Abstract
Stability-indicating LC methods were developed and validated for the quantitative determination of doripenem, meropenem and tebipenem in the presence of their degradation products formed during forced degradation studies. Isocratic HPLC and UHPLC separations were performed with a core–shell Kinetex 1.7, 2.6 and 5 µm, all C18, 100A, 100 × 2.1 mm columns and the mobile phase composed of acetonitrile and 12 mmol L−1 ammonium acetate in different ratios. The flow rates of the mobile phase were: 0.5 mL min−1 for 1.7 µm column, and 1.0 mL min−1 for 2.6 and 5 µm ones. Detection wavelength was 298 nm and temperature was set at 30 °C. All analysed drugs were exposed to stress conditions which caused their hydrolysis and thermal degradation. The methods were validated by evaluation of linearity, accuracy, precision, selectivity and robustness. Proposed methods were successfully applied for the determination of investigated antibiotics during kinetic studies in aqueous solutions and in the solid state. The advantages of chromatographic procedures which are based on the use of C18 stationary phases with different particle sizes in the analysis of selected carbapenems were discussed.
Avoid common mistakes on your manuscript.
Introduction
Significant susceptibility of some drugs to degradation requires proper selection of conditions of chromatographic separation. During the optimization of chromatographic procedure, the following features should be taken into account: limitation of reagents, which can catalyze degradation of labile analytes, short run time and achievement of required validation parameters. Among all validation parameters, the most essential is achievement of selectivity [1]. The achievement of suitable selectivity is critical parameter during method validation when the kind of degradation products depends on the affecting factors. Carbapenems doripenem (DOR), meropenem (MER) and tebipenem (TEB) were chosen as model drugs, which exhibit a significant instability in aqueous solutions and in the solid state. Available studies report that in the case of abovementioned antibiotics, formed degradation products depend on the concentration of investigated drug, pH of the solution, temperature and presence of acids and bases. It has been proved that different products as the result of carbapenem degradation were formed in solid state as well as in aqueous solutions [2]. The main degradation pathway of carbapenems degradation leads to the cracking of β-lactam bond in a bicyclic-fused rings structure. Existing papers have been reported that most of the chromatographic procedures were based on columns filled with 5 μm particles [2–7]. Application of columns with smaller particle sizes was reported for only a few β-lactam antibiotics [8–10]. However, the benefits of such analytical methods can be significant, taking into account higher reduction of the run time of analysis and shorter contact with reagents, which cause the hydrolytic degradation of investigated substances during sample preparation and chromatographic analysis [11, 12]. Therefore, it is such an important study of transfer from HPLC procedure to UPLC separation in the case of labile drugs. The aim of our studies was to select the best analytical conditions for the determination of labile carbapenem analogues, in the presence of their degradation products.
Experimental
Standards and Reagents
DORIBAX™ (Janssen-Cilag International NV, Beerse, Belgium) containing doripenem monohydrate as an anhydrous basis form was used as doripenem standard. Meropenem (purity >98 %) was obtained from CHEMOS (Regenstauf, Germany). Tebipenem (purity >98 %) was supplied by Pharmachem International (Wuhan, China). All other chemicals and solvents were obtained from Merck KGaA (Darmstadt, Germany) and were of analytical grade. High-quality pure water was prepared using a Millipore Exil SA 67120 purification system (Millipore, Molsheim, France).
Instrumentation
The Dionex Ultimate 3000 analytical system consisted of a quaternary pump, an autosampler, a column oven and a diode array detector. Three Kinetex, C18, 100A, 100 × 2.1 mm columns (Phenomenex, Torrance, USA) with different particle sizes (1.7, 2.6 and 5 µm) were tested. The mobile phase composed of acetonitrile and 12 mmol L−1 ammonium acetate in different ratios (Table 1). The flow rate of the mobile phase was 1.0 mL min−1. The wavelength of the DAD detector was set at 298 nm. Separation was performed at 30 °C.
Procedure for Forced Degradation Study of API’s
Degradation in Aqueous Solutions
The degradation of analysed API’s (active pharmaceutical ingredient) in aqueous solutions was studied at 293 K in sodium hydroxide solution (0.3 mol L−1) and at 353 K in hydrochloric acid (0.3 mol L−1). The ionic strength of all solutions was adjusted to 0.5 mol L−1 with a solution of sodium chloride (4.0 mol L−1). Degradation was initiated by dissolving an accurately weighed 5.0 mg of analysed drug in 10.0 mL of the solution equilibrated to the desired temperature in stoppered flasks. At specified times, samples of the reaction solutions were neutralised and instantly cooled with a mixture of ice and water.
Degradation in Solid State
5.0 mg samples of analysed API’s were weighed into 5 mL glass vials. To evaluate the influence of temperature and humidity, the samples were placed in a desiccator containing saturated solution of sodium chloride (RH ~76.4 %) that was in incubators (Wamed, Warsaw, Poland) set to temperature 333 K. At specified time intervals, determined by the rate of degradation, the vials were removed, cooled to room temperature and their contents were dissolved in water. The so-obtained solutions were quantitatively transferred into measuring flasks and diluted with the same solvent to 10.0 mL.
Results and Discussion
Chromatographic studies of β-lactam analogues are complicated due to their significant susceptibility to degradation, as well as possibility of catalytic effect of their degradation products to stability of parent compound.
In this article, different analytical procedures based on core shell C18 stationary phases with various particle sizes were evaluated. The fully porous C18 columns filled with 5 µm particles allowed the determination of analysed carbapenems in about 15 min [2–8]. Employment of core shell 5 µm particles technology leaded to essential reduction of retention times (t R) (Table 1). The determination of carbapenem analogues based on the core shell C-18 columns with particles 2.6 µm did not give significant benefits in regard to the reduction of retention times. However, this application produces higher pressures. In the case of determination of carbapenems analogues using core shell C-18 columns with 1.7 µm particles, the retention times represented similar values as in separation based on stationary phase with greater particles, but simultaneously flow rate was twice less (Table 1). This approach is compromised between reduction of consumption organic solvents and high pressure which appears in developed method with 5 µm particles.
