Abstract
In order to compensate for the disadvantages of passive drainage of cerebrospinal fluid and to get information on optimal drainage, a mechatronic extracorporal ventricular drainage system has been developed at the Helmholtz institute. A PI-controller adjusts the position of a linear tube squeezer to obtain the target intracranial pressure value. This paper presents the developed active drainage and first measurements of the drainage on a test rig including results of an online algorithm to detect the system’s hydraulic compliance.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2011 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Krause et al. (2011). Brain Pressure Dynamics and Control with an External Ventricular Drainage. In: Jobbágy, Á. (eds) 5th European Conference of the International Federation for Medical and Biological Engineering. IFMBE Proceedings, vol 37. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-23508-5_82