Studies of Short — and Long-Term Threshold Changes

  • Chapter

Abstract

The measurement of diastolic stimulation threshold at pacemaker implantation is a necessary prerequisite for the verification of an optimally longterm functioning of implanted pulse generator systems (34). Thereby, the initially measured threshold energy can be primarily seen as an expression of optimal electrode positioning and a technically flawless system of impulse transmission by the pacemaker electrode. Measurements dealing with improved pacemaker electrode technology presuppose a homogeneous group of patients. A uniform basis for the evaluation of later developing pacemaker stimulatory threshold can only be obtained if physiological, patho-physiological and pharmacological conditions are kept constant. On a short-term basis, conditions can be kept constant, usually without difficulty, in order to observe the influences of the individual factors. Long-term continuous observations must on principle be able to differentiate between a summation of factors at implantation and later measurements.

This is a preview of subscription content, log in via an institution.

Buying options

Chapter
USD   29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD   84.99
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD   109.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Learn about institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. H. Avenhaus, H. Grohmann, “Die Refraktärzeit des menschlichen Herzens unter Sparteinsulfat, Untersuchungen mit elektrischer Stimulation”, Z. Kreisl. Forsch. 58, 54 (1969)

    Google Scholar 

  2. J.T. Bigger, D.J. Weinberg, A.T. Kovalik, “Effects of diphenylhydantoin on excitability and automaticity in the canine heart”, Circulation Res. 26 /1, 1–5 (1970)

    Google Scholar 

  3. Ch. Büchner, W. Drägert, Th. Arnold, V. Schlosser, B. Nuber, “Schrittmachertherapie des Herzens”, Forum Cardiologicum 14 (1973) Boehringer Mannheim GmbH

    Google Scholar 

  4. W.M. Chardack, A.A. Gage, A.J. Federico, G. Schimert and W. Greatbatch, “Five years’ clinical experience with an implantable pacemaker”, An appraisal, Surg. 58, 915 (1965)

    Google Scholar 

  5. S.J. Choi, J. Roberts, G.J. Kelliher, “The effect of propranolol and quinidine on Na and K-exchange in the cat papillary muscle”, Eur. J. Pharmacol. 20, 10 (1972)

    Article  Google Scholar 

  6. M. Diewitz und O. Baldus, “Beeinflussung der myokardialen Reizschwelle durch verschiedene Antiarrhythmika und Sympathikomimetica”, Arzneimittel-Forsch. 23, 511 (1973)

    Google Scholar 

  7. M. Diewitz, D. Heinrich, O. Baldus, “Über das Verhalten der myokardialen Reizschwelle nach intrakardialer Elektrodenimplantation”, Verh. Dtsch. Ges. Inn. Med. 77 (1971)

    Google Scholar 

  8. M. Diewitz, G. Schramm und D. Heinrich, “Vor-und Nachteile der zweiseitigen Schrittmacherimplantation”, Minch. Med. Wschr. 114, 731 (1972)

    Google Scholar 

  9. P. Doenecke, G. Rettig, L. Bette, “Änderungen der myokardialen Reizschwelle unter Elektrostimulation”, Z. Kardiol. 63, 131 (1974)

    Google Scholar 

  10. G. Faucon, J.C. Evreux, M. Ollagnier, R. Bazaugour, “Effects isoles des digitaliques sur l’automatisme cardiaque normal et ectopique”, Comptes Rend. Soc. Biol. 165 /1, 136 (1971)

    Google Scholar 

  11. L.S. Gettes, R. Shabetai, T.A. Downs and B. Surawicz, “Effect of changes in potassium and calcium concentrations on diastolic threshold and strength interval relationships of the human heart”, Ann. N.Y. Acad. Sci. 167, 693 (1969)

    Article  Google Scholar 

  12. U. Gleichmann, L. Seipel, F. Loogen, “Der Einfluß von Antiarrhythmika auf die intrakardiale Erregungsleitung (His-Bündel-Elektrokardiographie) und Sinusknotenautomatie beim Menschen”, Dtsch. med. Wschr. 32, 1487 (1973)

    Article  Google Scholar 

  13. E.A. Gubarev. T.A. Butseva, “Diprazine and benadryl action on the bathmotropic effect of strophanthin”, Farmakol. Toksikol. 33 /5, 560 (1970)

    Google Scholar 

  14. N. Heinz, “Beeinflussung der diastolischen Reizschwelle des menschlichen Herzens durch Antiarrhythmika”, Arzneim.-Forsch. 21, 1189 (1971)

    Google Scholar 

  15. R. Koch, “Klinische und experimentelle Untersuchungen über die Wirkung und Wirkdauer von N-n-Propylajmaliniumhydrogentartrat bei Herzrhythmusstörungen”, Arzneim.-Forsch. 22, 2079 (1972)

    Google Scholar 

  16. B. Luederitz, H.D. Bolte, “Zur kardialen Wirkung von Glucagon. Elektrophysiologische Messungen am Papillarmuskel des Herzens”, Z. Kreist. Forsch. 60 /2, 130 (1971)

    Google Scholar 

  17. K. Noeske, W. Knothe, “Pathologische Herzveränderungen im Bereich von Schritt-macherelektroden”, Thoraxchir. 13, 381 (1965)

    Google Scholar 

  18. J.C. Norman, R. Lightwood and L.D. Abrahams, “Surgical treatment of Adams-Stokes syndrome using long-term inductive coupled coil pacemaking”, Ann. Surg. 159, 344 (1964)

