1 Introduction

Membrane-facilitated molecular separations play a key role in many large-scale industrial processes, ranging from seawater desalination to fruit juice clarification, and also in milliliter-scale sample processing in biomedical research laboratories (van Reis and Zydney 2007; Saxena et al. 2009; Kazemi et al. 2016). Miniaturization towards microliter volumes can be achieved by integrating membranes with microfluidic systems (de Jong et al. 2006; Chen et al. 2016; Roelofs et al. 2015; Wardrip and Arnusch 2016; Han and Hwang 2018). Microfluidic chips also enable in situ microscopy of the membrane, e.g., to study membrane fouling, and electrical or optical characterization of the permeate and retentate phases, facilitating separation performance evaluation of novel membrane materials and transport mechanisms (Roelofs et al. 2015; Kwak et al. 2013; Kornreich et al. 2014). In ultrafiltration, e.g., for protein concentration or buffer exchange, mass transport over the membrane is driven by pressure-induced flux; while mass transport in a dialysis process is primarily determined by concentration gradients over the membrane (Saxena et al. 2009).

Dialysis efficiency is significantly enhanced by minimizing the diffusion distance to the membrane and by maintaining a steep concentration gradient (Roelofs et al. 2015). Microdialysis flow cells offer the advantages of a sub-millimeter height of the feed channel and a continuous flow of fresh solution, buffer or pure water, through the dialysate channel. In a theoretical study, Yeh and Hsu evaluated the mass transfer efficiency for three different flow configurations (Yeh and Hsu 1999). Counter-current flow, where the sample and the dialysate phase flow in opposite directions along the membrane, gave the highest mass exchange; while co-current flow was the least effective. Cross-flow, implying that the dialysate stream is perpendicular to the sample stream (e.g., crossing a bundle of hollow fiber membranes), had an intermediate performance (Yeh and Hsu 1999). It was also demonstrated experimentally, for a flow cell with a planar microporous cellulose ester membrane, that counter-current operation results in more efficient dialysis than co-current flow (Yeh and Chang 2005).

Microfluidic dialysis cells have been described for a variety of applications, including small-molecule, peptide or protein extraction from protein mixtures (Perozziello et al. 2015), cell lysates (Xiang et al. 1999; Liu and Verma 1999) or blood (Kurita et al. 2006) for spectroscopic/spectrometric characterization, the exchange of metabolites and inhibitory species for cell-free protein synthesis (Timm et al. 2016), and the ionic strength modulation of oligonucleotide separation (Sheng and Bowser 2012) and virus phase transitions (Kornreich et al. 2014). Furthermore, many microfluidic devices, incorporating hollow fiber (Liu and Verma 1999; Canarelli et al. 2002; Jakubowski et al. 2005; Sun et al. 2008), planar polymer (Xu et al. 1998; Liu et al. 1998; Xiang et al. 1999) or planar inorganic (Tibavinsky et al. 2015; Wu et al. 2018) membranes, have been developed for the on-line desalting of biological samples for mass spectrometry analysis. The membrane pore size is always determined by the separation requirements, while the channel dimensions follow from the typical sample volume and the microfabrication methodology. The counter-flow configuration is most commonly employed, but the effect of the flow rate is typically not described, or is only investigated within a limited range, while the separation outcome tends to be assessed indirectly, for example as the signal-to-noise ratio in a mass spectrum (Xu et al. 1998; Liu and Verma 1999; Canarelli et al. 2002; Wu et al. 2018).

