Abstract
Introduction and hypothesis
Objective of this study was to develop an MRI-based finite element model and simulate a childbirth considering the fetal head position in a persistent occiput posterior position.
Methods
The model involves the pelvis, fetal head and soft tissues including the levator ani and obturator muscles simulated by the hyperelastic nonlinear Ogden material model. The uniaxial test was measured using pig samples of the levator to determine the material constants. Vaginal deliveries considering two positions of the fetal head were simulated: persistent occiput posterior position and uncomplicated occiput anterior position. The von Mises stress distribution was analyzed.
Results
The material constants of the hyperelastic Ogden model were measured for the samples of pig levator ani. The mean values of Ogden parameters were calculated as: μ1 = 8.2 ± 8.9 GPa; μ2 = 21.6 ± 17.3 GPa; α1 = 0.1803 ± 0.1299; α2 = 15.112 ± 3.1704. The results show the significant increase of the von Mises stress in the levator muscle for the case of a persistent occiput posterior position. For the optimal head position, the maximum stress was found in the anteromedial levator portion at station +8 (mean: 44.53 MPa). For the persistent occiput posterior position, the maximum was detected in the distal posteromedial levator portion at station +6 (mean: 120.28 MPa).
Conclusions
The fetal head position during vaginal delivery significantly affects the stress distribution in the levator muscle. Considering the persistent occiput posterior position, the stress increases evenly 3.6 times compared with the optimal head position.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
In the past decades, the concept of pelvic floor trauma was attributed largely to perineal, vaginal and anal sphincter injuries. In recent years with advances in magnetic resonance imaging (MRI) and four-dimensional sonography, it has also become evident that levator ani muscle (LAM) injuries represent an important part of pelvic floor trauma [1, 2]. LAM injuries occur in 13–36% of women who have vaginal delivery (VD), and there is conclusive evidence of a connection between these injuries and pelvic floor dysfunction (PFD) [3, 4]. During VD, the LAM undergoes extremely large deformations to allow the passage of the fetal head [5]. These deformations can lead to injuries such as muscle tearing and striated muscle atrophy, owing to pudendal denervation [6]. This trauma usually causes lifelong complications [7].
A persistent occiput posterior (OP) position in the second stage of delivery carries an increased risk of labor complications, obstetric interventions, severe maternal perineal lacerations and anal sphincter injury [8, 9]. This condition is found in up to 5% of deliveries [10] and occurs significantly more often in first labors.
Several virtual models of VD have already been developed. Most of them are focused on the second stage of labor, starting from full dilatation of the cervix to the birth of the fetus [11,12,13,14]. Although these models provide satisfactory insights, they are strongly limited by constitutive data or boundary conditions representing the real anatomy and physiology. Material properties are usually derived from cadaveric LAM uniaxial or biaxial tests [15]. To the authors' knowledge, there are no studies describing the material properties of non-cadaveric LAM. The objective of this study was to develop a sophisticated MRI-based finite element model of the female pelvic floor to simulate a child birth for fetal head position in persistent OP and analyze the stress distribution in the LAM with respect to the cardinal movements of labor.
Materials and methods
The study was approved by the local ethics committee of the authors' institution. An MRI-based three-dimensional computer model of the bony pelvis, pelvic floor muscles, pelvic floor organs and fetal head was created. All supporting structures were replaced by boundary conditions. The geometry of these structures was based on live-subject MRI data. Female characteristics were 25 years old, BMI 21.9 kg/m2, nulliparous, Caucasian, no previous vaginal delivery, normal POP-Q points, and absence of PFD symptoms. The neonate was 1 day old, after uncomplicated VD at term, with neurological indications for an MRI brain scan. The volunteer/legal representative gave written consent. Our method of obtaining the three-dimensional computer model and the imaging protocol have been published previously [16]. The objective of this study was to simulate changes in the LAM during vaginal delivery with respect to the cardinal movements of labor.
