Abstract
Several newtonian liquids with kinematic viscosities below about 1 mm2/s are investigated in order to find out if the commonly used purity specifications are sufficient to qualify the liquids as viscosity standards which can be used to determine the kinetic energy correction of glas capillary viscometers.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
References
DIN 51562 (1978): Messung der kinematischen between 67.1 and 67.9 (average value 67.4, standard Viskosität mit dem Ubbelohde-Viskosimeter.
Cannon, M. R., R. E. Manning, J. D. Bell, Anal. Chem. [3] also gave an average value of 67.4. If, however, 32, 255 (1960).
Meerlender, G., H. Bauer, Rheol. Acta (to be published).
Supplier 1: E. Merck, Labor- und Industriechemikalien, Darmstadt (FRG). Supplier 2: J. T. Baker, Chemicals B. V., Deventer (The Netherlands).
Author information
Authors and Affiliations
Editor information
Rights and permissions
Copyright information
© 1982 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Bauer, H., Meerlender, G. (1982). Newtonian reference materials with v < 1 mm2/s and their application in extending the range of capillary viscometers. In: Giesekus, H., Kirschke, K., Schurz, J. (eds) Progress and Trends in Rheology. Steinkopff, Heidelberg. https://doi.org/10.1007/978-3-662-12809-1_37
Download citation
DOI: https://doi.org/10.1007/978-3-662-12809-1_37
Publisher Name: Steinkopff, Heidelberg
Print ISBN: 978-3-7985-0612-1