To establish to which extent the separation performance can be modulated by the hydrodynamic conditions, we quantified salt transfer over a regenerated cellulose membrane in a laser-machined microfluidic counter-flow dialyzer for a wide range of feed and dialysate flow rates. We obtained retentate salt concentrations between 0.1 and 99% of the input salt concentration, indicating that with the appropriate water and sample flow settings, any desired salt transfer efficiency can be realized. Theoretical mass transfer analysis for the counter-flow dialysis geometry closely matched the experimental observations, consistent with a diffusion-limited membrane transport mechanism. At a high protein concentration of 50 mg/mL albumin, dialysis performance was only marginally affected over a period of 16 h, suggesting minimal fouling of the 5 kDa MWCO cellulose membrane. Flow-controlled microfluidic diffusion dialysis, hence, enables applications such as tuneable desalting of biomolecular samples, where a physiological salt concentration interferes with analytical assays but too low a salt concentration induces protein denaturation or modifies biomolecular interactions (Xu et al. 1998; Zhou and Pang 2018).

2 Experimental

2.1 Materials

Planar regenerated cellulose and polyethersulfone ultrafiltration (UF) membranes with molecular weight cut-off (MWCO) values of 1, 5 and 10 kDa were obtained from Sartorius Stedim Biotech as membrane disks of 25 mm diameter and 180 µm thickness. Polycarbonate track-etch membrane filters of 25 mm diameter and 20 µm thickness, with a 50 nm pore diameter and 1.18% porosity, were from Avanti Polar Lipids. Acrylic sheets of 5 mm thickness were purchased from RS Components. Sodium chloride and fluorescently (fluorescein isothiocyanate) labeled bovine serum albumin (BSA) were from Sigma Aldrich. Deionized water from a Milli-Q system was used for all experiments. Prior to use, to remove the humectant, the membranes were placed in a beaker containing deionized water for 1 h, after which they were rinsed with fresh water and gently wiped with tissue paper.

2.2 Membrane characterization

Membrane thickness was verified by scanning electron microscopy of sputter-coated membrane cross sections with a Zeiss EVO LS25 microscope. The porosity of the UF membranes was determined by conductometry (Garg et al. 2012, 2014). After washing with deionized water, membranes were clamped between two Delrin chambers of 1 mL volume with a 5 mm diameter aperture in the side wall, where the membranes were in contact with an electrolyte solution of 10 mM NaCl. Ag/AgCl electrodes of 0.8 mm diameter were inserted in each chamber, DC voltages up to 1 V were applied and the current flowing over the membrane was measured. The conductance of the electrolyte-exposed membrane area was then determined from the IV relationship. The porosity of the UF membranes was estimated according to:

$$P_{\text{UF}} = P_{\text{ref }} \frac{{L_{\text{UF}} \times G_{\text{UF}} }}{{L_{\text{ref}} \times G_{\text{ref}} }},$$
(1)

where G is the conductance, L is the pore length (i.e., membrane thickness) and P is the porosity of the membrane. From the conductance measurements of the UF and track-etch membranes, the porosity of the UF membranes can be approximated, neglecting the influence of pore geometry and assuming no distinct interactions of ions with the membrane material, using the known porosity and thickness of a track-etch membrane as reference values (Kumar and Chakarvarti 2008).

2.3 Counter-flow dialysis cell

A microfluidic dialyzer was constructed by clamping a planar membrane between two optically transparent plates of poly(methyl methacrylate) (PMMA) with micromachined channels. Acrylic sheets of 5 mm thickness were cut into plates of 3 × 3 cm with an Epilog Mini 30 W CO2 laser cutter, and a screw hole was cut in each corner of the plate. In the top plate, a feed channel of 0.2 mm depth and 1.8 mm width was laser ablated, while a water channel of 3.5 mm depth and 3.0 mm width was defined in the bottom plate. Both channels are of spiral geometry, with a length of ~ 75 mm, giving a feed and a water channel volume of approximately 25 and 800 µL, respectively. These different dimensions imply a different relation between the volumetric flow rate and the linear velocity for the feed and water channels, as indicated in the text. At the start and end of each channel, a 0.8 mm diameter through-plate hole was laser machined as a tubing connector. A water-washed circular membrane sheet was positioned on the bottom plate and then the top plate was put into place. Both plates were brought into tight contact by fastening the M4 screws in the corners of the device. The membrane surface in contact with the feed channel is ~ 1.35 × 10−4 m2. Figure 2a shows an assembled dialyzer with connected microfluidic PTFE tubing (inner and outer diameter 0.3 mm and 0.8 mm, respectively). Counter-flow dialysis was implemented with Harvard Apparatus 11 Plus syringe pumps, with the feed and the water phase flowing in opposite directions along the membrane. The conductivity of aliquots of the retentate outflow was measured with a Horiba B-173 conductivity meter with a 100 µL sample compartment and converted to NaCl concentration by a calibration curve. All experiments were performed at room temperature.

