The winner of the Lead quantitation challenge (published in volume 411 issue 9) is: Sami T. Tuomivaara, University of California, San Francisco, CA 94107, USA. The award entitles the winner to select a Springer book of her choice up to a value of €100,-. Our Congratulations!
Mass spectrometry has reached a broad diffusion as a diagnostic tool in many branches of life sciences and analytical chemistry. In the field of elemental analysis, inductively coupled plasma mass spectrometry (ICPMS) has become the standard technique for performing quantitation of elements at the ultratrace level. In a typical analysis, the ICPMS response of standard solutions is first acquired and plotted against concentration. Then, the calibration curve is obtained by linear fitting and is used to calculate analyte concentration in the samples. In mass spectrometric analysis, one must bear in mind that the detector responds to the isotopes of an element and not to the element itself. For example, in the case of lead, the analyst can choose to acquire m/z of 204 (204Pb), 206 (206Pb), 207 (207Pb), or 208 (208Pb). Usually, quantitation is obtained on m/z 208, which corresponds to the most abundant isotope in the mass spectrum of lead (Fig. 1).
In the Lead quantitation challenge [2], we consider a situation where quantitation of lead is performed by two laboratories on a water sample. In both cases, signals corresponding to 206Pb, 207Pb, and 208Pb were calculated and used for quantitation. Surprisingly, large discrepancies were obtained between and within laboratory when quantitation was performed on the basis of different isotopes.
For the interpretation of this result, we must consider that the abundance of lead isotopes can vary in nature by more than 20%. For a quantitation model that does not take this effect into account, large biases are therefore expected. In this case, a single isotope is not sufficient to represent the element and a comparison between sample and standard cannot be made relying uniquely on a single isotope without information on its abundance in the material.
In this challenge, the data provided [2] were simulated starting from the isotopic compositions listed in Table 1. For the water sample, the lead isotopic composition was consistent with the average values reported by Millot and Négrel in French groundwater [3]. The two laboratories used primary standard solutions modeled after the isotopic composition of two international standards: the ERM-EB400 [4] and the HIPB-1 CRM [5]. As shown in Table 1, the lead isotopic composition for these materials is quite different and such discrepancies must be considered for data analysis.
Laboratory 1 normalized results:
Laboratory 2 normalized results:
When the results are normalized by taking into account the isotopic composition of lead, the biases disappear and a general agreement within and between laboratory is observed.
In mass spectrometry analysis, this correction is mostly overlooked because, for the majority of elements, the differences between isotopic composition of samples and standards are too small to be noticed. However, for metrological applications, this assumption should be verified in each case.
The exercise proposed for this challenge constitutes a simple scenario where the samples are compared to the primary standards with a linear external calibration curve. Further investigations are needed to evaluate the effects of variable isotopic composition when more complex quantitation models, such as standard addition [6] or isotope dilution [7], are employed.
References
Meija J, Coplen TB, Berglund M, Brand WA, Biėvre PD, Gröning M, Holden NE, Irrgeher J, Loss RD, Walczyk T, Prohaska T. Isotopic compositions of the elements 2013 (IUPAC Technical Report). Pure Appl Chem 2016;88:293–306. https://doi.org/10.1515/pac-2015-0503.
Pagliano E. Lead quantitation challenge. Anal Bioanal Chem 2019;411:1675–1676. https://doi.org/10.1007/s00216-019-01592-w.
Millot R, Négrel P. Lead isotope systematics in groundwater: implications for source tracing in different aquifer types. Procedia Earth Planet Sci 2015;13:7–10. https://doi.org/10.1016/j.proeps.2015.07.002.
Vogl J, Yim YH, Lee KS, Goenaga-Infante H, Malinovskiy D, Hill S, Ren T, Wang J, Vocke RD, Murphy KE, Nonose N, Rienitz O, Noordmann J. Certification of ERM- EB400, the first matrix reference material for lead isotope amount ratios, and ERM-AE142, a lead solution providing a lead isotopic composition at the edge of natural variation. Geostand Geoanalytical Res 2019;43:23–37. https://doi.org/10.1111/ggr.12253.
Tong S, Meija J, Zhou L, Methven B, Mester Z, Yang L. High-precision measurements of the isotopic composition of common lead using MC-ICPMS: comparison of calibration strategies based on full gravimetric isotope mixture and regression models. Anal Chem 2019;91:4164–4171. https://doi.org/10.1021/acs.analchem.9b00020.
Meija J, Pagliano E, Mester Z. Coordinate swapping in standard addition graphs for analytical chemistry: a simplified path for uncertainty calculation in linear and nonlinear plots. Anal Chem 2014;86:8563–8567. https://doi.org/10.1021/ac5014749.
Pagliano E, Mester Z, Meija J. Reduction of measurement uncertainty by experimental design in high-order (double, triple, and quadruple) isotope dilution mass spectrometry: application to GC-MS measurement of bromide. Anal Bioanal Chem 2013;405:2879–2887. https://doi.org/10.1007/s00216-013-6724-5.
Author information
Authors and Affiliations
Corresponding author
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This article is the solution to the Analytical Challenge to be found at https://doi.org/10.1007/s00216-019-01592-w
Rights and permissions
About this article
Cite this article
Pagliano, E. Solution to lead quantitation challenge. Anal Bioanal Chem 411, 6533–6534 (2019). https://doi.org/10.1007/s00216-019-02056-x
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00216-019-02056-x