The original MRI-based 3D finite element model of the female pelvic floor already developed by these authors was improved and used for further applications. Compared with the original model, the following changes were implemented: (1) the real, not simplified geometry of the obturator internus muscle (OIm) was involved, (2) the LAM was clearly split into three individual parts, the iliococcygeal muscle (ICm), puborectal muscle (PRm), and pubovisceral muscle (PVm), based on in-vivo MRI, (3) the real passive biomechanical properties of pelvic floor muscles obtained from measurements were newly used, (4) the boundary conditions representing the pelvic floor-supporting apparatus were particularized in detail, and (5) OP fetal head positions were considered.
Model geometry and finite element mesh (FEM)
The initial model geometry was reconstructed from in vivo scanned MRI using a free semiautomatic software (3D Slicer, 3.0; BWH, Boston, MA, USA). The outlines of relevant structures were digitized from these consecutive axial MRI scans. The resulting geometry and mesh were created in HyperMesh commercial software (11.0; Altair, MI, USA). Rigid parts, such as the female pelvis and fetal head, were constructed with 2D triangular mesh including 119,373 elements in summary. Deformable parts, such as muscles, were modeled by 3D tetrahedral mesh consisting of 527,164 elements.
The accuracy and efficiency of the finite element simulations considered in the presented work are highly predisposed to the quality of the finite element mesh [17]. Thus, the process of element quality control was not neglected. The tetrahedral elements (four-node version) were chosen to discretize the model geometry. The aspect ratio (AR), measured as the ratio of the longest edge length divided by the minimum altitude of the smallest side, represents a simple parameter quantifying the element shapes. Prior models recommended that the percentage of the tetrahedral element ARs > 3 should be < 5% to produce the smallest errors [18, 19]. In this work, 1.19% of element ARs > 3 was considered. In addition, respecting the published data [20], only dihedral angles between 30°and 120° were accepted in this study. The size of the element edges was also limited; the range of 1–3 mm was considered. Finally, the element Jacobians were also checked. This parameter represents the determinant of the Jacobian matrix containing information regarding the volume, shape and orientation of the element. The following criteria were strictly used: (1) a positive value [19]; (2) > 0.2 in magnitude [18, 21]; (3) < 5% of all Jacobians below a magnitude of 0.7 [22].
Material properties
During vaginal delivery, the LAM undergoes extremely large deformation; therefore, the hyperelastic nonlinear Ogden material model was used [23]. The Ogden material model is a mathematical model developed by Raymond Ogden in 1972. This model describes hyperelastic models having non-linear stress-strain behavior. The model is suitable for complex materials such as biological tissues because its material characteristic is very similar. In our work, the human cadaveric muscle samples were replaced by animal tissues removed in vivo. The material parameters were based on experimental measurements of porcine muscle samples.
Experimental procedure—porcine LAM specimen
The experiments conducted were in accordance with the animal welfare regulations and guidelines for the country in which the experiments were performed. Twenty LAM samples from ten female pigs (age: 12–14 weeks, weight: 33–45 kg, nulliparous) were obtained from practice surgical operations designed for medical students. The LAM samples were surgically removed during general anesthesia. Two slabs of LAM were obtained from each pig (from the right and left side). After that, the LAM specimens were individually stored in glass laboratory jars and immediately deep frozen (−20 °C) until further processing. The effects of careful deep freezing on the hyperelastic and tensile properties are negligible, as reported in the literature [24].
Experimental measurement—passive biomechanical LAM properties to determine the stress-strain curve
The frozen LAM slabs were kept at room temperature (22 °C) to thaw out for 3 h before mechanical measurement following the recommended procedure [24]. The muscles were cleared of fascia and cut into cuboid-shaped samples suitable for the measurement device (about 6 mm wide, 4 mm thick and 10 mm long), always in the direction of the muscle fibers. The fibers were always identified anteriorly to the ischial spine, posteriorly to the pubic rami, medially to the arcus tendineus musculi levatoris ani (ATML) and laterally to the vaginal wall. The LAM specimens were clamped in a mechanical test system in the direction of muscle fibers, and a uniaxial mechanical test was performed. The Zwick/Roell Z050 traction machine (Zwick/Roell, Ulm, Germany) was used for measurement. The exact width and thickness of the initial unloaded samples were measured by a digital caliper to obtain the initial cross-sectional areas. The lengths of unloaded samples were given by the initial distance between the grips of the measuring apparatus. Thus, the original length of each specimen was found to be 5 or 10 mm. The measuring apparatus does not allow keeping the samples in liquid during the measurement. Therefore, they were sprayed with 0.9% NaCl (sodium chloride) to prevent dehydration. Each sample underwent preconditioning consisting of 20 cycles up to 15% of its original length. The uniaxial extension was performed until the muscle ruptured. The loading velocity was 6 mm/min, as reported in [25]. The tensile force and sample elongations were recorded.