2.4 Mass transfer analysis

For urea dialysis over a microporous cellulose ester membrane in a planar counter-flow dialyzer, Yeh and Zhang observed good agreement between experimental results and theoretical predictions from a mass transfer model (Yeh and Chang 2005). This model is based on the classic solutions from Graetz and Lévêque for heat transfer to tube walls for a laminar flow, which can be extrapolated to mass transfer because of the identical forms of the convection–diffusion equation (Kirtland 2010). As an idealization, it is assumed that the solute is consumed at the channel wall, i.e., the membrane in a dialyzer, that the time scale for solute diffusion is longer than the time scale of axial convection (~ flow rate), and that there is no radial convection. This enables calculation of the diffusive solute flux to this boundary wall and its scaling with the flow rate as a function of the distance from the feed channel inlet, which takes the form of a 1/3 power law for the parabolic velocity profile of a fully developed laminar flow (Kirtland 2010; Haase et al. 2015). Here, we follow the model derived by Yeh et al. (Yeh and Hsu 1999; Yeh and Chang 2005), where the total mass transfer rate M (mol/s) in a planar counter-flow dialyzer is given by Eq. (2). This analysis also assumes: steady-state diffusion, uniform solute concentration over the cross section of flow, a constant overall mass transfer coefficient, constant flow rates, constant physical properties of the fluid and no chemical reactions (Yeh and Chang 2005).

$$M = \frac{{\left( {C_{\text{f,in}} - C_{\text{w,in}} } \right) \left( {1 - {\text{e}}^{{ - K S \left[ {\frac{1}{{Q_{\text{f}} }} - \frac{1}{{Q_{\text{w}} }}} \right]}} } \right)}}{{\left( {\frac{1}{{Q_{\text{f}} }}} \right) - \left( {\frac{1}{{Q_{\text{w}} }}} \right) {\text{e}}^{{ - K S \left[ {\frac{1}{{Q_{\text{f}} }} - \frac{1}{{Q_{\text{w}} }}} \right]}} }}$$
(2)

Here, Cf,in is the solute concentration (mol/m3) of the pre-dialysis feed solution and Cw,in of the pre-dialysis water flow (i.e., no solute). Qf and Qw are the flow rates (m3/s) of the feed and the water phase, respectively, and S is the feed-exposed mass transfer area of the membrane sheet (m2). The overall mass transfer coefficient K (m/s) is:

$$\frac{1}{K} = \frac{1}{{k_{\text{f}} }} + \frac{1}{{k_{\text{m}} }} + \frac{1}{{k_{\text{w}} }},$$
(3)

where kf and kw are the solute mass transfer coefficients for the feed and the water phase, respectively, and km for the membrane, derived as:

$$k_{\text{f}} = 0.816\left( {\frac{{6 Q_{\text{f}} D_{\text{f}}^{2} }}{{W_{\text{f}} H_{\text{f}}^{2} L_{\text{f}} }}} \right)^{{\frac{1}{3}}},$$
(4)
$$k_{\text{w}} = 0.816\left( {\frac{{6 Q_{\text{w}} D_{\text{w}}^{2} }}{{W_{\text{w}} H_{\text{w}}^{2} L_{\text{w}} }}} \right)^{{\frac{1}{3}}},$$
(5)
$$k_{\text{m}} = \frac{{D_{\text{m}} \varepsilon }}{\tau t}.$$
(6)

Here, W, H and L are the width, height and length (m) of the feed or the water channel, and D is the diffusion coefficient (m2/s) of the solute in each phase, as indicated by the subscript. The structural parameters of the membrane are included as membrane porosity ε, tortuosity τ and thickness t (m) (Yeh 2008).

