Recommendations of the American Association of Physicists in Medicine TG-43U1 report have been followed to characterize the Varisource VS2000 192Ir high dose rate brachytherapy source, provided by Varian Oncology Systems. Monte Carlo calculations with PENELOPE code have been carried out in order to obtain dosimetric parameters for this source: air kerma strength, dose rate constant, radial dose function and anisotropy function. These quantities are compared to the source reference data, finding results in agreement with them taking respective uncertainties into account. Outcomes stated in this paper provide a complete parameter collection, which can be used for dosimetric calculations as well as a means of comparison with other datasets from this source
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2009 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Casado, F.J., Mateo, B., Cenizo, E., García-Pareja, S., Galán, P., Bodineau, C. (2009). Monte Carlo dosimetry with PENELOPE code of the VariSource VS2000 192Ir high dose rate brachytherapy source. In: Dössel, O., Schlegel, W.C. (eds) World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany. IFMBE Proceedings, vol 25/1. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-03474-9_43
Download citation
DOI: https://doi.org/10.1007/978-3-642-03474-9_43
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-03472-5
Online ISBN: 978-3-642-03474-9
eBook Packages: EngineeringEngineering (R0)