Characterisation of Vitamin E–Blended UHMWPE for Higher In Vivo Performance in Orthopaedic Arthroplasty

  • Chapter
  • First Online:
Total Hip Arthroplasty

Abstract

State-of-the-art cross-linked UHMWPE and recent findings about UHMWPE oxidation from in vivo stresses lead to the need of a better understanding of why anti-oxidant additives have to be added in orthopaedics and which anti-oxidant additivation method is the best option for the use in joint replacements. A GUR 1020 cross-linked vitamin E–blended UHMWPE has been investigated to provide an accurate outline of its properties.

DSC and FTIR measurements, together with ageing and tensile tests, were performed on compression-moulded blocks, as well as biocompatibility tests, including implantation on rabbits. Moreover, wear simulations on finished components (Delta acetabular liners, LimaCorporate, Italy) have been completed.

All the test procedures have been repeated for a reference material, a GUR 1050 cross-linked and remelted standard UHMWPE (UHMWPE X-Lima, LimaCorporate, Italy), and the outcomes have been compared to the cross-linked vitamin E–blended UHMWPE ones.

On the additivated UHMWPE, we found an ultimate tensile strength of 43 MPa, a yield strength value of 25 MPa and an elongation to breakage equal to 320%. The degree of crystallinity was 45 ± 2%, and no signal of creation of oxidation products was detected up to 2,000 h of permanence in oxidant ambient after the ageing test. The reference material showed lower mechanical resistance values (σ* = 33 MPa, σ y = 19 MPa, 305% elongation), a crystallinity of 46 ± 2% and the creation of oxidation products starting from 700 h in O2 ambient. The biocompatibility tests indicate that the additivated material is biocompatible, as the reference X-Lima UHMWPE.

Wear tests gave a wear rate of 5.12 mg/million cycles against 6.13 mg/million cycles of the reference material and no sign of run-in wear rate.

Our results indicate that there is a loss in mechanical properties in the reference material due to the post-irradiation thermal treatment. DSC measurements instead show no change in crystallinity between the additivated UHMWPE and the reference material. The blend between polymer and additive assures a uniform concentration of vitamin E across the whole thickness of the moulded block, and ageing test results on additivated UHMWPE have shown that the material possesses a superior resistance to degradation phenomena.

Biocompatibility assesses that the presence of vitamin E is not detrimental for the in vivo use of the material, and wear results indicate a better wear resistance of the material, especially in the first stages of the wear process.

From these considerations, it can be concluded that the choice of a vitamin E–blended material, in respect to the standard cross-linked UHMWPE, is highly resistant to oxidation phenomena and retains the mechanical properties thanks to the absence of a post-irradiation thermal treatment; therefore, it is expected to have superior in vivo endurance performance.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 84.99
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 109.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

References

  1. ASTM International (2002) ASTM F2003–02. Standard practice for accelerated aging of ultra-high molecular weight polyethylene after gamma irradiation in air. West Conshohocken

    Google Scholar 

  2. ASTM International (2010) ASTM D638–10. Standard test method for tensile properties of plastics. West Conshohocken

    Google Scholar 

  3. ASTM International (2007) ASTM F648–07. Standard specification for ultra-high-molecular-weight polyethylene powder and fabricated form for surgical implants. West Conshohocken

    Google Scholar 

  4. ASTM International (2007) ASTM F2695–07. Standard specification for ultra-high molecular weight polyethylene powder blended with alpha-Tocopherol (Vitamin E) and fabricated forms for surgical implant applications. West Conshohocken

    Google Scholar 

  5. Bhateja S, Duerst R, Aus E et al (1995) Free radicals trapped in polyethylene crystals. J Macromol Sci Phys B34:263–272

    CAS  Google Scholar 

  6. Blundell DJ, Beckett DH, Wilcocks PH (1981) Routine crystallinity measurements of polymers by DSC. Polymer 22:704–711

    Article  CAS  Google Scholar 

  7. Blunn G, Brach del Prever EM, Costa L et al (2002) Ultra high molecular-weight polyethylene (UHMWPE) in joint replacement: fabrication, sterilisation and wear. J Bone Joint Surg Br 84:946–949

    Article  PubMed  CAS  Google Scholar 

  8. Bracco P, Oral E (2011) Vitamin E-stabilized UHMWPE for total joint implants. Clin Orthop Relat Res 469:2286–2293

    Article  PubMed  Google Scholar 

  9. Bracco P, Brunella V, Zanetti M et al (2007) Stabilisation of UHMWPE with vitamin E. Polym Degrad Stab 92:2155–2162

    Article  CAS  Google Scholar 

  10. Costa L, Bracco P (2009) Mechanisms of crosslinking, oxidative degradation and stabilization of UHMWPE. In: Kurtz SM (ed) The UHMWPE biomaterials handbook, 2nd edn. Elsevier Academic Press, Amsterdam

    Google Scholar 

  11. Costa L, Brach del Prever EM (2000) UMWPE for arthroplasty. Minerva Medica, Torino

    Google Scholar 

  12. Costa L, Luda MP, Trossarelli L (1997) Ultra-high molecular weight polyethylene. I. Mechano-oxidative degradation. Polym Degrad Stab 55:329–338

    Article  CAS  Google Scholar 

  13. Costa L, Luda MP, Trossarelli L (1997) Ultra-high molecular weight polyethylene. II. Thermal- and photo-oxidation. Polym Degrad Stab 58:41–54

