Skip to main content

Kontinuierliche Messung des gemischt-venösen Sauerstoffpartialdrucks mittels einer Katheterlektrode

  • Chapter
Leistungsbegrenzung von seiten der Lunge

Abstract

A method for continuous long term measurement of mixed-venous oxygen pressure by means of a catheter electrode is described. Preliminary results in 6 patients show a good correlation to discontinuous measurements with a Radiometer electrode (r = 0, 95, S. E. E. = ± 1, 9 mm Hg). Provided the electrode is calibrated every 3 hours the average deviation of the electrode reading compared with the discontinuous measurement is less than 10%. Changes in the mixture of venous blood coming from the superior and inferior Vena cava as well as variations in the position of the electrode may simulate real alterations of mixed-venous oxygen pressure.

The influence of a changing arterial oxygen pressure and of short term hemodynamic effects on mixed-venous oxygen pressure is demonstrated in some examples. Essentially the importance of this method is to show short term changes of mixed-venous oxygen pressure.

Zusammenfassung

Es wird eine Methode zur kontinuierlichen Langzeitmessung des gemischt-venösen Sauerstoffpartialdrucks mittels einer Katheterelektrode beschrieben. Erste Erfahrungen mit dieser Messung an insgesamt 6 Patienten zeigen eine gute Korrelation mit diskontinuierlich gewonnenen Vergleichswerten (r = 0, 95, Standardschätzfehler ± 1, 9 mm Hg). Bei etwa 3-stündlicher Nacheichung der Elektroden ist mit einer mittleren Abweichung des Elektrodenmeßwertes vom diskontinuierlichen Vergleichswert um weniger als 10% zu rechnen. Die kontinuierliche Messung des rechtsatrialen Sauerstoffpartialdrucks ist problematisch, da wechselnde Mischungsverhältnisse des venösen Blutes aus oberer und unterer Hohlvene sowie geringe Lageänderungen der Meßsonde tatsächliche Änderungen des gemischt-venösen Sauerstoffpartialdrucks vortäuschen können. Der Einfluß von Änderungen des arteriellen Sauerstoffpartialdrucks und von kurzfristigen Änderungen der Hämodynamik auf den gemischt-venösen Sauerstoffpartialdruck wird an Beispielen demonstriert. Die Bedeutung dieser Methode liegt insbesondere in der Darstellung kurzzeitiger Änderungen des gemischt-venösen Sauer stoffpartialdrucks.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 44.99
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 59.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Literatur

  1. Eberhardt, P., Fehlmann, W., Schiebli, R., Büsser, E.: Kontinuierliche PO2-Messung mittels intravasaler Sonden. Verh. Dtsch. Ges. Biomed. Techn. 15. -17. 5. 1974, Hannover

    Google Scholar 

  2. Goeckenjan, G., Schneider, P., Heidenreich, J.: Kontinuierliche P02Uberwachung mittels intravasaler Sauerstoffelektroden. Dtsch. med. Wschr. 101 (1976)

    Google Scholar 

  3. Lee, J., Wright, F., Barber, R., Stanley, L.: Central venous oxygen saturation in shock: a study in man. Anesthesiol. 36, 472–478 (1972)

    Article  CAS  Google Scholar 

  4. Lutch, J. S., Murray, J. F.: Continuous positive pressure ventilation: Effects on systemic oxygen transport and tissue oxygenation. Ann. Intern. Med. 76, 193–202 (1972)

    Article  CAS  PubMed  Google Scholar 

  5. Mindt, W.: Sauerstoffsensor für in vivo-Messung. Dtsch. Ges. Biomed. Techn. 24. -25. 5. 1973, Erlangen

    Google Scholar 

  6. Mithoefer, J. C., Holford, F. D., Keighley, J. F. H. The effect of oxygen administration on mixed venous oxygenation in chronic obstructive pulmonary disease. Chest 66, 122–132 (1974)

    Article  CAS  PubMed  Google Scholar 

  7. Scheinman, M. M., Brown, M. A., Rapaport, E.: Critical assessment of use of central venous oxygen saturation as a mirror of mixed venous oxygen in severely ill cardiac patients. Circulation 40, 165–172 (1969)

    Article  CAS  PubMed  Google Scholar 

  8. Schröder, H. H., Schmidt, K., Beckmann, O. W.: Die Beurteilung des gemischt-venösen Sauer stoffpartialdrucks zur Beurteilung der Herz - leistung insbesondere beim kardiogenen Schock. Intensivmed. 12, 133146 (1975)

    Google Scholar 

  9. Schulz, V., Schnabel, K. H., Erdmann, W.: Beatmung mit positivem end-exspiratorischem Druck–funktionsdiagnostische Untersuchungen und klinische Erfahrungen. Intensivmed. 12, 153–164 (1975)

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 1976 Springer-Verlag Berlin Heidelberg

About this chapter

Cite this chapter

Goeckenjan, G., Schneider, P., Heidenreich, J. (1976). Kontinuierliche Messung des gemischt-venösen Sauerstoffpartialdrucks mittels einer Katheterlektrode. In: Ulmer, W.T. (eds) Leistungsbegrenzung von seiten der Lunge. Verhandlungen der Gesellschaft für Lungen- und Atmungsforschung. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-662-25411-0_20

Download citation

  • DOI: https://doi.org/10.1007/978-3-662-25411-0_20

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-662-23364-1

  • Online ISBN: 978-3-662-25411-0

  • eBook Packages: Springer Book Archive

Publish with us

Policies and ethics