Left ventricular assist device: Experimental and clinical study

  • Chapter
Artificial Heart 2
  • 89 Accesses

Summary

Recently, intra-aortic balloon pumping (IABP) has been widely employed clinically as a means of mechanical circulatory assistance. However, there have been a few patients who could not be saved from severe heart failure after a cardiac operation in spite of IABP assistance. An evaluation of the left ventricular assist device (LVAD) was experimentally carried out for profound heart failure after acute myocardial infarction, which was produced by multiple ligations of the coronary arteries. Consequently, the efficacy of the LVAD was clearly confirmed hemodynamically in comparison with another means of mechanical assistance such as VA bypass and IABP + VA bypass.

On the basis of our experimental studies, the LVAD was successfully employed in a 68-year-old female patient who underwent closure of a ventricular septal rupture (VSR) and AC bypass for VSR following acute myocardial infarction (AMI). She was weaned from the LVAD and IABP 7 days after surgery. Our experience concerning the LVAD in mechanical circulatory assistance is reported in detail.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 39.99
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 54.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Okada M, Nakamura K (1986) Indications and limitations of IABP. Jap J Thorac Surg 39: 172–181

    PubMed  CAS  Google Scholar 

  2. Pennington DG, Swartz M, Codd JE, Merjavy JP, Kaiser GC (1983) Intraaortic balloon pumping in cardiac surgical patients: A nine-year experience. Ann Thorac Surg 36: 125–131

    Google Scholar 

  3. Goldberger M, Tabak SW, Shak PK (1986) Clinical experience with intraaortic balloon counterpulsation in 112 consecutive patients. Am Heart J 111: 497–502

    Article  PubMed  CAS  Google Scholar 

  4. Sanfelippo PM, Baker NH, Ewy HG, Moore PJ, Thomas JW, Brahos GJ, McVicker RF (1986) Experience with intraaortic balloon counterpulsation. Ann Thorac Surg 41: 36–41

    Article  PubMed  CAS  Google Scholar 

  5. Heitmiller R, Jacobs ML, Daggett WM (1986) Surgical management of postinfarction ventricular septal rupture. Ann Thorac Surg 41: 683–691

    Article  PubMed  CAS  Google Scholar 

  6. Alcan KE, Stertzer SH, Wallsh E, Bruno MS, DePas- 20. quale NP (1984) Current status of intra-aortic balloon counterpulsation in critical care cardiology. Critical Care Medicine 12: 489–495

    CAS  Google Scholar 

  7. Edelman SK (1980) Timing during concomitant abdominal left ventricular assist device (ALVAD) and infra- 21. aortic balloon pumping (IABP) in man. Trans Am Soc Artif Inter Organs 26: 450–454

    CAS  Google Scholar 

  8. Pae WE, Rosenberg G, Donachy H, Landis DL, Phillips WM, Parr GVS, Prophet GA, Pierce WS (1980) Mechanical assistance for postoperative cardiogenic shock: A three year experience. Trans Am Soc Artif 22. Intern Organ 26: 256–261

    Google Scholar 

  9. Atsumi K (1984) Ventricular assistance-Development and clinical application of a new device. In: Unger F (ed) Assisted circulation, 2nd edn. Springer, Berlin pp 100–114

    Chapter  Google Scholar 

  10. Dew PA, Olsen DB, Kessler TR, Coleman DL, Kolff 23. WJ (1984) Mechanical failure in vivo and in vitro studies of pneumatic total artificial hearts. Trans Am Soc Artif Intern Organs 30: 112–116

    CAS  Google Scholar 

  11. Takano H, Nakatani T, Taenaka Y, Umezu M (1986) Development of the ventricular assist pump system: 24. Experimental and clinical studies. In: Akutsu T (ed) Artificial heart 1 Springer, Tokyo, pp 141–151

    Google Scholar 

  12. Nitta S, Kagawa Y, Hongo T, Horiuchi T, Katahira Y, Fujimasa I, Imachi K, Atsumi K (1986) Clinical experi- 25. ence of left and right ventricular assist devices. In: Akutsu T (ed) Artificial heart 1. Springer, Tokyo, pp 153–158

    Google Scholar 

  13. Swan HJ, Forrester JS, Diamond G, Chatterjee K, Parmely WW (1972) Hemodynamic spectrum of myo- 26. cardial infarction and cardiogenic shock. A conceptual model. Circulation 45: 1097–1110 27.

    Google Scholar 

  14. Resnekov L (1983) Cardiogenic shock. Chest 83: 893898 28.

    Google Scholar 

  15. Okada M, Asada S (1979) Assisted circulation-especially IABP. Jpn J Thorac Surg 32: 645–657 29.

    Google Scholar 

  16. Okada M, Horii H, Ikuta H, Shimizu K, Isaji S, Tsuruta H, Takahashi H, Matsuda S, Yano H, Yoneda K, Kawai M, Iizuka M, Hironaka S, Shiozawa T, Okuno 30. K, Nakamura K (1982) The effects of mechanical circulatory assistances for acute myocardial infarction- 31. IABP and left heart bypass (using roller pump, or artificial heart). Jpn J Artif Organs 11: 114–119 32.