    Article  CAS  Google Scholar 

  14. Costa L, Luda MP, Trossarelli L et al (1998) Oxidation in orthopaedic UHMWPE sterilized by gamma radiation and ethylene oxide. Biomaterials 19:659–668

    Article  PubMed  CAS  Google Scholar 

  15. Costa L, Luda MP, Trossarelli L et al (1998) In vivo UHMWPE biodegradation of retrieved prosthesis. Biomaterials 19:1371–1385

    Article  PubMed  CAS  Google Scholar 

  16. Costa L, Jacobson K, Bracco P et al (2002) Oxidation of orthopaedic UHMWPE. Biomaterials 23:1613–1624

    Article  PubMed  CAS  Google Scholar 

  17. Gencur SJ, Rimnac CM, Kurtz SM (2006) Fatigue crack propagation resistance of virgin and highly crosslinked thermally treated UHMWPE. Biomaterials 27:1550–1557

    Article  PubMed  CAS  Google Scholar 

  18. Gray P (1970) Polymer crystallinity determination by DSC. Thermochimica Acta 1:563–577

    Article  CAS  Google Scholar 

  19. Harris W (1995) The problem is osteolysis. Clin Orthop Relat Res 311:46–53

    PubMed  Google Scholar 

  20. Hermida JC, Bergula A, Chen P et al (2003) Comparison of the wear rates of twenty-eight and thirty-two-millimeter femoral heads on cross-linked polyethylene acetabular cups in a wear simulator. J Bone Joint Surg Am 85:2325–2331

    PubMed  Google Scholar 

  21. International Organisation for Standardisation (1997) UNI EN ISO 527–2. Determination of tensile properties test conditions for moulding and extrusion plastics. Geneva

    Google Scholar 

  22. International Organisation for Standardisation (2002) ISO 14242. Implants for surgery – wear of total hip-joint prostheses. Geneva

    Google Scholar 

  23. International Organisation for Standardisation (2009) ISO 10993. Biological evaluation of medical devices. Geneva

    Google Scholar 

  24. Keller A, Ungar G (1983) Radiation effects and crystallinity in polyethylene. J Phys Chem 22:155–181

    CAS  Google Scholar 

  25. Marshall A (2011) Effect of UHMWPE particles on mesenchymal stem cell replication. In: Proceedings of 5th UHMWPE international meeting, Philadelphia, 22–23 Sept 2011

    Google Scholar 

  26. McKellop H, Shen FW, Lu B et al (2000) Effect of sterilization method and other modifications on the wear resistance of acetabular cups made of ultra-high molecular weight polyethylene: a hip simulator study. J Bone Joint Surg Am 82:1708–1725

    PubMed  Google Scholar 

  27. Oral E, Wannomae KK, Hawkins NE et al (2004) α-Tocopherol doped irradiated UHMWPE for high fatigue resistance and low wear. Biomaterials 25:5515–5522

    Article  PubMed  CAS  Google Scholar 

  28. Oral E, Greenbaum ES, Malhi AS et al (2005) Characterization of irradiated blends of alpha-tocopherol and UHMWPE. Biomaterials 26:6657–6663

    Article  PubMed  CAS  Google Scholar 

  29. Oral E, Christensen S, Malhi A et al (2006) Wear resistance and mechanical properties of highly crosslinked UHMWPE doped with vitamin E. J Arthroplasty 21:580–591

    Article  PubMed  Google Scholar 

  30. Oral E, Rowell SL, Muratoglu OK (2006) The effect of α-tocopherol on the oxidation and free radical decay in irradiated UHMWPE. Biomaterials 27:5580–5587

    Article  PubMed  CAS  Google Scholar 

  31. Oral E, Wannomae KK, Rowell SL et al (2007) Diffusion of vitamin E in ultra-high molecular weight polyethylene. Biomaterials 28:5225–5237

    Article  PubMed  CAS  Google Scholar 

  32. Oral E, Godesky Beckos C, Malhi A et al (2008) The effects of high dose irradiation on the cross-linking of vitamin E-blended UHMWPE. Biomaterials 29:3557–3560

    Article  PubMed  CAS  Google Scholar 

  33. Parth M, Aust N, Lederer K (2002) Studies on the effect of electron beam irradiation on the molecular structure of ultra-high molecular weight polyethylene under the influence of alpha-tocopherol with respect to its application in medical implants. J Mater Sci Mater Med 13:917–921

    Article  PubMed  CAS  Google Scholar 

  34. Regis M, Bracco P, Giorgini L et al (2012) J Biomed Mater Res B: Appl Biomat (in press)

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Luigi Costa .

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2012 EFORT

About this chapter

Cite this chapter

Costa, L., Regis, M., Bracco, P., Giorgini, L., Fusi, S. (2012). Characterisation of Vitamin E–Blended UHMWPE for Higher In Vivo Performance in Orthopaedic Arthroplasty. In: Knahr, K. (eds) Total Hip Arthroplasty. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-27361-2_5

Download citation

  • DOI: https://doi.org/10.1007/978-3-642-27361-2_5

  • Published:

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-642-27360-5

  • Online ISBN: 978-3-642-27361-2

  • eBook Packages: MedicineMedicine (R0)

Publish with us

Policies and ethics