Skip to main content
Log in

Experimental Determination of the Maximum Permissible Coil Misalignment in Adaptive Transcutaneous Energy Transfer Systems

  • Published:
Biomedical Engineering Aims and scope

The effect of the symmetrical coil couple geometry on the value of lateral coil misalignment that corresponds to zero coupling in inductive energy transfer systems was studied. It was shown that the distance between coil turns and the number of coil turns have rather small effect on the zero-coupling distance. Increasing the outer diameter of the coils was shown to be the only way to increase the zero-coupling distance at given parameters of the system. It was found that changes in the outer diameter of the coils had more significant effect on the zero-coupling distance when the outer diameter was more than twice as great as the inner diameter.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Agarwal, K., Jegadeesan, R., Guo, Y.-X, and Thakor, N. “Wireless power transfer strategies for implantable bioelectronics: Methodological review,” IEEE Rev. Biomed. Eng., 10, 1-28 (2017).

    Article  Google Scholar 

  2. Danilov, A. A., Itkin, G. P., and Selishchev, S. V., “Progress in methods for transcutaneous wireless energy supply to implanted ventricular assist devices,” Biomed. Eng., 44, No. 4, 125-129 (2010).

    Article  Google Scholar 

  3. Zeng, F.-G., Rebscher, S., Harrison, W., Sun, X., and Feng, H., “Cochlear implants: System design, integration, and evaluation,” IEEE Rev. Biomed. Eng., 1, 115-142 (2008).

    Article  Google Scholar 

  4. Clark, G., Cochlear implants: Fundamentals and Applications, Springer (2003).

  5. North, R., “Neural interface devices: Spinal cord stimulation technology,” Proc. IEEE, 96, 1108-1119 (2008).

    Article  Google Scholar 

  6. Eldridge, P., Simpson, B. A., and Gilbart, J., “The role of rechargeable systems in neuromodulation,” Eur. Neurol. Rev., 6, No. 3, 187-192 (2011).

    Article  Google Scholar 

  7. Pelletier, B., Spiliopoulos, S., Finocchiaro, T., Graef, F., et al., “System overview of the fully implantable destination therapy – ReinHeart-total artificial heart,” Eur. J. Card.-Thor. Surg., 47, No. 1, 80-86 (2015).

    Article  Google Scholar 

  8. Weiland, J. D. and Humayun, M. S., “Visual prosthesis,” Proc. IEEE, 96, No. 7, 1076-1084 (2008).

    Article  Google Scholar 

  9. Li, X., Yang, Y., and Gao, Y., “Visual prosthesis wireless energy transfer system optimal modeling,” Biomed. Eng. Online, 13, No. 3, 1-11 (2014).

    Google Scholar 

  10. Burny, F., Donkerwolcke, M., Moulart, F., et al., “Concept, design and fabrication of smart orthopedic implants,” Med. Eng. Phys., 22, 469 (2000).

    Article  Google Scholar 

  11. Danilov, A. A., Mindubaev, E. A., and Selishchev, S. V., “Design and evaluation of an inductive powering unit for implantable medical devices using GPU computing,” Prog. Electromagn. Res. B, 69, 61-73 (2016).

    Article  Google Scholar 

  12. Danilov, A. A., Mindubaev, E. A., and Selishchev, S. V., “Space-frequency approach to design of displacement tolerant transcutaneous energy transfer system,” Prog. Electromagn. Res. M, 44, 91-100 (2015).

  13. Danilov, A. A., Mindubaev, E. A., and Selishchev, S. V., “Methods for compensation of coil misalignment in systems for inductive transcutaneous power transfer to implanted medical devices,” Biomed. Eng., 51, No. 1, 56-60 (2017).

    Article  Google Scholar 

  14. Zierhofer, C. M. and Hochmair, E.S., “Geometric approach for coupling enhancement of magnetically coupled coils,” IEEE Trans. Biomed. Eng., 43, No. 7, 708-714 (1996).

    Article  Google Scholar 

  15. Danilov, A. A., Mindubaev, E. A., and Selishchev, S. V., “The effect of transmitter coil size on the optimal implantation depth of receiver coil in transcutaneous inductive energy transfer systems,” Biomed. Eng., 51, No. 5, 354-357 (2017).

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to K. O. Gurov.

Additional information

Translated from Meditsinskaya Tekhnika, Vol. 51, No. 6, Nov.-Dec., 2017, pp. 40-43.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Gurov, K.O., Mindubaev, E.A., Selishchev, S.V. et al. Experimental Determination of the Maximum Permissible Coil Misalignment in Adaptive Transcutaneous Energy Transfer Systems. Biomed Eng 51, 432–436 (2018). https://doi.org/10.1007/s10527-018-9764-5

Download citation

  • Received:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10527-018-9764-5

Navigation