Abstract
Spin Echo pulse sequences follow similar contrast behavior to that described in the previous section; however, there are some key differences between Gradient Echo and Spin Echo pulse sequences (Fig. 11.1). In the case of spin echo, a 180° refocusing pulse is applied at a time equal to half the echo time. This reverses the effect of the field inhomogeneities and the amplitude of the transverse magnetization at the echo times dependent on the pure T2 relaxation time of the tissue. As the T2 values for tissues are longer than the T2* values, the echo times chosen to achieve T2 weighting with Spin Echo are also longer than the echo times required to achieve T2* weighting with gradient echo sequences. The spin echo pulse sequence uses an excitation pulse that is normally 90°, so at each excitation pulse all of the z magnetization is transferred into the transverse plane. This combined with the refocusing pulse gives the largest possible signal provided the magnetization is allowed to recover sufficiently between repetitions.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Author information
Authors and Affiliations
Rights and permissions
Copyright information
© 2010 Springer-Verlag London Limited
About this chapter
Cite this chapter
Plein, S., Greenwood, J.P., Ridgway, J.P. (2010). Gradient Echo Versus Spin Echo. In: Cardiovascular MR Manual. Springer, London. https://doi.org/10.1007/978-1-84996-362-6_11
Download citation
DOI: https://doi.org/10.1007/978-1-84996-362-6_11
Published:
Publisher Name: Springer, London
Print ISBN: 978-1-84996-361-9
Online ISBN: 978-1-84996-362-6
eBook Packages: MedicineMedicine (R0)