Abstract
The work presents the experimental results concerning the electromagnetic characteristics of a shielded room destined to bio-electromagnetism researches and to the utilization of a high resolution SQUID magnetometer. The experimental results regarding the dependence of the shielding factor on the applied field frequency are presented. The phase of the residual magnetic field and its dependence on the frequency of the external field has also been determined. The experimental results confirm the theoretical studies on the utilization of the non-ferromagnetic material shields.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2011 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Rau, M.C., Baltag, O., Rau, I. (2011). Applications of the Hybrid Shielding in Biomagnetometry. In: Vlad, S., Ciupa, R.V. (eds) International Conference on Advancements of Medicine and Health Care through Technology. IFMBE Proceedings, vol 36. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-22586-4_31
Download citation
DOI: https://doi.org/10.1007/978-3-642-22586-4_31
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-22585-7
Online ISBN: 978-3-642-22586-4
eBook Packages: EngineeringEngineering (R0)