Abstract
A determination of the sensitometric parameters of screen-film systems to evaluate their qualities was performed. The quality control of the automatic film processor was carried out to ensure a high level of efficiency. Based on ISO 9236-3, the following potentials were applied on the X-ray tubes: 25 kV, 28 kV, 30 kV and 35 kV. Four different mammography films from different manufacturers with and without screens were tested for curve shape, speed and average gradient. The results indicated that film 1 exhibited better contrast, film 3 demonstrated the highest energy dependence, and film 4 presented the largest base+fog density. None of the four mammographic films tested achieved satisfactory results in all parameters analyzed. Improvements in the manufacturing process for these films must be completed to avoid losses in the image quality.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2015 Springer International Publishing Switzerland
About this paper
Cite this paper
Medeiros, L.L., Magalhaes, L.A.G., Estrada, J.J.S., Ferreira, N.M.P.D., de Almeida, C.E. (2015). Sensitometric analyses of screen-film systems for mammography exams in Brazil. In: Jaffray, D. (eds) World Congress on Medical Physics and Biomedical Engineering, June 7-12, 2015, Toronto, Canada. IFMBE Proceedings, vol 51. Springer, Cham. https://doi.org/10.1007/978-3-319-19387-8_4
Download citation
DOI: https://doi.org/10.1007/978-3-319-19387-8_4
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-19386-1
Online ISBN: 978-3-319-19387-8
eBook Packages: EngineeringEngineering (R0)