Abstract
Determination of aluminium in serum of patients on chronic intermittent haemodialysis is of paramount importance in the prevention or early diagnosis of aluminium intoxication. We present a new method based on flameless atomic absorption spectroscopy, in which the serum matrix is destroyed by oxygen. A comparison has been made between the described method and another procedure which is used in the Laboratory of Toxicology in Gent, with favourable results. In addition, a method is presented for the determination of aluminium in hair, in which special attention has been paid to the cleaning of the hair samples prior to destruction. As yet it cannot be concluded whether aluminium concentrations in hair give a better representation of the body burden than serum levels do.
Similar content being viewed by others
References
Alfrey, A.C., G.R. Le Gendre andW.D. Kaehny (1976)N. Engl. J. Med. 294, 184–188.
Alfrey, A.C., A. Hegg andP. Craswell (1980)Am. J. Clin. Nutr. 33, 1509–1516.
Davison, A.M., G.S. Walker, H. Oli andA.M. Lewins (1982)Lancet II, 785–787.
Fleming, L.W., W.K. Stewart, G.S. Fell andD.J. Halls (1982)Clin. Nephrol. 17, 222–227.
Flendrig, J.A., H. Kruis andH.A. Das (1976)Lancet I, 1235.
Gardiner, P.E., J.M. Ottaway, G.S. Fell andD.J. Halls (1981)Anal. Chim. Acta 128, 57–66.
Gorsky, J.E., andA.A. Dietz (1978)Clin. Chem. 24, 1485–1490.
Gorsky, J.E., A.A. Dietz, H. Spencer andD. Osis (1979)Clin. Chem. 25, 1739–1743.
Graf, H., H.K. Stummvoll, V. Meisinger, J. Kovarik, A. Wolf andW.F. Pinggera (1981)Kidney Int. 19, 587–592.
Guillard, O., A. Piriou, P. Mura andD. Reiss (1982)Clin. Chem. 28, 1714–1715.
Laker, M. (1982)Lancet II, 260–262.
Majelijne, W., P. D'Haese, F. Van Den Vijver, D. Pollet, F.A. De Wolff, G. Verpooten, M.E. De Broe andA. Heyndrickx (1982) Abstracts of the International Meeting of the International association of forensic toxicologists, Sevilla.
Parkinson, I.S., M.K. Ward, T.G. Feest, R.W.P. Fawcett andD.N.S. Kerr (1979)Lancet I, 406–409.
Parkinson, I.S., M.K. Ward andD.N.S. Kerr (1981)J. Clin. Pathol. 34, 1285–1294.
Schmidt, W., andF. Dietl (1981)Fresenius Z. Anal. Chem. 308, 129–132.
Smeyers-Verbeke, J., D. Verbeelen andD.L. Massert (1980)Clin. Chim. Acta 108, 67–73.
Trapp, G.A., andJ.B. Cannon (1981)N. Engl. J. Med. 304, 172.
Van De Vijver, F., J.P. Van Waeleghem, M.E. De Broe, P. D'haese andA. Heyndrickx (1982)Lancet II, 1106.
Yokel, R.A. (1982)Clin. Chem. 28, 662–665.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
De Groot, H.J., De Haas, E.J.M., D'Haese, P. et al. Determination by flameless atomic absorption of aluminium in serum and hair for toxicological monitoring of patients on chronic intermittent haemodialysis. Pharmaceutisch Weekblad Scientific Edition 6, 11–15 (1984). https://doi.org/10.1007/BF01960191