The absolute salt ion mass (mol) transferred from the feed to the water phase is obtained from the total mass transfer rate M (mol/s) by considering the membrane contact time (s) of a segment of the laminar feed stream at a given flow rate. For example, given the ~ 25 µL volume of the feed channel, a feed flow of 10 µL/min corresponds to a dialyzer flow-through time of ~ 2.5 min (150 s). The theoretical post-dialysis salt concentration of the feed solution can then be calculated by subtracting the theoretical amount of salt transferred to the water phase from the pre-dialysis amount of salt in the feed channel volume.

3 Results and discussion

3.1 Membrane characterization

Membranes with a nominal MWCO of 1–10 kDa are permeable to salt ions but reject most protein species. Cross sections of regenerated cellulose UF membranes were visualized with electron microscopy, revealing the typical ultrastructure of membranes synthesized by phase inversion (Cuperus and Smolders 1991). Figure 1a shows a UF membrane of ~ 20 µm thickness on top of its ~ 180 µm thick microporous support sheet. The asymmetric structure of the UF membrane is apparent at higher magnification (Fig. 1b), showing a dense skin layer of ~ 2 µm thickness on top of a ~ 20 µm thick sub-layer with a ~ 200 nm diameter pore structure. The pores in the sample-facing skin layer, which determine the separation performance, will have a diameter of ~ 2 nm (Cuperus and Smolders 1991; Romero et al. 2013).

Fig. 1
figure 1

a Scanning electron microscopy image of a cross section of a regenerated cellulose (RC) ultrafiltration membrane on top of its microporous support sheet. b Cross section of the membrane itself, with the dense skin layer visible in the top left corner and the sub-layer, partially covered by debris, in the center of the image. c Current–voltage relationships for RC and polyethersulfone (PES) ultrafiltration membranes of various MWCO specifications and a reference polycarbonate (PC) track-etch membrane of known porosity

The porosity of the membranes was estimated by electrical measurements. The IV relationship for the UF membranes and for the reference track-etch membrane is depicted in Fig. 1c. The conductance of 1, 5 and 10 kDa polyethersulfone membranes was 104, 116 and 173 µS, respectively; while the conductance of the 50 nm pore diameter track-etch membrane was 184 µS. Estimating the UF and the track-etch membrane thickness as 20 µm and taking the porosity of the reference track-etch membrane, as specified by the supplier, as 1.18%, the porosity of the 1, 5 and 10 kDa membranes can be approximated as, respectively, 0.67, 0.74 and 1.11%. The conductance measured for a 5 kDa regenerated cellulose membrane was 112 µS, corresponding to a porosity of 0.72%. These values agree with previously reported values for UF membrane porosities, which range from 0.1 to 10% but are typically around 0.5% (Cuperus and Smolders 1991).

3.2 Effect of water flow rates on static sample dialysis

Regenerated cellulose membranes of 5 kDa MWCO were clamped between two acrylic plates with spiral microfluidic channels defined by laser machining, matching the circular outline of the membrane (Fig. 2a). Tight contact between the top and bottom plate ensured a good seal with the polymer membrane, as previously demonstrated for other microfluidic dialyzers (Kurita et al. 2006; Sheng and Bowser 2012; Perozziello et al. 2015). Initial desalting experiments were performed with a static feed solution of 150 mM NaCl and 2 mg/mL fluorescently labeled BSA (to visually verify BSA retention in the feed channel) with co-lyophilized buffer ions (170 mM total salt) and a static dialysate phase of deionized water, which was refreshed at different time intervals. After removal from the dialyzer, the conductivity of the dialysate was measured to determine the amount of membrane-permeated salt ions, enabling estimation of the remaining salt concentration in the feed chamber after each refreshment step.