Experimental procedure—Ogden material identification
The simple homogeneous deformations were considered during uniaxial tension: xi = λi Xi, where i = 1, 2, 3 and xi are the corresponding coordinates after deformations, λi are constants referred to as the principal stretches of deformations, and Xi are coordinates identifying material particles in some unstressed configuration. The deformation gradient, F, in terms of principal stretches is then: F = diag [λ1, λ2, λ3]. For an incompressible material, the principal stretches satisfy the constant: λ1 λ2 λ3 = 1. Thus, for the Ogden hyperelastic material, the strain function is given by the following formula:
where μj [GPa] and αj [−] are the unknown Ogden parameters and N is the number of terms in the Ogden series; for the purpose of this work, N = 2. Assuming uniaxial tension, the unknown variables are then: λ2 = λ3 = λ1–1/2, J = 1. Thus, the parameter K is absolutely not significant for this purpose. Moreover, it is assumed that the muscular tissue exhibits an incompressible behavior. The measured tensile force and elongation were used to calculate the stress-strain curve for each muscle sample. The stress was defined as the actual force divided by the initial cross section. The strain was defined as the actual specimen elongation divided by the initial specimen length. The final stress-strain curves were used to identify unknown parameters in the specific strain-energy function by means of the least squares technique. For the case of uniaxial tension, the nominal stress-stretch relation for Ogden material can be written as: \( \mathrm{P}=\frac{\updelta \mathrm{W}}{\updelta \uplambda} \). The method consists of minimizing the stress error and thus:
where Pitest is the measured nominal stress for the nominal strain measure, Piderived is the stress obtained from the strain energy function, and ND is the number of data points. In this study, the function lsqcurvefit in the Optimization Toolbox of MATLAB (R2013a; The MathWorks, Inc., Natick, MA, USA) was used. The lower and upper limits were added to ensure the following inequality that must be satisfied for all Ogden parameters: \( \sum \limits_{\mathrm{i}=1}^{\mathrm{N}}{\upmu}_{\mathrm{i}}{\upalpha}_{\mathrm{i}}>0 \) [16].
The density of general mammalian skeletal muscle tissues is approximately 1.06 kg/l [26]; the value of Poisson’s ratio is 0.499, as usual. The friction between the fetal head and female pelvic floor muscles is 0.1 assuming that the contact is well lubricated during the vaginal delivery.
Initial and boundary conditions
The initial (ICs) and boundary (BCs) conditions were considered to simulate the real behavior of pelvic structures during VD. The ICs reflect the initial positions and rotations of all observed components. The BCs represent the surrounding support of all model structures. The pelvis was modeled by a rigid body fixed for all degrees of freedom in 3D space and represents the mechanical frame for soft tissues. In addition, it forms the anatomical pelvis axis limiting the head trajectory during labor. Passive connective tissue in the model followed the outlines of relevant structures. The obturator membrane was replaced by external supporting forces. The OIm also provides support for the LAM and was fixed to the bony pelvis (ischiopubic ramus) and at the points where the OIm leaves the internal pelvic floor in the direction of muscle insertion—the medial aspect of the greater trochanter. The ATML traverses the anteromedial aspect of the OIm. To create the LAM model, three sections of levator ani muscles, based on existing descriptions, were included [27]. The ICm originates from the ATML. These nodes were fixed to the OIm. Laterally, the ICm is attached to the posterior ilia and dorsal to the sacrum; this part of the muscle is also translationally and rotationally fixed. Nodes along the upper cranial third of the ICm were limited to move only in the plane defined by the sacrum and posterior ilia in the interior and lateral directions. This condition represents the LAM support provided by ligaments and tendons attached to the inner pelvic walls. The PVm and PRm originate from the fibrous enthesis on the dorsal surface of the pubic bone. These muscle parts are fixed in the space. The upper arc of the PVm is connected to the ICm; the lower part of the PRm is then connected to the PVm. All other nodes of the presented model were unconstrained.
