of HAP and the mechanical performance of metals.18 HAP is a widely used coating material for orthopedic implants because its mechanical properties closely match with those of human bone mineral, and it is also biocompatible with human bone tissues.19 HAP coating of orthopedic implants can be an effective method of improving their physiological response leading to an overall improved performance of the implants.20

HAP coatings can be deposited on a metallic substrate using techniques such as direct laser melting, sol–gel, and vapor deposition processes, including plasma spraying, which is a commercially viable technique for clinical applications.20,21,22,23,24,25,26, 27 Nevertheless, HAP coatings produced by plasma-spray bond poorly with metal substrate and are a candidate to easy wear and tear, which results in undesirable debris formation.28 Synthesizing a composite layer with HAP on the implant surface is one way to solve these problems. This can be achieved by melting the preplaced HAP on the implant surface using a laser.24 The primary aim of this research is to assess the feasibility of synthesizing multifunctional orthopedic coatings on 254SS implant surfaces using laser rapid manufacturing (LRM), which can prevent infection while still promoting osseointegration. LRM is an additive manufacturing (AM) process, which can be used not only for preparing customized implants but also for treating their surfaces in several ways.24 , 29 , 30

Experimental Work

An LRM AM system equipped with a 2.0-kW continuous wave, Ytterbium-doped, fiber laser with a beam size of 2.5 mm was used. For direct laser melting of the preplaced HAP powder bed on the surface of a high nitrogen stainless steel (254SS) substrate (100 × 50 × 2 mm3 sheet), several coating samples, over a 10 × 10 mm2 area, were prepared at 1000-W laser power and 1-m/min scan speed. These process parameter settings were finalized based on Refs. 24 and 25, and preliminary trials were conducted. At 800-W and 900-W laser power with 0.8-m/min, 0.9-m/min, and 1.0-m/min scan speeds, discontinuous tracks with weak bonding were produced. At 1000-W power with 0.8-m/min and 0.9-m/min scan speeds, discontinuous tracks were produced while 1200-W power with 0.8-m/min, 0.9-m/min, and 1.0-m/min scan speeds resulted in red hotness. These outcomes are undesirable for orthopedic coatings, and hence, the laser power and scan speed were finally fixed at 1000 W and 1 m/min, respectively.

The HAP-coated 254SS samples were sectioned and metallographically prepared for microstructural studies according to ASTM E3-11. Microstructure and surface morphology studies were conducted using an optical microscope and a scanning electron microscope (SEM, ZEISS EVO 50) equipped with an energy-dispersive spectrometer (EDS), respectively. The phase composition of the coated samples was examined using an x-ray diffraction (XRD; RIGAKU Model: Smart lab 3 KW). The microhardness of pristine and coated 254 stainless steel was determined using a Vickers microhardness tester. The antibacterial efficacy of pristine and coated 254 stainless steel samples was determined using the fluorescence activated cell sorting (FACS) technique, which is one of the best, modern technologies for cell sorting.31 Finally, the apatite-forming ability of the coated samples was determined by the simulated body fluid (SBF) immersion test. The samples were taken out from SBF after 2 days of immersion, washed with distilled water, dried at 100°C, and then examined under SEM for apatite layer formation. It is believed that forming bone-like apatite in vitro is critical because apatite is the dominant inorganic phase of hard tissue. Thus, bone-like apatite formation indicates good osteoconductivity in biomaterials.32 Nevertheless, extensive in vivo studies are required to confirm the osseointegration, cytotoxicity, and biocompatibility of UNS S31254 stainless steel.

Results and Discussion

Surface Morphology Characterization

In Figs. S1, S2 and S3 of the supplementary material, the cross sections of clad, heat-affected zone (HAZ), substrate, and successful deposition of coating on the substrate are clearly visible.

Figure 1 presents the SEM micrographs of the cross-sectional microstructure of the coated sample. Austenite dendrites and some interdendritic phases are clearly visible in the coated samples as shown in Fig. 1. The SEM micrograph of Fig. 1 confirms that the interface between HAP coating and the 254SS substrate is sound and diffused, which are the characteristics of the laser deposited coatings.29 , 30 , 33,34, 35 Contrarily, HAP coatings deposited by other methods such as plasma spraying have a sharp interface, which is unacceptable for their long-term success.29

Fig. 1
figure 1

SEM micrograph of the interface microstructure of the HAP-coated 254SS sample

A typical SEM microstructure image of the HAP-coated 254SS sample is shown in Fig. 2. Calcium is present in interdendritic phases, chromium is present in austenite dendrites phases, and iron is present in both austenite dendrites and interdendritic phases as clearly demonstrated in Fig. 2.