Fig. 2
figure 2

a Counter-flow microfluidic dialyzer with an ultrafiltration membrane sealed between 3 × 3 cm acrylic plates with laser-machined feed and water channels in a spiral geometry. b Residual salt concentration of a static sample phase as a function of periodic refreshment, at different time intervals, of a static water phase

Figure 2b shows that, in general, a higher frequency of water refreshment (i.e., shorter water phase resident time in the dialyzer) results in a lower residual salt concentration in the feed chamber. For example, after 15 min, when the water phase has been refreshed one, two, three or 15 times, the residual salt amounts to 25%, 19%, 14% or 8%, respectively, of the original concentration. This corresponds to a molar solute flux across the membrane of 2–3 × 10−5 mol/m2/s. Figure 2b also illustrates that the initial water refreshment steps yield the most pronounced reduction in salt concentration of the static feed sample. This is expected because the feed/dialysate salt concentration gradient, the driving force for dialysis, is largest at the initial stage of desalting. When the water phase is refreshed every minute, the salt concentration in the feed sample is reduced to approximately 1 mM after 30 min. By varying the amount of time that the water phase is in contact with the static salt sample and the rate at which the water is refreshed, the extent of desalting can be modulated.

3.3 Effect of flow rates on counter-flow dynamic dialysis

A continuously flowing feed stream enables higher sample throughput and facilitates dialysis quantification by in-line conductivity measurements of the post-dialysis retentate phase, i.e., the desalted feed. Deionized water continuously flowing in the opposite direction enables constant refreshment of the permeate (‘water’) phase, maintaining maximal concentration gradients. To quantify the separation performance of the counter-flow dialyzer, we determined the post-dialysis retentate salt concentration for a wide range of flow conditions: 10–600 µL/min (0.016–0.95 mm/s linear velocity) for the water phase and 0.5–50 µL/min (0.023–2.3 mm/s) for the feed phase. Data points represent single measurements. For a number of flow combinations, repeat experiments were performed, indicating a variation of a few mM.

Figure 3a depicts the percentage of salt retained at the feed side of the 5 kDa membrane at a fast water flow of 600 µL/min, which approximates the highest frequency of water refreshment in the quasi-dynamic dialyzer operation (Fig. 2b), and at a slow flow of 10 µL/min. For the 600 µL/min water flow, it can be seen that a slow feed flow of 0.5 µL/min leads to almost complete desalting (0.2 mM retentate salt, i.e., 0.1% of salt retained) and a fast feed flow of 50 µL/min to minimal desalting (92% of salt retained). Intermediate feed flow rates result in intermediate salt transfer, suggesting that any particular desalting efficiency can be obtained with the appropriate feed flow rate. A typical syringe pump can set the flow rate in increments of 0.01 µL/min.

Fig. 3
figure 3

Retentate salt concentration after dynamic counter-flow dialysis with a 5 kDa MWCO regenerated cellulose ultrafiltration membrane at various feed and water flow rate combinations, as a percentage of the initial salt concentration. a Feed flows from 0.5 to 50 µL/min at water flow rates of 10 and 600 µL/min. b Water flows from 10 to 600 µL/min at feed flow rates of 0.5 and 1.0 µL/min. c Feed flows from 0.5 to 50 µL/min at water flows of 10, 20, 50, 150, 300, 500 and 600 µL/min. Lines connecting data points are a guide to the eye. d Water flows from 10 to 600 µL/min at six different feed flow rates between 2.5 and 50 µL/min. The sample is 150 mM NaCl with 2 mg/mL fluorescently labeled BSA with co-lyophilized buffer ions: 170 mM total salt. Note that volumetric feed and water flow rates correspond to different flow velocities because of the larger volume of the water channel