The fetal head was also replaced by a rigid body without any possibility to be deformed [16]. For station, following obstetric convention, the pelvis was divided above and below the ischial spine into levels (expressed in centimeters). There is a difference of head descent when the occiput is posterior. Fetuses with occiput anterior (OA) position follow the curved path of the birth canal, and the fetal head undergoes positional changes through the birth canal (cardinal movements). In this study, the fetal head entered the pelvic inlet in the left occiput posterior (LOP) position, and the internal rotation moved the occiput toward the os sacrum. At the level of the ischial spine (station 0), the internal rotation was completed and the occiput could remain in the direct occiput posterior position. After the ischial spine has been passed and the forehead has completely passed the pubic bone, flexion of the head becomes possible, and it abruptly twists up toward the outlet (Fig. 1). A biomechanical description of the fetal OP head trajectory in the birth canal is shown in Fig. 2.
Stress distribution in the MLA
A vaginal delivery scaled in seconds was simulated for the head position in OP. At the beginning of the simulation, the LAM remained at rest without any external load. The distribution of von Mises stress generated in the LAM during vaginal delivery was analyzed using the FEM method and the Virtual Performance Solution commercial software (VPS 9.0; ESI Group, Paris, France). The von Mises stress allows any arbitrary three-dimensional stress to be represented as a single positive stress value. All three principal stresses could be computed and shown.
Results
The calculated Ogden parameters, μ1, μ2, α1 and α2, computed for each pig’s LAM specimen are summarized in Table 1. Two members of the Ogden series were considered. The mean values ± standard deviations of the Ogden parameters were calculated: μ1 = 8.9 ± 8.2 GPa; μ2 = 21.6 ± 17.3 GPa; α1 = 0.1803 ± 0.1299; α2 = 15.112 ± 3.1704. The von Mises stress distribution in the LAM is shown in Table 2. For comparison, the table also contains the von Mises stress distribution in the LAM during vaginal delivery with respect to the cardinal movements of labor where the fetal head is in the OA position. The results are presented respecting the fetal head descent, considering the range from −1 up to +8. The fetal head initially distends the ICm portion of the LAM at station −1 and 0, respecting the OP and OA fetal head position. The mean stress values at this station were 1.15 MPa and 0.09 MPa, respectively. The PVm complex portion was distended at station +2 and +1 for the OP and OA head position. The mean stress values in these stations were 2.56 MPa and 0.01 MPa. Finally, the PRm portion was also distended at station +2 and + 1. The stress values were 120.28 MPa and 32.55 MPa, respectively. The maximal stress for the OP head position was found already at station +6 in the PRm muscle part. The mean value was 120.28 MPa. However, the maximum for the OA position was at station +8 in the PVm part. The value was 44.53 MPa. The color-coded versions of the von Mises stress distribution in the LAM during VD considering the LOP initial fetal head position is shown for four head positions (Fig. 3). In this figure, the MLA elongation in the caudal direction as a result of distention by the fetal head is also depicted. Considering both initial head positions, the MLA was elongated by nearly 1.5 times versus its initial resting position.
Discussion
Failure of the occiput to spontaneously rotate to the anterior position has been associated with increased maternal and neonatal morbidity. Its clinical significance has long been a subject of controversy among obstetricians. Despite several studies on this topic, the optimal method of decreasing the maternal pelvic floor and perineal trauma remains uncertain.
