Summary
This report describes the imaging characteristics of titanium alloy aneurysm clips and our clinical experience with these clips in more than 300 patients. Phantom and clinical investigations showed that clip artifacts on CT and MR are minor as compared to the cobalt alloy clips used previously. Spiral CT angiography (CTA) in combination with titanium alloy clips could be shown to be a feasible mode of postoperative control and can be used to determine completeness of aneurysm elimination, patency of adjacent arteries as well as vasospasm. In contrast, magnetic resonance angiography (MRA) proved to be unfeasible as a method of postoperative vascular imaging since the titanium clips still produce a shadow in the order of size of the entire aneurysm. Therefore, completeness of aneurysm elimination cannot be judged on magnetic resonance angiography. The clinical experience in more than 300 cases showed that the titanium alloy clips essentially performed well. However, the limits of elastic deformation appear to be somewhat inferior to cobalt alloy clips. The standard appliers do not open the blades of the titanium clips quite as far as with the comparable cobalt alloy clips and the corresponding appliers. Therefore the titanium aneurysm clips are not quite as well suited for large broad based aneurysms. Furthermore, the increased susceptibility of the new clips with regard to abuse suggests to follow the recommendations of the manufacturer not to recycle and re-sterilize clips that have been tried but not permanently implanted.
Keywords
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
References
Dorsch NWC, Young N, Kingston RJ, Compton JS (1995) Early experience with spiral CT in the diagnosis of intracranial aneurysms. Neurosurgery 36: 230–238
Fisher RS, Ehsan T, Smith K, Lawton MT, Bichard WD, Spetzler RF (1996) Titanium aneurysm clips, part II. Seizure and electroencephalographic studies in implanted rabbits. Neurosurgery 38: 1165–1169
Harbaugh RE, Schlusselberg DS, Jeffery R, Hayden S, Cromwell LD, Pluta D, English RA (1995) Three-dimensional computed tomographic angiography in the preoperative evaluation of cerebrovascular lesions. Neurosurgery 36: 320–327
Huston III J, Rufenacht DA, Ehmann RL, Wiebers DO (1991) Intracranial aneurysms and vascular malformations: comparison of time-of-flight and phase contrast MR angiography. Radiology 181: 721–730
Kato Y, Sano H, Katada K, Ogura Y, Ninomiya T, Okuma I, Kanno T (1996) Effects of new titanium cerebral aneurysm clips on MRI and CT images. Minim Invasive Neurosurg 39: 82–85
Kassell NF, Torner JC, Haley EC Jr (1990) The international cooperative study on the timing of aneurysm surgery, part 1: overall management results. J Neurosurg 73: 18–36
Lawton MT, Ho JC, Bichard WD, Coons SW, Zabramski JM, Spetzler RF (1996) Titanium aneurysm clips, part I: mechanical, radiological, and biocompatibility testing. Neurosurgery 38: 1158–1164
Lawton MT, Heiserman JE, Prendergast VC, Zabramski JM, Spetzler RF (1996) Titanium aneurysm clips, part III: clinical application in 16 patients with subarachnoid hemorrhage. Neurosurgery 38: 1170–1175
Macdonald RL, Wallace MC, Kestle JRW (1993) Role of angiography following aneurysm surgery. J Neurosurg 79: 826–832
Marchal G, Bosmans H, Van Fraeyenhoven L, Wilms G, Van Hecke P, Plets C, Baert AL (1990) Intracranial vascular lesions: optimization and clinical evaluation of three-dimensional time-of-flight MR angiography. Radiology 175: 443–448
Mayberg MR, Batjer HH, Dacey R, Diringer M, Haley EC, Heros RC, Sternau LL, Torner J, Adams HP, Feinberg W, Thies W (1994) Guidelines for the management of aneurysmal subarachnoid hemorrhage. A statement for healthcare professionals from a special writing group of the stroke council, American Heart Association. Circulation 90: 2592–2505
Payner TD, Tew JM Jr, Steiger HJ (1995) Aneurysm clips. In: Wilkins RH, Rengachary SS (eds) Neurosurgery, 2nd edn. McGraw Hill, New York, pp 2271–2276
Piepgras A, Guckel F, Weik T, Schmiedeck P (1995) Titanium aneurysm clips and their advantages in diagnostic imaging. Radiologe 35: 830–833
Ross JS, Masaryk TJ, Modic MT (1990) Intracranial aneurysms: evaluation by MR angiography. AJNR Am J Neuro-radiol 11: 449–456
Schmid UD, Steiger HJ, Huber P (1987) Accuracy of high resolution computed tomography in direct diagnosis of cerebral aneurysms. Neuroradiology 29: 152–159
Schwartz RB, Tice HM, Hooten SM, Hsu L, Stieg PE (1994) Evaluation of cerebral aneurysms with helical CT: correlation with conventional angiography and MR angiography. Radiology 192: 717–722
Tampieri D, Leblanc R, Oleszek J, Pokrupa R, Melançon D (1995) Three-dimensional computed tomographic angiography of cerebral aneurysms. Neurosurgery 36: 749–755
van Loon JJL, Yousry TA, Fink U, Seelos KC, Reulen HJ, Steiger HJ (1997) Postoperative spiral computed tomography and magnetic resonance angiography after aneurysm clipping with titanium clips. Neurosurgery 41: 851–857
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 1999 Springer-Verlag Wien
About this chapter
Cite this chapter
Steiger, HJ., van Loon, J.J.L. (1999). Virtues and Drawbacks of Titanium Alloy Aneurysm Clips. In: Langmoen, I.A., Lundar, T., Aaslid, R., Reulen, HJ. (eds) Neurosurgical Management of Aneurysmal Subarachnoid Haemorrhage. Acta Neurochirurgica Supplements, vol 72. Springer, Vienna. https://doi.org/10.1007/978-3-7091-6377-1_7
Download citation
DOI: https://doi.org/10.1007/978-3-7091-6377-1_7
Publisher Name: Springer, Vienna
Print ISBN: 978-3-7091-7309-1
Online ISBN: 978-3-7091-6377-1
eBook Packages: Springer Book Archive