Figure 3a also shows that the water flow rate modulates the dialysis performance. At a slow water flow of 10 µL/min, there is more salt retained in the sample for all feed flow rates. Salt transfer over the membrane is minimal for feed flows above 20 µL/min (0.93 mm/s) but the lowest flow rate of 0.5 µL/min results in efficient desalting (0.6% of salt retained). Figure 3b shows the percentage of salt retained in the feed stream at the lowest sample flow rates of 0.5 and 1.0 µL/min, where salt transfer is extremely efficient. At 0.5 µL/min, over 99% of the salt ions diffuse over the membrane to the water phase for the entire water flow range of 10–600 µL/min. At 1.0 µL/min (0.046 mm/s) feed flow this necessitates water flow rates from 50 µL/min (0.079 mm/s), with a 10 and 25 µL/min water flow giving 6.5% (outside the frame of Fig. 3b) and 1.8% retentate salt, respectively.

For water flow rates of 10, 50, 150, 300, 500 and 600 µL/min, the retentate residual salt percentage is shown as a function of the feed flow rate in Fig. 3c. The residual salt fraction is plotted as a function of the water flow rate in Fig. 3d. For all water flow rates, the salt dialysis efficiency clearly depends on the sample flow rate; a slower sample flow results in a lower retentate salt concentration (Fig. 3c), i.e., more efficient salt transfer over the membrane. At higher sample flow rates, the efficiency is reduced because the residence time of the sample in the feed channel becomes shorter, as previously observed for other microdialyzers (Jakubowski et al. 2005; Sheng and Bowser 2012). A short diffusion distance to the membrane is crucial for efficient dialysis (Roelofs et al. 2015). The Brownian motion diffusion coefficient of NaCl at ~ 1–150 mM concentration at 25 °C is ~ 1.5 × 10−5 cm2/s (Robinson and Stokes 2002), which corresponds to a net ion displacement of ~ 400 µm/min; while the height of the feed channel is 200 µm. This implies that for optimal salt transfer efficiencies, the contact time of the sample stream with the membrane should be at least 30 s. Given the 75 mm length of the feed channel, this implies a linear velocity of at most 2.5 mm/s, i.e., a sample flow rate of at most 2.5 µL/min.

At a particular sample flow rate, a faster water flow, in the opposite direction, generally improves salt transfer, but this modulation by the water counter-flow is modest (Fig. 3d). For example, at a sample flow of 10 µL/min (0.46 mm/s), for the slowest water flow of 10 µL/min (0.02 mm/s) ~ 57% of salt is retained in the feed; while this is ~ 44% for the highest flow rate of 600 µL/min (0.95 mm/s). For several feed flow rates, the desalting efficiency improves most when the water counter-flow is increased from 10 to ~ 150 µL/min (Fig. 3d). This is due to the relative residence times of the feed and the water phase (Yeh and Hsu 1999); at the lowest water flow rates, the linear velocity of the water stream is considerably smaller than the velocity of the feed stream. For example, the 10 µL/min feed velocity is nearly 30 times faster than the 0.016 mm/s velocity of the 10 µL/min water flow. Under these conditions, the water counter-flow is, hence, somewhat static with respect to the feed flow, which diminishes the salt concentration gradient over the membrane. Increasing the water flow velocity, up to 0.32 mm/s (200 µL/min), can, thus, noticeably enhance the dialysis efficiency.

3.4 Mass transfer analysis

The mass transfer over the membrane in the counter-flow dialyzer was modeled as described in Sect. 2.4. Mass transfer coefficients for the feed and the water phase were calculated for flow rates up to 50 and 600 µL/min, respectively. For km we used the experimentally determined membrane porosity of 0.72% and a membrane skin layer thickness of 2 µm (Sect. 3.1). Ultrafiltration membranes have a 3D pore network; hence, the tortuosity was approximated as 2 (Kokubo and Sakai 1998; Bakeri et al. 2015). The diffusion coefficient of the salt ions inside the pores was assumed to be identical to bulk solution, i.e., 1.5 × 10−5 cm2/s (Robinson and Stokes 2002), giving a km of 2.7 × 10−6 m/s. The overall mass transfer coefficient K (m/s) for transport from the feed to the water phase was then obtained according to Eq. (3), and subsequently the overall mass transfer rate M (mol/s) was derived according to Eq. (2).