Fig. 2
figure 2

SEM microstructure image of the HAP-coated 254SS sample

To acquire information about the composition and phase of the elements present in the deposited HAP coatings, EDS elemental mapping is carried out in this research. The EDS elemental mapping images of the HAP-coated 254SS sample are shown in Figs. S4, S5 and S6 of the supplementary material.

The consistent surface morphology of the LRM-coated 254SS sample at 2000 times magnification is shown in Fig. 3.

Fig. 3
figure 3

SEM micrograph of the coated 254SS sample at 2000 times magnification

The XRD diffractographs of the substrate, HAP powder, and HAP coating are presented in Fig. 4. The XRD results confirm the presence of HAP in crystalline format 2θ (31.66°, 34.48°, 35.98°, 39.04°, 47.26°, 56.58°, 62.74°, 64.44°) with minor traces of Fe3P at 2θ (41.64°).

Fig. 4
figure 4

XRD diffractograph of 254SS substrate, HAP powder, and HAP coating

Microhardness

Microhardness tests were conducted to confirm the coatings’ resistance to plastic deformation at a load of 300 g applied for 30 s as per Ref. 36. The hardness of 504 ± 6 HV of HAP-coated 254SS substrate was found to be significantly higher than the hardness of the bare substrate 265 ± 14 HV, as confirmed by the difference in size of the indents in the optical micrograph of Fig. 5.

Fig. 5
figure 5

Optical micrograph of the HAP-coated 254SS sample after microhardness test

Antibacterial Efficacy

The in vitro antibacterial efficacy of the bare and HAP-coated 254SS samples was investigated against Escherichia coli by using the previously used FACS technique.37 This is to confirm the reattainment of the antibacterial property in HAP coatings synthesized by LRM. Compared with the bare 254SS implant surface, the surface morphology of the composite layer of HAP produced on the implant surface helps in preventing bacterial attachment. This results in reduced bacterial infection, as confirmed by the FACS results and the literature.1 Figures S7 to S10 in the supplementary material show the FACS graphs that represent the relative data of E. coli death, i.e., 32.2% for the HAP-coated 254SS sample, with respect to the bare 254SS sample.

The main influencing factor in reducing bacterial infection is the increased surface roughness. As previously reported in the literature,38 the antibacterial effect can be put down to the scale of surface irregularities, which are comparable with the size of bacteria. The higher roughness seems to limit the anchoring points for bacteria and reduces the area in contact with their membrane.38

Bioactivity

The bioactivity of HAP-coated 254SS substrate was examined by conducting an immersion test in SBF. Figure 6 shows the surface morphology of the HAP-coated substrate after immersion in SBF. The growth of apatite layers is clearly visible after the immersion test, resulting in an average increase of nearly 0.16% in the weight of HAP-coated substrate.39

Fig. 6
figure 6

SEM micrograph of HAP-coated 254SS sample after immersion test

The overall results of the present research confirm that coating of HAP on 254SS by LRM is useful for orthopedic implant applications. It overcomes the limitations of plasma-sprayed HAP coatings, such as poor bonding strength and wear resistance.

Conclusion

In this research, a successful attempt has been made to treat the surface of 254SS, a high nitrogen stainless steel, with HAP using an LRM system. The coatings showed a composite microstructure with austenite dendrites and interdendritic HAP with small traces of Fe3P phase. A Vickers microhardness test confirms the higher hardness of HAP-coated 254SS substrate than that of the bare 254SS substrate. FACS tests confirm that the coated samples have a superior antibacterial property when compared with the pristine samples. An SBF immersion test confirms the rapid formation of apatite on the HAP-coated surface. The present research ably demonstrates that it is feasible to synthesize multifunctional orthopedic coatings that can prevent infection while still promoting osseointegration on 254SS implant surfaces using the LRM additive manufacturing process.