Calibration of Therapy-level Dosimeters in Japan by Association for Nuclear Technology in Medicine (ANTM)

  • Conference paper
World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany

Part of the book series: IFMBE Proceedings ((IFMBE,volume 25/1))

  • 105 Accesses

Abstract

In 2004, the calibration system of therapy-level dosimeter sensitivity has been reformed in Japan. An organization of new dosimeter calibration system by Association for Nuclear Technology in Medicine (ANTM) was launched from April, 2004 instead of the old system which was operated by Japan Radiological Society (JRS). Since then, about 5 years have passed, and the calibration work has been carried out favorably in general. It is thought that the transfer of dosimeter calibration system was performed smoothly.

The reference dosimeters of ANTM are calibrated by the comparison with the standard dosimeter owned by National Institute of Advanced Industrial Science and Technology (AIST) that is the PSDL of radiation dose in Japan. ANTM acts as a SSDL practically, i.e. ANTM calibrates dosimeters used in clinical institutions in whole Japan. As the old system by JRS had several calibration steps, the reduction of uncertainties accompanying with standard transfer steps would be expected in new system by ANTM.

The total of the dosimeter calibrated by ANTM from April/’04 to March/’09 was 3,407 and the total of ionization chamber was 7,746 that are divided into two groups of 4,900 cylindrical type (63%) and 2,846 shallow type (37%). The average of ratios of calibration factor to previous one was almost 1.000 and the standard deviation of ratios was almost 0.006 every year. Therefore, the variation of calibration factors delivered by ANTM are so small that the calibration system by ANTM works in stable and traceability of therapy level dosimeter in Japan is maintained effectively and in high level.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 259.00
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 329.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2009 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Sakata, S. et al. (2009). Calibration of Therapy-level Dosimeters in Japan by Association for Nuclear Technology in Medicine (ANTM). In: Dössel, O., Schlegel, W.C. (eds) World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany. IFMBE Proceedings, vol 25/1. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-03474-9_273

Download citation

  • DOI: https://doi.org/10.1007/978-3-642-03474-9_273

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-642-03472-5

  • Online ISBN: 978-3-642-03474-9

  • eBook Packages: EngineeringEngineering (R0)

Publish with us

Policies and ethics