These theoretical values for K and M are shown in Fig. 4a, b, respectively, for the experimental water flow rates as well as for smaller values of 1 and 5 µL/min, as a function of the sample flow rate (0.5–50 µL/min). K varies between ~ 0.3 and 1.2 µm/s, increasing with faster sample and water flows. The steep increase in M for sample flow rates up to ~ 10 µL/min represents the transition from a static to an effective dynamic feed phase, substantially increasing the salt concentration gradient over the membrane. Also the mass transfer increase with higher water flow rates, from ~ 3 to 25 nmol/s, can be related to a larger concentration gradient. However, the effect of operating the counter-flow dialyzer at a water flow beyond 300 µL/min is relatively small; the salt gradient is not expected to increase significantly in this fast flow regime.

Fig. 4
figure 4

a Theoretical overall mass transfer coefficient K and b theoretical mass transfer rate M for the counter-flow dialyzer for feed flows from 0 to 50 µL/min at discrete water flow rates between 1 and 600 µL/min. c Comparison between the theoretically predicted (connected blue markers) and the experimentally obtained (black markers) retentate salt concentration for feed flow rates of 1, 2.5, 5, 10, 20, 30 and 50 µL/min

For selected feed flow rates, the modeled post-dialysis salt concentration of the feed phase, obtained from M as described in Sect. 2.4, was compared with the experimentally determined concentrations. Figure 4c shows that the theoretical and experimental values are well matched, generally deviating by at most ~ 10% retentate salt concentration. Such good agreement between diffusion theory and experiment was also observed for counter-flow dialysis of urea over a cellulose ester membrane (Yeh and Chang 2005), whose model was applied here, and of fluorescein over a track-etched polycarbonate membrane (Sheng and Bowser 2012), as well as for co-flow dialysis of salt ions over a nanoporous alumina membrane (Tibavinsky et al. 2015). The models of the latter two studies likewise assume advection-dominated solute transport along the feed channel and diffusion-dominated transport over the membrane (i.e., minimal transmembrane water flow and hence no or negligible solute advection over the membrane), with feed flow rates determining the membrane contact time and water flow rates modulating the salt concentration gradient over the membrane.

3.5 Dialysis performance at high protein concentration

Counter-flow salt dialysis was also performed with a sample of 170 mM salt and 50 mg/mL (0.75 mM) BSA (2 mg/mL fluo-BSA and 48 mg/mL non-labeled BSA), approximating the total protein concentration in human blood serum, using membranes with different MWCO values. At the lowest sample flow rate of 0.5 µL/min, the post-dialysis retentate conductivity was found to correspond to ~ 1% of the input salt concentration for 5 and 10 kDa MWCO membranes and to ~ 2% for a 1 kDa MWCO membrane, for all investigated water flow rates (in the range of 10–750 µL/min, data not shown). For the three membranes, Fig. 5 depicts the retentate salt concentration for sample flow rates from 0.5 to 50 µL/min at a fast water flow. As previously observed for a model sample with 2 mg/mL BSA and a dialysis cell with a 5 kDa MWCO membrane (Fig. 3a), the dialysis efficiency depends on the feed flow rate. At feed flows from 10 µL/min, compared to the 5 kDa MWCO membrane, the 10 kDa membrane results in ~ 10% more salt transfer to the water phase, with respect to the input salt concentration, and the 1 kDa membrane in ~ 10% less. Qualitatively, this trend reflects the different porosities of these membranes (Sect. 3.1).