Our study provides original data on LAM behavior during the active second stage of labor in fetuses with a persistent OP position. It confirmed the assumption of direct dependence of the stress distribution in the LAM on the fetal head position. This work shows the significant stress increase for the case of the LOP position compared with the optimal OA position. The mechanism of VD of OP fetuses requires an increased distension of the LAM to accommodate a greater degree of downward descent of the fetal head until the forehead has completely passed the pubic bone. In the ICm, the stress values were comparable for both positions. Nevertheless, the von Mises stress generated in the PVm part was more than two times higher in the OP than OA position. In addition, the stress values in the PRm even increased 3.6 times. Nevertheless, the muscle elongation in the caudal direction was also comparable for both head positions.
In the presented model, the largest LAM stretch ratio was 1.5 in the inferior direction. Compared with the range of 1.62–3.76, based on a statistical study of 227 women [28], the LAM seems to be slightly stiffer. This could be caused by the pig muscle specimens used for material constant identification. Nevertheless, the model was well validated comparing the results of the OA head position with the published data. The maximal mean stress in the ICm was 15.09 MPa, which corresponds to the results in the literature [16, 29]. The maximal stress values in the PVm complex were generated during fetal head extension. The obtained value of 44.53 MPa is also comparable to the literature [16]. Finally, the maximum value in the posteromedial inner area of the PRm was 32.55 MPa at station +4, corresponding with data from [13]. The authors were unable to find any literature describing the simulation of the OP FEM model; thus, we were unable to validate these results. Nevertheless, the FEM model was validated for the LOA head position. Therefore, the outputs for the OP position can also be considered as correct. This increase in LAM stress distribution in OP may be associated with a variety of factors. Incomplete flexion of the fetal head increases the presenting diameter and alters the mechanics of labor. Fetuses with OA follow the curved path of the birth canal, whereas fetuses with OP show a different head trajectory (Fig. 4). This different trajectory has been sonographically witnessed among fetuses with persistent OP positions in the active second stage [30]. This confirms our observations that the mechanism of VD in OP fetuses requires an increased LAM distension to accommodate a greater degree of downward fetal head descent.
Concerning the material modeling, the presented model has advantages and also some limitations. The first limitation of this model is that the parameters of the pig LAM were measured instead of human tissues. The porcine LAM has the same structure, function and attachments as the human. Nevertheless, porcine bodies have a different anatomy. They move on four legs; moreover, their fetuses have different weights and sizes in the ratio of female parameters compared with the human body. Therefore, the external forces acting on the pig LAM are dissimilar to those of humans, and the mechanical properties could be slightly different. The second limitation is that we did not measure in vivo muscles. The results of measurements of mechanical properties could be influenced by dehydration of the tissue. However, this potential source of bias was minimalized by the short duration of the experiment (< 10 min) and by spraying the tissue with a saline solution. The third limitation is that the muscle samples were mechanically dissected from the porcine pelvis and thus the tissue could be slightly deformed involving some volume changes. The next limitation is that the measurement and model were simplified. The LAM is an anisotropic biological tissue. Therefore, the use of uniaxial testing together with the isotropic constitutive model is not fully sufficient to describe their mechanical response. For our purpose, this analysis is acceptable. Nevertheless, to characterize the anisotropic properties, a cross-fiber experiment should be performed. Another limitation is that the head was driven at a given trajectory with a homogeneous progression. Linear progress is used as a simplification of real labor in our numerical simulation. We plan to consider more realistic motion of the fetal head (progress and standstill) in our future studies.
Unlike other studies in which the head model is oversimplified as a sphere, we used the geometry of an actual neonatal head based on MRI. The effects of fetal head molding were not considered. However, the use of a deformable fetal head leads to an almost 20% reduction of the reaction forces on the pelvic floor muscles compared with a rigid head [31]. Nevertheless, to the authors' knowledge, there is no data set in the literature describing head molding during VD considering the OP head position.
In the future, this model will be improved in a few steps: a biaxial tensile test will be performed to obtain the material constants to allow the simulation of non-homogeneous behavior of muscle tissue; a molding head for the LOA position will be used.
Our results have clearly shown that a persistent OP position is a significant risk factor for LAM stress injury. Recognition of this greater susceptibility to severe birth-related LAM trauma at OP delivery should help reduce its occurrence.
