  • Chapter
Ophthalmic Echography

Part of the book series: Documenta Ophthalmologica Proceedings Series ((DOPS,volume 48))

  • 20 Accesses

Abstract

Although various methods exist for confirming the power of an intraocular lens (IOL) prior to implantation, few clinicians perform this step due to the improved quality control by manufacturers, 5, 6, 7, the risk of contamination and the increased surgical time. After the lens has been implanted, however, there has been no method described for directly measuring the power. We present a method using slit lamp photography which measures with an average error of only 0.25 D the IOL power within the eye.

A method is described for determining the power of an intraocular lens (IOL) within the eye by measuring the horizontal dimension of the corneal reflected image (Purkinje-Sanson I) and the anterior IOL reflected image (Purkinje-Sanson III) as seen through a standard slit lamp with a target approximately 68 mm in front of the cornea. The horizontal K-reading (180 degrees) and the anterior chamber depth are the two other parameters necessary to calculate the exact power of the IOL. Seven tables have been provided which utilize these four measurements, eliminating the need for complex calculations.

To determine the accuracy of his technique, ten implanted IOL’s ranging from 9 D to 27 D were chosen, their power calculated and compared to the actual IOL power. The largest error was 0.5 D and the average error was 0.25D.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 39.99
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 54.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 1987 Martinus Nijhoff/Dr W. Junk Publishers, Dordrecht

About this chapter

Cite this chapter

Prager, T.C. et al. (1987). Measuring intraocular lens power within the eye. In: Ossoinig, K.C. (eds) Ophthalmic Echography. Documenta Ophthalmologica Proceedings Series, vol 48. Springer, Dordrecht. https://doi.org/10.1007/978-94-009-3315-6_6

Download citation

  • DOI: https://doi.org/10.1007/978-94-009-3315-6_6

  • Publisher Name: Springer, Dordrecht

  • Print ISBN: 978-94-010-7988-4

  • Online ISBN: 978-94-009-3315-6

  • eBook Packages: Springer Book Archive

Publish with us

Policies and ethics