Fig. 5
figure 5

Dynamic counter-flow dialysis with 1, 5 and 10 kDa polyethersulfone ultrafiltration membranes of a feed solution with 170 mM salt and a protein concentration of 50 mg/mL BSA, for feed flow rates in the range of 0.5–50 µL/min at a water flow rate of 750 µL/min

For feed flow rates up to 30 µL/min, the residual salt percentage obtained with a 5 kDa MWCO membrane is ~ 15% higher than for the model sample with 2 mg/mL BSA (Fig. 3a), indicating reduced mass transfer due to membrane fouling associated with the high protein concentration of 50 mg/mL. It should be noted that BSA has been employed as a model foulant in studies of ultrafiltration cells (Wang and Tang 2011), which are prone to fouling because of the applied transmembrane pressure. To assess any further decrease in separation performance over time, a counter-flow dialysis cell with a 5 kDa MWCO membrane was operated continuously for 16 h with a feed and water flow rate of 0.5 and 750 µL/min, respectively. Towards the end of this period, the salt concentration of the retentate had increased by just ~ 0.1 mM with respect to the start of the dialyzer operation. Hence, any membrane fouling does not become more severe by prolonged counter-flow dialysis, where the feed flowing parallel to the membrane surface is expected to minimize membrane fouling because of shear stress (Casey et al. 2011).

4 Conclusions

We characterized the dialysis performance of a microfluidic counter-flow dialyzer with a ~ 25 µL feed and a 800 µL dialysate channel of 75 mm length, separated by a commercial 5 kDa MWCO cellulose membrane. The feed and dialysis flow rates determine the retentate salt concentration, indicating that the separation efficiency can be precisely controlled by setting these flow rates. Specifically, with a feed flow between 0.5 and 50 µL/min (0.023 and 2.3 mm/s), the retentate salt concentration varies from ~ 0.1 to ~ 99% of the input sample of 170 mM salt. For a given feed flow rate, a faster water flow also improves salt transfer by increasing the concentration gradient over the membrane. But because the counter-flow configuration is very efficient in realizing a steep gradient, this effect is relatively modest (up to a ~ 20% improvement). Mass transfer modeling shows that a faster sample flow actually also results in a higher mass transfer rate (up to ~ 5–10 nmol/s) because in a counter-flow geometry, this again increases the concentration gradient. However, the dominant effect of a faster sample flow is a shorter dialyzer flow-through time and hence a reduced membrane contact time of the feed phase, resulting in less mass transfer and hence less efficient desalting of the feed flow.

The time required for desalting a sample by counter-flow dialysis could be reduced by implementing a feed channel shallower than 200 µm, reducing the diffusion distance to the membrane; while the volume throughput of the dialyzer could be maintained with a wider feed channel or multiple parallel channels. However, sample volumes of ~ 100 µL are sufficient for many analytical techniques and can be dialyzed in a reasonable timeframe with the present geometry. For longer flow-through dialysis times (i.e., larger sample volumes), the desalting performance did not significantly deteriorate with 50 mg/mL of the model foulant protein BSA, which is a major advantage of dialysis with respect to a pressure-driven separation process such as ultrafiltration (Shi et al. 2014). Salt transfer efficiency could also be increased, and process time decreased, by employing a membrane of higher porosity. For the 1–10 kDa MWCO regime which enables sample retention (membrane rejection) of biomarker proteins, planar membranes as supplied for ultrafiltration applications are of very low porosity, but the synthesis of nanoporous membranes of reduced thickness and higher porosity is an active research area (DesOrmeaux et al. 2014; Mireles and Gaborski 2017; Wang et al. 2017; Wu et al. 2018).

Generally, our rapid prototyping approach of clamping membranes between rigid laser-machined acrylic plates enables implementation of tuneable dynamic dialysis for any planar membrane material and any diffusion-driven separation process. Specifically, we anticipate that tuneable desalting will benefit molecular analysis methods such as field effect transistor biosensing (Noor and Krull 2014; Kaisti 2017; Lowe et al. 2017), where a reduced salt concentration is required for sensing but excessive desalting can disrupt biomolecular structure and interactions.