Abstract
Clinical evaluation of dose reduction studies in x-ray breast imaging is problematic because it is difficult to justify imaging the same patient at a variety of radiation doses. One common alternative is to use simulation algorithms to manipulate a standard-dose exam to mimic reduced doses. Although there are several dose-reduction simulation methods for full-field digital mammography, the availability of similar methods for digital breast tomosynthesis (DBT) is limited. This work proposes a method for simulating dose reductions in DBT, based on the insertion of noise in a variance-stabilized domain. The proposed method has the advantage of performing signal-dependent noise injection without knowledge of the noiseless signal. We compared clinical low-dose DBT projections and reconstructed slices to simulated ones by means of power spectra, mean pixel values, and local standard deviations. The results of our simulations demonstrate low error (<5 %) between real and simulated images.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
References
Samei, E., Eyler, W., Baron, L.: Effects of anatomical structure on signal detection. In: Handbook of Medical Imaging, pp. 655–682 (2000)
Saunders, R.S., Samei, E.: A method for modifying the image quality parameters of digital radiographic images. Med. Phys. 30, 3006–3017 (2003)
Båth, M., Håkansson, M., Tingberg, A., Månsson, L.G.: Method of simulating dose reduction for digital radiographic systems. Radiat. Prot. Dosimetry 114, 253–259 (2005)
Kroft, L.J.M., Veldkamp, W.J.H., Mertens, B.J.A., Van Delft, J.P.A., Geleijns, J.: Detection of simulated nodules on clinical radiographs: dose reduction at digital posteroanterior chest radiography. Radiology 241(2), 392–398 (2006)
Veldkamp, W.J.H., Kroft, L.J.M., Van Delft, J.P.A., Geleijns, J.: A technique for simulating the effect of dose reduction on image quality in digital chest radiography. J. Digit. Imaging 22(2), 114–125 (2009)
Svalkvist, A., Båth, M.: Simulation of dose reduction in tomosynthesis. Med. Phys. 37, 258–269 (2010)
Mackenzie, A., Dance, D.R., Workman, A., Yip, M., Wells, K., Young, K.C.: Conversion of mammographic images to appear with the noise and sharpness characteristics of a different detector and x-ray system. Med. Phys. 39, 2721 (2012)
Mackenzie, A., Dance, D.R., Diaz, O., Young, K.C.: Image simulation and a model of noise power spectra across a range of mammographic beam qualities. Med. Phys. 41, 121901-1–121901-14 (2014)
Borges, L.R., Oliveira, H.C.R., Nunes, P.F., Vieira, M.A.C.: Method for inserting noise in digital mammography to simulate reduction in radiation dose. In: Proceedings of the SPIE 9412, Medical Imaging, 94125J (2015)
Cockmartin, L., Bakic, P.R., Bosmans, H., Maidment, A.D., Gall, H., Zerhouni, M., Marshall, N.W.: Power spectrum analysis of an anthropomorphic breast phantom compared to patient data in 2D digital mammography and breast tomosynthesis. In: Fujita, H., Hara, T., Muramatsu, C. (eds.) IWDM 2014. LNCS, vol. 8539, pp. 423–429. Springer, Heidelberg (2014)
Acknowledgements
The authors would like to thank São Paulo Research Foundation (FAPESP grant# 2013/18915-5) and the Brazilian Foundation for the Coordination of Improvement of Higher Education Personnel (CAPES grant# 88881.030443/2013-01) for the financial support given to this project. The authors would also like to acknowledge the support of the National Institutes of Health/National Cancer Institute grant 1R01-CA154444. The content of this paper is solely the responsibility of the authors and does not necessarily represent the official views of the funding agencies. We thank Real Time Tomography (RTT) for providing assistance with image reconstruction. ADAM is a member of the scientific advisory board and shareholder of RTT.
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2016 Springer International Publishing Switzerland
About this paper
Cite this paper
Borges, L.R., Guerrero, I., Bakic, P.R., Maidment, A.D.A., Schiabel, H., Vieira, M.A.C. (2016). Simulation of Dose Reduction in Digital Breast Tomosynthesis. In: Tingberg, A., Lång, K., Timberg, P. (eds) Breast Imaging. IWDM 2016. Lecture Notes in Computer Science(), vol 9699. Springer, Cham. https://doi.org/10.1007/978-3-319-41546-8_43
Download citation
DOI: https://doi.org/10.1007/978-3-319-41546-8_43
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-41545-1
Online ISBN: 978-3-319-41546-8
eBook Packages: Computer ScienceComputer Science (R0)