Abbreviations
- MRI:
-
Magnetic resonance imaging
- LAM:
-
Levator ani muscle
- VD:
-
Vaginal delivery
- PFD:
-
Pelvic floor dysfunction
- OP:
-
Occiput posterior
- OIm:
-
Obturator internus muscle
- ICm:
-
Iliococcygeal muscle
- PRm:
-
Puborectal muscle
- PVm:
-
Pubovisceral muscle
- FEM:
-
Finite element mesh
- ICs:
-
Initial conditions
- BCs:
-
Boundary conditions
- ATML:
-
Arcus tendineus musculi levatoris ani
- OA:
-
Occiput anterior
- LOP:
-
Left occiput posterior
- AR:
-
Aspect ratio
- GPa:
-
Gigapascal
- MPa:
-
Megapascal
References
De Lancey JO, Kearney R, Chou Q, Speights S, Binno S. The appearance of levator ani muscle abnormalities in magnetic resonance images after vaginal delivery. Obstet Gynecol. 2003;101:46–53.
Dietz HP, Janzarone V. Levator trauma after vaginal delivery. Obstet Gynecol. 2005;106:707–12.
Schwertner-Tiepelmann N, Thakar R, Sultan AH, Tunn R. Obstetric levator ani muscle injuries: current status. Ultrasound Obstet Gynecol. 2012;39:372–83.
Handa VL, Blomquist JL, Roem J, Muñoz A, Dietz HP. Levator morphology and strength after obstetric avulsion of the levator ani muscle. Female Pelvic Med Reconstr Surg. 2018. https://doi.org/10.1097/SPV.0000000000000641.
Jansova M, Kalis V, Rusavy Z, Zemcik R, Lobovsky L, Laine K. Modelling manual perineal protection during vaginal delivery. Int Urogynecol J. 2014;25:65–71.
Lien KC, Morgan DM, DeLancey JO. Pudendal nerve stretch during vaginal birth: a 3D computer simulation. Am J Obstet Gynecol. 2005;192:1669–76.
Mant J, Painter R, Vessey M. Epidemiology of genital prolapse: observations from the Oxford family planning association study. Brit J Obstet Gynecol. 1997;104:579–85.
Cheng YW, Shaffer BL, Caughey AB. The association between persistent occiput posterior position and neonatal outcomes. Obstet Gynecol. 2006;107:837–44.
Akmal S, Tsoi E, Howard R, Osei E, Nicolaides KH. Investigation of occiput posterior delivery by intrapartum sonography. Ultrasound Obstet Gynecol. 2004;24:425–8.
Simkin P. The fetal occiput posterior position: state of the science and a new perspective. Birth. 2010;37:61–7.
Li X, Kruger JA, Nash MP, Nielsen PM. Anisotropic effects of the LA muscle during childbirth. Biomech Model Mechanobiol. 2011;10:485–94.
Ashton-Miller JA, DeLancey OL. On the biomechanics of vaginal birth and common sequelae. Annu Rev Biomed Eng. 2009;11:163–76.
Hoyte L, Damaser MS, Warfield SK, Chukkapalli G, Majumdar A, Choi DJ. Quantity and distribution of LA stretch during simulated vaginal childbirth. Am J Obstet Gynecol. 2008;199:198.e1–5.
Lien KC, DeLancey JO, Ashton-Miller JA. Biomechanical analysis of the efficacy of patterns of maternal effort on second-stage progress. Obstet Gynecol. 2009;113:873–80.
Nagle AS, Barker A, Kleeman SD, Haridas B. Passive biomechanical properties of human cadaveric levator ani muscle at low strains. J Biomech. 2014;47:583–6.
Krofta L, Havelkova L, Urbankova I, Krcmar M, Hyncik L, Feyereisl J. Finite element model focused on stress distribution in the levator ani muscle during vaginal delivery. Int Urogynecol J. 2017;28:275–84.
Burkhart TA, Andrews DM, Dunning CE. Finite element modeling mesh quality, energy balance and validation methods: a review with recommendations associated with the modeling of bone tissue. J Biomech. 2013;46:1477–88.