    Google Scholar 

  17. Okada M, Shiozawa T, Iizuka M, Okuno K, Hironaka S, Matsuda S, Yoneda K, Yano A, Kawai M, Naka- 33. mura K, Asada S (1984) The effect of intraaortic balloon pumping ( IABP) on cardiogenic shock due to acute myocardial infarction. Kobe J Med Sci 30: 35–51

    Google Scholar 

  18. Kubota M, Okada M, Imai M, Nakamura K (1986) Ex- 34. perimental studies on assisted circulation by centrifugal pump. Presented at the 39 th J Jap Assoc Thorac Surg, Oct. Tokyo

    Google Scholar 

  19. Okada M (1987) An evaluation of simultaneous appli- 35. cation of IABP during ventricular assist device. Jpn Ann Thorac Surg 7: 230–235

    CAS  Google Scholar 

  20. Okada M, Kubota M, Imai M, Koyama Y, Koterazawa T, Nakamura K (1987) Left ventricular assist device: Experimental and clinical study. Proceedings of 2nd international symposium on artificial heart and assist device. p 24

    Google Scholar 

  21. Okada M, Shiozawa T, Iizuka M, Okuno K, Chen CC, Matsuda S, Yoneda K, Yano A, Kawai M, Asada S (1979) Experimental and clinical studies on the effect of intra-aortic balloon pumping for cardiogenic shock following acute myocardial infarction. Artif Organs 3: 241–246

    Google Scholar 

  22. Okada M, Matsuda S, Kusumoto N, Shio K, Ohta T, Shiozawa T, Okuno K, Iizuka M, Kozawa S, Nakamura K (1984) Intraaortic balloon pumping ( IABP) in patients with cardiogenic shock and mechanical complication due to acute myocardial infarction. Jpn Ann Thorac Surg 4: 926–932

    Google Scholar 

  23. Matsuda S, Okada M, Yamamoto S, Ohta T, Ohyabu H, Kuris S, Shida T, Shio J, Tachibana F, Nakamura K, Tsuruta H, Goto T, Ogawa K (1983) Combined use of ECM() with IABP for cardiogenic shock. Jpn J Artif Organs 12: 362–366

    Google Scholar 

  24. Bernhard WF, Clay W, Gernes D, Hougen T, Sherman C, Burke D, Schoen TJ, Poirier VL (1985) Temporary and permanent left ventricular bypass: Laboratory and clinical observations. World J Surg 9: 54–64

    Google Scholar 

  25. Pennington DG, Bernhard WF, Golding LR, Berger RL, Watson JT (1985) Long-term follow-up of postcardiotomy patients with profound cardiogenic shock salvaged by ventricular assist device. Circulation 70 (Suppl II): 70–84

    Google Scholar 

  26. Kafrouni G (1984) Intraaortic balloon counterpulsation. Am J Surg 147: 731–734

    Article  PubMed  CAS  Google Scholar 

  27. Kantrowitz A (1986) The ninth Hastings lecture, Spectrum. Artif Organs 10: 497–510

    Article  PubMed  CAS  Google Scholar 

  28. Pierce Ws (1986) The artificial heart-1986: Partial fulfillment of a promise. ASAIO Trans 32: 5–10

    Google Scholar 

  29. Kresh JY, Kerkhof PL, Goldman SM, Brockman SK (1986) Heart-mechanical assist device interaction. ASAIO Trans 32: 437–443

    Article  PubMed  CAS  Google Scholar 

  30. Cortesini R, Berloco P (1986) Alternatives to heart transplantation. Cardiologia 31: 1065–1968

    PubMed  CAS  Google Scholar 

  31. Wolner E (1986) Artificial heart in heart transplantation-A review. Z Kardiol 75: 125–126

    PubMed  Google Scholar 

  32. Firth BG (1987) Replacement of the failing heart. Am J Med Sci 293: 50–65

    Article  PubMed  CAS  Google Scholar 

  33. Vasku J (1987) Use of total heart replacement in clinical practice. Importance of experimental preparation in relation to legal, ethical and social aspects. Cas Lek Cesk 126: 33–40

    PubMed  CAS  Google Scholar 

  34. Copeland JG, Levinson NM, Smith R, Icenogle TB, Vaughn C, Chung K, Otto R (1986) The total artificial heart as a bridge to transplantation. A report of two cases. JAMA 256: 2991–2995

    Google Scholar 

  35. Griffith BP, Hardesty RL, Kormos RL, Trento A, Borovetz HS, Thompson ME, Bahnson HT (1987) Temporary use of the Jarvik-7 total artificial heart before transplantation. N Engl J Med 316: 130–134

    Article  PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 1988 Springer Japan

About this chapter

Cite this chapter

Okada, M., Kubota, M., Imai, M., Koyama, Y., Nakamura, K. (1988). Left ventricular assist device: Experimental and clinical study. In: Akutsu, T., et al. Artificial Heart 2. Springer, Tokyo. https://doi.org/10.1007/978-4-431-65964-8_21

Download citation

  • DOI: https://doi.org/10.1007/978-4-431-65964-8_21

  • Publisher Name: Springer, Tokyo

  • Print ISBN: 978-4-431-70544-4

  • Online ISBN: 978-4-431-65964-8

  • eBook Packages: Springer Book Archive

Publish with us

Policies and ethics