Performance of Systolic Blood Pressure estimation from radial Pulse Arrival Time (PAT) in anesthetized patients

  • Conference paper
  • First Online:
EMBEC & NBC 2017 (EMBEC 2017, NBC 2017)

Abstract

The performance of estimating Systolic Blood Pressure (SBP) in anesthetized patients via Pulse Arrival Time (PAT) techniques was studied with respect to the minimum required time in between two recalibration procedures.

Materials: a clinical trial [NCT02651558] involving 14 patients was conducted measuring PAT from an ECG and an arterial line inserted into the radial artery.

Methods: comparison of BP estimates from PAT measurements against invasive BP values was performed in terms of mean error and standard deviation of the error (AAMI/ANSI/ISO 81060-2), cumulative percentage of readings falling within 5, 10 and 15 mmHg (BHS criteria), and MAD - Mean Absolute Difference (IEEE Std 1708). Two calibration strategies were explored, involving time between recalibration periods ranging from 10 seconds to 8 minutes.

Results: assuming an affine calibration function between PAT and SBP, different slope (Mean Slope: -1.45, CI: -1.64 to -1.27 mmHg/ms) and offset values (Mean Offset: 575, CI: 517 to 633 mmHg) were found in between patients. In addition, given a patient, affine calibration functions at different anesthesia phases also showed to be variable. When assessing agreement in terms of existing international standards it was found that PATbased SBP estimates complied with requirements when time between two calibrations was smaller than 60 seconds.

Conclusions: the use of anesthetic agents compromises the implementation of PAT-based techniques to estimate SBP.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  • Mukkamala et al, Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and Practice. IEEE Trans Biomed Eng. 2015 Aug;62(8):1879-901

    Google Scholar 

  • Chiu et al, Determination of pulse wave velocities with computerized algorithms. Am Heart J. 1991 May;121(5):1460-70

    Google Scholar 

  • Sola, Continuous non-invasive blood pressure estimation. ETHZ Thesis No 20093, 2011.

    Google Scholar 

  • IEEE Std. 1708-2014 - IEEE Standard for Wearable Cuffless Blood Pressure Measuring Devices.

    Google Scholar 

  • ISO, ANSI/AAMI/ISO 81060-2:2013 Non-invasive sphygmo manometers — Part 2: Clinical investigation of automated measurement type

    Google Scholar 

  • ESH, Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults, Blood Press Monit. 2002 Feb;7(1):3-17

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Josep Solà .

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2018 Springer Nature Singapore Pte Ltd.

About this paper

Cite this paper

Solà, J. et al. (2018). Performance of Systolic Blood Pressure estimation from radial Pulse Arrival Time (PAT) in anesthetized patients. In: Eskola, H., Väisänen, O., Viik, J., Hyttinen, J. (eds) EMBEC & NBC 2017. EMBEC NBC 2017 2017. IFMBE Proceedings, vol 65. Springer, Singapore. https://doi.org/10.1007/978-981-10-5122-7_216

Download citation

  • DOI: https://doi.org/10.1007/978-981-10-5122-7_216

  • Published:

  • Publisher Name: Springer, Singapore

  • Print ISBN: 978-981-10-5121-0

  • Online ISBN: 978-981-10-5122-7

  • eBook Packages: EngineeringEngineering (R0)

Publish with us

Policies and ethics