Abstract
Quality assurance is designed to ensure that radiation is delivered correctly. AAPM TG-40 and TG-142 recommendations set a standard of practice for radiation oncology practices. Machine characteristics are measured on a daily, monthly and annual basis, and compared to tolerance levels (unacceptable in the long term) and action levels (unacceptable at any time). Different devices are used to measure these QA checks. Daily checks can use less precise but less cumbersome instruments, while monthly and annual checks use more precise instruments.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Author information
Authors and Affiliations
Rights and permissions
Copyright information
© 2014 Springer Science+Business Media New York
About this chapter
Cite this chapter
Chang, D.S., Lasley, F.D., Das, I.J., Mendonca, M.S., Dynlacht, J.R. (2014). Linac Quality Assurance. In: Basic Radiotherapy Physics and Biology. Springer, Cham. https://doi.org/10.1007/978-3-319-06841-1_13
Download citation
DOI: https://doi.org/10.1007/978-3-319-06841-1_13
Published:
Publisher Name: Springer, Cham
Print ISBN: 978-3-319-06840-4
Online ISBN: 978-3-319-06841-1
eBook Packages: MedicineMedicine (R0)