Skip to main content

Influence of the position of intravascular ultrasound catheters on the depiction of cross sectional areas in arterial vessels

Einfluß der Position intravaskulärer Ultraschallkatheter auf die Wiedergabe von arteriellen Gefäßquerschnitten

  • Chapter
Diätetik und Arteriosklerose

Abstract

Intravascular ultrasound depicts detailed cross-sectional areas of vessels. Since the catheter position is variable in peripheral vessels, the influence of central/ eccentric and coaxial/angulated catheter position on the measurement of cross-sectional areas in severely atherosclerotic arteries was analysed. In 78 segments of postmortem human iliac arteries (mean vessel diameter 0.96 ± 0.21 cm2 with central catheter position) the mean malestimation of cross-sectional areas with eccentric catheter position was 12.6 %. In 112 segments of aortic and iliac vessels (mean vessel diameter 0.90 ± 0.23 cm2 with coaxial catheter position) catheter angulation of 10 degrees in one plane led to a false estimation up to 5 % in 54 segments, up to 10 % in 33 and more than 10 % in 25 segments.

In atherosclerotic arteries, the measurement of crbss sectional areas requires a central and coaxial catheter position. For comparative measurements markings along the ultrasound catheters minimize intrapositional variability in normal arteries.

Zusammenfassung

Die intravaskuläre Ultraschalldiagnostik ermöglicht die exakte Abbildung von Gefäßen als transaxiale Querschnittsbilder. Da die Katheterposition in peripheren Gefäßen variabel ist, wurde unter in-vitro-Bedingungen in ausgeprägt arteriosklerotischen Gefäßsegmenten der Einfluß der Katheterposition (zentral/exzentrisch und koaxial/anguliert) auf die Bestimmung der Querschnittsflächen ermittelt. Bei 78 Segmenten postmortal entnommener lliakalarterien (mittlerer Gefäßquerschnitt 0,96 ± 0,21 cm2 bei zentraler Katheterposition) betrug die mittlere Abweichung bei exzentrischer Katheterausrichtung 12,6 %. Bei 112 autoptisch gewonnenen iliakalen und aortalen Segmenten (mittlerer Gefäßquerschnitt 0,90 ± 0,23 cm2) führte die Angulation des Katheters um 10 Grad in einer Ebene zu einer Fehleinschätzung der Querschnittsflächen bis zu 5 % bei 54 Segmenten, bis 10 % bei 33 und mehr als 10 % bei 25 Segmenten. Vergleichende Messungen von Gefäßquerschnitten erfordern eine standardisierte Reproduktion der Katheterposition. Unter in-vivo-Bedingungen wurden am Mischlingshund acht verschiedene aortale und 15 verschiedene iliakale Gefäßquerschnitte wiederholend gemessen; die Transducerposition wurde durch Kathetermarkierungen an der arteriellen Schleuse reproduziert. Es ergab sich eine Intraobserver-Variabilität aortal von 1,47 mm2 (aortal) bzw. von 0,64 mm2 (iliakal); die intrapositionelle Variabilität betrug aortal 5,47 mm2 und iliakal 2,05 mm2.

In arteriosklerotischen Gefäßsegmenten ist bei Bestimmung der Querschnitts-flächen eine zentrale und koaxiale Position des Ultraschallkatheters erforderlich. Bei vergleichenden Messungen nichtarteriosklerotischer Gefäßquerschnitte tragen markierte Kathetersysteme zu einer niedrigen intrapositionellen Variabilität bei.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 44.99
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 59.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Literaturverzeichnis

  1. Bauriedel G, Mio SJ, Höfling B. Role of angioscopy in the treatment of peripheral vascular disease with percutaneous atherectomy. Am J Cardiol 1991; 68: 226–231.

    Article  PubMed  CAS  Google Scholar 

  2. Davidson CJ, Sheikh KH, Kisslo KB, Phillips HR, Peter RH, Behar VS, Yihong K, Krucoff M, Ohman EM, Tcheng JE, Stack RS. Intracoronary ultrasound evaluation of interventional technologies. Am J Cardiol 1991; 68: 1305–1309.

    Article  PubMed  CAS  Google Scholar 

  3. Ge J, Erbel R, Seidel I, Gorge G, Reichert T, Gerber T, Meyer J. Experimentelle Überprüfung der Genauigkeit und Sicherheit des intraluminalen Ultraschalls. Z Kardiol 1991; 80: 595–601.

    PubMed  CAS  Google Scholar 

  4. Neville RF, Bartorelli AL, Sidawy AN, Almagor Y, Potkin B, Leon M. An in vivo feasibility study of intravascular ultrasound imaging. Am J Surgery 1989; 158: 142–145.

    Article  CAS  Google Scholar 

  5. Sheikh KH, Davidson CJ, Kisslo KB, Harrison JK, Himmelstein SI, Kisslo J, Bashore T. Comparison of intravascular ultrasound, external ultrasound and digital angiography for evaluation of peripheral artery dimensions and morphology. Am J Cardiol 1991; 67: 817–822.

    Article  PubMed  CAS  Google Scholar 

  6. Weidinger F, Schwarzacher S, Bohm G, Moritz A, Losert U, Glogar D. Direkte Diagnostik der atherosklerotischen Gefäßwand: Möglichkeiten und Grenzen des intravaskulären Ultraschalls. Z Kardiol 1992; 81: 1–8.

    PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 1993 Springer Fachmedien Wiesbaden

About this chapter

Cite this chapter

Kerber, S. et al. (1993). Influence of the position of intravascular ultrasound catheters on the depiction of cross sectional areas in arterial vessels. In: Heinle, H., Schulte, H., Schaefer, H.E. (eds) Diätetik und Arteriosklerose. Vieweg+Teubner Verlag, Wiesbaden. https://doi.org/10.1007/978-3-663-01942-8_35

Download citation

  • DOI: https://doi.org/10.1007/978-3-663-01942-8_35

  • Publisher Name: Vieweg+Teubner Verlag, Wiesbaden

  • Print ISBN: 978-3-663-01943-5

  • Online ISBN: 978-3-663-01942-8

  • eBook Packages: Springer Book Archive

Publish with us

Policies and ethics