Abstract
We have proposed a new gelatin phantom for evaluating the spatial resolution of Optical Computed Tomography (OCT), MRI and X-ray CT (X-CT) as a gel dosimeter readout device. In addition, MTF evaluation was conducted using the proposed gelatin phantom. Consequently, 0.19 (MRI), 1.07 (X-CT), 0.88 (0-D OCT), and 0.89 (2-D OCT) were obtained as a results of modulation transfer factor at 0.3 spatial frequency (cycle/mm).
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
References
Baldock C., De Deene Y., Doran S., Ibbott G., Jirasek A., Lepage M., McAuley KB., Oldham M., Schreiner LJ.: Polymer gel dosimetry. Phys Med Biol. 55(5):R1-63 (2010). https://doi.org/10.1088/0031-9155/55/5/r01. Epub 2010 Feb 11.
Vandecasteele J., De Deene Y.: Evaluation of radiochromic gel dosimetry and polymer gel dosimetry in a clinical dose verification. Phys Med Biol. 58(18):6241–62 (2013). https://doi.org/10.1088/0031-9155/58/18/6241. Epub 2013 Aug 22.
Watanabe Y., Nakaguchi Y.: 3D evaluation of 3DVH program using BANG3 polymer gel dosimeter. Med Phys. 40(8):082101 (2013).
Maryanski MJ., Schulz RJ., Ibbott GS., Gatenby JC., Xie J., Horton D., Gore JC.: Magnetic resonance imaging of radiation dose distributions using a polymer-gel dosimeter. Phys Med Biol. 39(9):1437–55 (1994).
Maryanski MJ., Gore JC., Kennan RP., Schulz RJ.: NMR relaxation enhancement in gels polymerized and cross-linked by ionizing radiation: a new approach to 3D dosimetry by MRI. Magn Reson Imaging. 11:253–258 (1993).
P. Y. Guo, J. A. Adamovics, M. Oldham: Characterization of a new radiochromic three-dimensional dosimeter. Med Phys. 33(5): 1338–1345 (2006).
Oldham M., Kim L.: Optical-CT gel-dosimetry. II: Optical artifacts and geometrical distortion. Med Phys. 31(5):1093–1104 (2004).
Kawamura H., Takanashi T., Shimada Y., Sato Y., Abe S., Hara S.: Optical computed tomography for polymer gel dosimetry. Igaku Butsuri.; 37(2): 111–116 (2017).
Ichikawa K., Muramatsu Y. Hyoujun X Sen CT Gazou Keisoku, Ohmsha Japan (2009).
Nickoloff EL: Measurement of the PSF for a CT scanner: appropriate wire diameter and pixel size. Phys Med Biol. 33(1): 149–155 (1988).
Shashadhar M. Mohapatra, Jeff D. Turley, John R. Prince, Joseph C. Blechinger, and Don A Wilson: Transfer function measurement and analysis for a magnetic resonance imager. Med Phys. 18: 1114 (1991). https://doi.org/10.1118/1.596622.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2019 Springer Nature Singapore Pte Ltd.
About this paper
Cite this paper
Takanashi, T., Kawamura, H. (2019). Evaluation of Spatial Resolution of MRI, Optical CT and X-Ray CT Using MTF for Gel Dosimeter. In: Lhotska, L., Sukupova, L., Lacković, I., Ibbott, G.S. (eds) World Congress on Medical Physics and Biomedical Engineering 2018. IFMBE Proceedings, vol 68/1. Springer, Singapore. https://doi.org/10.1007/978-981-10-9035-6_6
Download citation
DOI: https://doi.org/10.1007/978-981-10-9035-6_6
Published:
Publisher Name: Springer, Singapore
Print ISBN: 978-981-10-9034-9
Online ISBN: 978-981-10-9035-6
eBook Packages: EngineeringEngineering (R0)