Validation of analytical methods for the determination of carbapenems using C-18 stationary phase with different particle sizes was conducted in regarding to linearity, precision, accuracy, LOD and LOQ. The significant reduce t R of carbapenem after application of core shell columns did not influence on separation of main peaks (meropenem, doripenem and tebipenem) from degradation products. Suitable symmetry of peaks of main substances as well as related products was unchanged. Order of time retentions for all analytes was the same as in the separation based on fully porous stationary phases. For all the analogues of carbapenems, the linear relationships between concentration of analytes and response of detector were achieved. The widest ranges of linear response were obtained when carbapenems were determined with usage of core shell columns with 5 µm particles. Those determinations were characterised by the most satisfied regression coefficients. Table 2 shows linear response function of concentration over the range 0.5–3.0 µg and regression coefficients (r) were >0.995 for all the carbapenems. Similar results were obtained for precision studies. It was shown that the lowest values of intra-day and inter-day precision were calculated in the cases of determination of DOR, MER and TEB using C-18 column filled with 5 µm particles. Method repeatability data as accuracy are summarised in Fig. 1.
Decreasing particle sizes from 2.6 to 1.7 µm in core shell columns did not significantly influence RDS % values characterising intra-day and inter-day precision. The accuracy of determination of carbapenem analogues was established at three concentration levels: 80,100 and 120 % of label claim. However, significant differences between accuracy of determinations of carbapenems analogues were not observed (Fig. 2a–c). The lowest values of LOD and LOQ for carbapenems were obtained when their determinations were conducted on core shell column filed with 5 µm particles.
Conclusions
In the presented study, the influence of the particle size of core shell stationary phase on parameters of labile drugs determination was examined. An achievement of the most satisfactory validation parameters was possible when the doripenem, meropenem and tebipenem were determined using core shell C-18, 5 µm stationary phases. The separation of carbapenems based on core shell C-18, 5 µm stationary phases is valuable due to short time of analysis, long vitality of column and possibility to work on classical HPLC systems (dedicated for max pressure 400 bars).
References
Sehrawat R, Maithani M, Singh R (2010) Regulatory aspects in development of stability-indicating methods: a review. Chromatographia 72(1–2):1–6
Mendez A, Chagastelles P, Palma E, Nardi N, Schapoval E (2008) Thermal and alkaline stability of meropenem: degradation products and cytotoxicity. Int J Pharm 350:95–102
Cielecka-Piontek J, Zalewski P, Barszcz B, Lewandowska K, Paczkowska M (2013) Stress degradation studies of tebipenem and a validated stability-indicating LC method. Chromatographia 76(7–8):381–386
Cielecka-Piontek J, Zajac M, Jelińska A (2008) A comparison of the stability of ertapenem and meropenem in pharmaceutical preparations in solid state. J Pharm Biomed Anal 46(1):52–57
Matsuda T, Ikawa K, Ikeda K, Morikawa N, Tsumura R, Shibukawa M, Iida K, Kurisu K (2009) LC method for the determination of meropenem in cerebrospinal fluid: application to therapeutic drug monitoring. Chromatographia 69(9–10):1031–1034
Kipper K, Anier K, Leito I, Karjagin J, Oselin K, Herodes K (2009) Rapid determination of meropenem in biological fluids by LC: comparison of various methods for sample preparation and investigation of meropenem stability. Chromatographia 70(9–10):1423–1427
Cielecka-Piontek J, Krause A, Zalewski P, Lunzer A, Jelińska A (2012) Development and validation of a stability-indicating LC–UV method for the determination of doripenem and biapenem in pharmaceutical dosage forms. Acta Chrom 24(2):207–219
Reddy PRM, Prathyusha S, Shanmugasundaram P, Naidu PY, Hanumanaraju GV, Karthikeyan R (2014) Stability indicating ultra-performance liquid chromatography method for the simultaneous estimation of doripenem and its degradation products. J Liq Chrom Relat Tech 37(3):298–310
Carlier M, Stove V, Roberts JA, Van de Velde E, De Waele JJ, Verstraete AG (2012) Quantification of seven β-lactam antibiotics and two β-lactamase inhibitors in human plasma using a validated UPLC-MS/MS method. Int J Antimicrob Agents 40(5):416–422
Wang L, Yuan-Qian Li Y-Q (2009) Simultaneous determination of ten antibiotic residues in milk by UPLC. Chromatographia 70(1–2):253–258
Cielecka-Piontek J, Zalewski P, Jelińska A, Garbacki P (2013) UHPLC: the greening face of liquid chromatography. Chromatographia 76(21–22):1429–1437
Gumustas M, Kurbanoglu S, Uslu B, Ozkan SA (2013) UPLC versus HPLC on drug analysis: advantageous applications and their validation parameters. Chromatographia 76(21–22):1365–1427
Acknowledgments
This study was supported by a grant from the State Committee for Scientific Research, Poland (No. N N405 683040).
Author information
Authors and Affiliations
Corresponding author
Additional information
Published in the topical collection Advances in Chromatography and Electrophoresis & Chiranal 2014 with guest editor Jan Petr.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
About this article
Cite this article
Zalewski, P., Talaczyńska, A., Korban, P. et al. An Approach to Transfer Methods from HPLC to UHPLC Techniques in Some Carbapenems. Chromatographia 77, 1483–1487 (2014). https://doi.org/10.1007/s10337-014-2725-6
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10337-014-2725-6