The Physikalisch-Technische Bundesanstalt (PTB) is developing a primary standard to determine the absorbed dose to water within a water phantom for X-ray brachytherapy sources such as I-125 and Pd-103. A candidate for the standard to be newly established is a water-equivalent parallel-plate ionization chamber with continuously adjustable electrode spacing, forming an extrapolation chamber in a phantom of water-equivalent material.
A newly contructed exemplar of this chamber was put into operation. The chamber design is presented and the method of evaluation is summarized briefly. Results obtained with this chamber for ISO reference x-ray radiation qualities N20, N25, N30 were compared to those obtained by a free-air chamber, when a conversion factor to the absorbed dose to water is applied.
A good agreement within 2 % of the values for the absorbed dose to water obtained by the two methods could be stated, which demonstrate that the operability of the chamber is now given.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2009 Springer-Verlag Berlin Heidelberg
About this paper
Cite this paper
Schneider, T., Meier, M., Selbach, H.J. (2009). A new device for the measurement of the absorbed dose to water for low energy x-ray sources used in brachytherapy. In: Dössel, O., Schlegel, W.C. (eds) World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany. IFMBE Proceedings, vol 25/1. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-03474-9_36
Download citation
DOI: https://doi.org/10.1007/978-3-642-03474-9_36
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-03472-5
Online ISBN: 978-3-642-03474-9
eBook Packages: EngineeringEngineering (R0)