Tsukerman I, Plaks A. Comparison of accuracy criteria for approximation of conservative fields in tetrahedra. IEEE Trans Magn. 1998;34:3252–5.
Quenneville CE, Dunning CE. Development of a finite element model of the tibia for short-duration high-force axial impact loading. CMBBE. 2011;14:205–12.
Klinger BM, Shewchuck JR. Aggressive tetrahedral mesh improvement. 16th International Meshing Roundtable. Springer-Verlag; 2007. p. 3–23.
Li ZH, Reed MP, Rupp JD, Hoff CN, Zhang J, Cheng B. Development, validation and application of a parametric pediatric head finite element model for impact simulations. Ann. Biomed Eng. 2011;39:2984–97.
Ray MH, Mongiardini M, Plaxico CA, Anghileri M. Recommended procedures for verification and validation of computer simulations used for roadside safety applications. National Cooperative Highway Research Program (NCHRP) Project 22–24. Final Report, available at. 2010http://onlinepubs.trb.org/onlinepubs/nchrp/nchrp_w179.pdf. Accessed 15 January 2012
Ogden RW, Saxxomandi G, Sgura I. Fitting hyperelastic models to experimental data. Comput Mech. 2004;34:484–502.
Moon DK, Woo SL-Y, Takakura Y, Gabriel MT, Abramowitch SD. The effects of refreezing on the viscoelastic and tensile properties of ligaments. J Biomech. 2006;39:1153–7.
Jing D. Experimental and theoretical biomechanical analysis of the second stage of labor, PhD Thesis. University of Michigan; 2010.
Hsu Y, Chen L, Huebner M, Asthon-Miller JA, DeLancey JO. Quantification of levator ani cross-sectional area differences between women with and those without prolapse. Obstet Gynecol. 2006;108:879–83.
Kearney R, Sawhney R, DeLancey JO. Levator ani muscle anatomy evaluated by origin-insertion pairs. Obstet Gynecol. 2004;104:168–73.
Svabik K, Shek KL, Dietz HP. How much does the levator hiatus have to stretch during childbirth? BJOG Int J Obstet Gynaecol. 2009;116:1657–62.
Jing D, Ashton-Miller JA, DeLancey JO. A subject-specific anisotropic vicso-hyperelastic finite element model of female pelvic floor stress and strain during the second stage of labor. J Biomech. 2012;45:445–60.
Ghi T, Maroni E, Youssef A, Morselli-Labate AM, Paccapello A, Montaguti A, et al. Sonographic pattern of fetal head descent: relationship with duration of active second stage of labor and occiput position at delivery. Ultrasound Obstet Gynecol. 2014;44:82–9.
Silva MET, Oliveira DA, Roza TH, Brandao S, Parente MPL, Mascarenhas T, et al. Study on the influence of the fetus head molding on the biomechanical behavior of the pelvic floor muscles, during vaginal delivery. J Biomech. 2015;48:1600–5.
Funding
This study was supported by the Charles University PROGRES projects; grant LO1506 of the Czech Ministry of Education, Youth and Sports under program NPU I; the European Regional Development Fund Project "Application of Modern Technologies in Medicine and Industry” (no. CZ.02.1.01/0.0/0.0/17_048/0007280); project no. 182 "Obstetrics 2.0—Virtual models for the prevention of injuries during childbirth” realized within the frame of the INTERREG V-A Program: Cross-border cooperation between the Czech Republic and the Federal State of Germany, Bavaria, Aim European Cross-border Cooperation 2014–2020. The project is supported financially by the European Regional Development Fund (85% of the costs) and the state budget of the Czech Republic (5%).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflicts of interest
None.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
This work was presented at AUGS PFD Week, October 9–13, 2018, Chicago, IL, USA
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Havelková, L., Krofta, L., Kochová, P. et al. Persistent occiput posterior position and stress distribution in levator ani muscle during vaginal delivery computed by a finite element model. Int Urogynecol J 31, 1315–1324 (2020). https://doi.org/10.1007/s00192-019-03997-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00192-019-03997-8