Abstract

Thickness irregularity of active layer is made to express density irregularity. True data by the X-rays are extracted by exposing Ultraviolet (UV) rays that prohibited exposure are exposed for radiochromic film (RF). When UV is exposed, the density irregularity is corrected. In addition, RF is initialized, thereby improving of data acquisition.

GAFCHROMIC EBT (GAF-EBT) film was exposed to UV light at 360 nm twice with a 2 hour interval. The distance from the UV tube to the GAF-EBT film was 165 cm. The GAF-EBT films were scanned after the first and second UV exposure, using a flat bed scanner. Analyses were performed by the density profile curve of the short and long axes of the GAFEBT.

The density irregularity of the active layer of the film was corrected to 2.19 ± 0.49 (pixel value) by subtracting the density obtained with the first UV exposure from that with the second UV exposure of GAF-EBT on the short axis. On the long axis, the density irregularity of the active layer of GAFEBT was corrected to 2.67 ± 0.65 (pixel value).

Uniform UV exposure data was identified. The measurement precision of the GAF-EBT film and its usefulness are improved by this method.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 259.00
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 329.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2009 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Katsuda, T. et al. (2009). Reducing Non-uniformity Error of Radiochromic Film in the Diagnostic Range by Ultraviolet Exposure: Preliminary Study. In: Dössel, O., Schlegel, W.C. (eds) World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany. IFMBE Proceedings, vol 25/3. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-03902-7_64

Download citation

  • DOI: https://doi.org/10.1007/978-3-642-03902-7_64

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-642-03901-0

  • Online ISBN: 978-3-642-03902-7

  • eBook Packages: EngineeringEngineering (R0)

Publish with us

Policies and ethics