Abstract

In this article, a new tool for the intraoperative measurement of distances within the middle ear by means of a micromanipulator is presented. The purpose of this work was to offer the surgeon a highly accurate tool for measuring the distances between two points in the 3D operational field. The tool can be useful in various operations; this article focuses, however, on measuring the distance between the stapes footplate and the long process of the incus of the middle ear. This distance is important for estimating the proper prosthesis length in stapedotomy for treating otosclerosis. We evaluated the system using a simplified mechanical model. Our results show that the system can measure distances with a maximum error of 0.04 mm.

Keywords

Middle Ear Surgery Telemanipulator Stapedotomy Measurement 

References

  1. 1.
    Strauss, G., Hipp, S., Fischer, M., Hofer, M., Munz, F., Dietz, A., Strauss, M., Lüth, T.: Intraoperative volumetry in ENT surgery. Objectivation of Surgical Success by the Volume Control System. HNO 58, 839–845 (2010)Google Scholar
  2. 2.
    Theissing, J., Rettinger, G., Werner, J.A.: ENT-Head and Neck Surgery: Essential Procedures. Thieme Medical Pub. (2010)Google Scholar
  3. 3.
    Edelstein, D.R.: Revision Stapedectomy and Stapedotomy. Thieme (2008)Google Scholar
  4. 4.
    Gros, A., Vatovec, J., Žargi, M., Jenko, K.: Success rate in revision stapes surgery for otosclerosis. Otology & Neurotology 26, 1143–1148 (2005)CrossRefGoogle Scholar
  5. 5.
    Heermann, R., Hauger, C., Issing, P.R., Lenarz, T.: Application of Optical Coherence Tomography (OCT) in Middle Ear Surgery. Laryngo-Rhino-Otologie 81(6), 400–405 (2002)CrossRefGoogle Scholar
  6. 6.
    Bozzato, A., Struffert, T., Hertel, V., Iro, H., Hornung, J.: Analysis of the accuracy of high-resolution computed tomography techniques for the measurement of stapes prostheses. European Radiology 20, 566–571 (2010)CrossRefGoogle Scholar
  7. 7.
    Maier, T., Strauss, G., Hofer, M., Kraus, T., Runge, A., Stenzel, R., Berger, T., Dietz, A., Lueth, T.C.: A New Micromanipulator System for Middle Ear Surgery. In: IEEE International Conference on Robotics and Automation, pp. 1568–1573 (2010)Google Scholar
  8. 8.
    Hein, A., Lueth, T.C.: Image-Based Control of Interactive Robotics Systems. In: Taylor, C., Colchester, A. (eds.) MICCAI 1999. LNCS, vol. 1679, pp. 1125–1132. Springer, Heidelberg (1999)CrossRefGoogle Scholar
  9. 9.
    Schauer, D., Hein, A., Lüth, T.C.: Dynamic force control for a miniaturised medical robot system. In: Proceedings 2003 IEEE/ASME International Conference on Advanced Intelligent Mechatronics (AIM 2003), pp. 1090–1095. IEEE, Port Island (2003)CrossRefGoogle Scholar
  10. 10.
    Koulechov, K., Rapoport, T., Lueth, T.C.: Miniaturized, Autoclavable Surgical Robot (Miniaturisierter, autoklavierbarer Roboter für die Chirurgie). at - Automatisierungstechnik 54, 213–221 (2006)CrossRefGoogle Scholar
  11. 11.
    Tokuda, J., Fischer, G.S., Papademetris, X., Yaniv, Z., Ibanez, L., Cheng, P., Liu, H., Blevins, J., Arata, J., Golby, A.J., Kapur, T., Pieper, S., Burdette, E.C., Fichtinger, G., Tempany, C.M., Hata, N.: OpenIGTLink: an open network protocol for image-guided therapy environment. The International Journal of Medical Robotics and Computer Assisted Surgery 5, 423–434 (2009)CrossRefGoogle Scholar

Copyright information

© Springer-Verlag Berlin Heidelberg 2011

Authors and Affiliations

  • Thomas Maier
    • 1
  • Gero Strauss
    • 2
  • Franz Bauer
    • 1
  • Andreas Grasser
    • 1
  • Nobuhiko Hata
    • 3
  • Tim C. Lueth
    • 1
  1. 1.Institute of Microtechnology and Medical Device Technology (MiMed)Technische Universität MünchenGermany
  2. 2.Department of ENT, University Hospital Leipzig and BMBF Innovation Center Computer Assisted Surgery (ICCAS)University LeipzigGermany
  3. 3.Department of Radiology, Harvard Medical SchoolBrigham and Women’s HospitalUSA

Personalised recommendations