Definition
The Beer-Lambert law for the absorption, A, of light by a sample of concentration C is
where ℓ is the length of the sample through which the light passes and ε is known as the (wavelength dependent) extinction coefficient. If C is measured in mol dm−3 and ℓ is measured in cm, then ε has units of mol−1 dm3 cm−1. If ε and ℓ are known, the absorbance at a given wavelength can then be used to determine concentration. Ideally, one uses a point of maximum absorbance as the absorbance signal will be less affected by small errors in wavelength calibration. However, sometimes this is not possible.
For DNA, one usually uses the absorbance signal at or near 260 nm (wherever the maximum is for a particular DNA). For DNA, the linear relationship between concentration and absorbance breaks down when the absorbance of a 1-cm path length solution exceeds ∼ 1.5–2 absorbance units, presumably due to DNA-DNA interactions. Some DNA extinction coefficients...
This is a preview of subscription content, log in via an institution.
References
Kelly SM, Jess TJ, Price NC (2005) How to study proteins by circular dichroism. Biochim Biophys Acta 1751:119–139
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Section Editor information
Rights and permissions
Copyright information
© 2018 European Biophysical Societies' Association (EBSA)
About this entry
Cite this entry
Rodger, A. (2018). Concentration Determination Using Beer-Lambert Law. In: Roberts, G., Watts, A. (eds) Encyclopedia of Biophysics. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-35943-9_775-1
Download citation
DOI: https://doi.org/10.1007/978-3-642-35943-9_775-1
Received:
Accepted:
Published:
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-35943-9
Online ISBN: 978-3-642-35943-9
eBook Packages: Springer Reference Biomedicine and Life SciencesReference Module Biomedical and Life Sciences