    Article  Google Scholar 

  19. M. Pieniak, R.J. Zochowski, M.J. Stopczyk, “Rising threshold to electrical stimulation in ischemic area during short coronary occlusion”, IVth Intern. Sympos. on Card. Pacing, Groningen 1973

    Google Scholar 

  20. T.A. Preston, R.D. Fletcher, B.R. Lucchesi, R.D. Judge, “Changes in myocardial threshold. Physiologic and pharmacologic in patients with implanted pacemaker”, Am. Heart J. 74, 235 (1967)

    Article  Google Scholar 

  21. B. Ramdohr, E.V. Leitner, “Zur Methodik der Reizschwellenbestimmung nach Herzschrittmacherimplantation”, Medizinalmarkt/Acta Medicotechnika 21, 35–37 (1973)

    Google Scholar 

  22. M.R. Rosen, H. Gelbrand, B.F. Hoffmann, “Quabain effects on single cardiac cells and their electrocardiagraphic corrolary”, Circulation 44/4 Suppl. 2 (1971) II - 84

    Google Scholar 

  23. H. Schieffer, L. Bette, P. Doenecke, G. Harbauer, H. Heinz, W. Hoffmann, K. Hoffmann, G. Hofmeier, Chr. Klein, H. Mannebach, G. Rettig und H. Steeb, “Erfahrungen mit intrakardialen Herzschrittmachern”, Minch. Med. Wschr. 51, 1712 (1971)

    Google Scholar 

  24. H.K. Schulten, H. Etzrodt, W. du Mesnil de Rochemont, H.W. Chriske, K.D. Grossner, G. Steinbrück, “Klinische und elektrophysiologische Beobachtungen bei der antiarrhythmischen Therapie mit Diphenylhydantoin”, Verh. Dt. Ges. inn. Med. 77, 952 (1971)

    Google Scholar 

  25. H.K. Schulten, K.D. Grossner, G. Steinbrück, “Langzeitmessungen von Reiz-schwellen bei endokardialen Schrittmachersonden”, Zschr. Kardiol. 62 /7, 617 (1973)

    Google Scholar 

  26. H.K. Schulten, K.D. Grosser, A. Vogel, G. Steinbrück, H. Etzrodt, H.W. Chriske, H. Redlich und W.D. Hirschmann, “Vor-und Nachteile der zweiseitigen Schrittmacherimplantation”, Intensivmedizin 9 /14, 214 (1972)

    Google Scholar 

  27. E. Sowton, I. Barr, “Physiological changes in threshold”, Ann. N.Y. Acad. Sci. 167, 679 (1969)

    Article  Google Scholar 

  28. F. Unger, K. Steinbach, “Die mathematische Formulierung des Reizschwellenverlaufes zur Objektivierung ihres Anstiegs”, II. Chirurg. Klinik und Kardiolog. Univ.-Klinik Wien (1973)

    Google Scholar 

  29. W.J. Walker, J.T. Elkins, L.W. Wood, “Effect of potassium in restoring myocardial response to a subthreshold cardiac pacemaker”, New Engl. J. Med. 271, 597 (1964)

    Article  Google Scholar 

  30. K. Wegener, “Pathologisch-anatomische Befunde an menschlichen Herzen nach Schrittmacherimplantation”, Verh. Dt. Ges. Kreisl. Forsch. 35, 246 (1969)

    Google Scholar 

  31. W. Weissenhofer, “Die myokardiale Reizschwelle nach Implantation intrakardialer Elektroden”, Biomed. Techn. 17, 13 (1972)

    Article  Google Scholar 

  32. J.C. Westerholm, “Threshold studies in transvenous cardiac pacemaker treatment”, Scand. J. Thorac a. Cardiovasc. Surg. Suppl. 8 (1971)

    Google Scholar 

  33. K.W. Westermann, O. Giebel, P. Kalmar, G. Priester, K.D. Scheppokat, “Das Verhalten der elektrischen Reizschwelle bei Stimulation mit endokardialen Elektroden”, Verh. Dtsch. Ges. Kreisl. Forsch. 35, 259 (1969)

    Google Scholar 

  34. E. Windisch, H. Fürnrohr, U. Worbs, “Reizschwellenmessungen als Voraussetzung für eine optimale Funktion implantierbarer Schrittmachersysteme”, Z. Kreisl. Forsch. 58 /7, 724 (1969)

    Google Scholar 

  35. M.W. Young, H. Meia, S. Furman, D.J.W. Escher, “Effects of lidocaine and quabain on myocardial threshold to pacer stimuli”, Circulation 42/4 Suppl. 111 209 (1970)

    Google Scholar 

  36. M.W. Young, H. Meia, S. Furman, D.J.W. Escher, “Effect of lidocaine on myocardial threshold to pacer stimuli”, Clin. Res. 18 /2, 335 (1970)

    Google Scholar 

Download references

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 1975 Springer-Verlag Berlin-Heidelberg

About this chapter

Cite this chapter

Doenecke, P., Flöthner, R., Rettig, G., Bette, L. (1975). Studies of Short — and Long-Term Threshold Changes. In: Schaldach, M., Furman, S. (eds) Engineering in Medicine. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-66187-7_17

Download citation

  • DOI: https://doi.org/10.1007/978-3-642-66187-7_17

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-642-66189-1

  • Online ISBN: 978-3-642-66187-7

  • eBook Packages: Springer Book Archive

Publish with us

Policies and ethics