Implementation and Clinical Use of Portal Imaging

  • Cynthia L. Thomason
Part of the Cancer Treatment and Research book series (CTAR, volume 93)

Abstract

Although introduced recently, on-line electronic portal imaging devices (EPIDs) are beginning to have an impact on radiation therapy, primarily in geometric verification of treatment fields. Using an EPID, a digital electronic portal image (EPI) can be obtained within a few seconds and be available for immediate review prior to completion of treatment. With the advent of more sophisticated techniques, such as conformal therapy and three-dimensional (3D) localization and treatment planning using computed tomography (CT) and magnetic resonance imaging (MRI), patient setup verification is perhaps the weak link in the planning and treatment process. Accuracy and precision of patient positioning is crucial to ensure delivery of the maximum dose to the target volume while avoiding normal and critical structures. Electronic portal imaging (EPIG) has the potential to strengthen that link at a lower cost than could be achieved with film. While it may not be possible to totally eliminate setup errors, effective use of an EPID will minimize both random and systematic errors, which may result in the reduction of the dose to the target volume and/or an overdose to normal critical structures.

Keywords

Reference Image Setup Error Portal Image Electronic Portal Imaging Device Patient Setup 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    AAPM Task Group No. 28. 1987. AAPM Report No. 24: Radiotherapy Portal Imaging Quality. American Institute of Physics.Google Scholar
  2. 2.
    Photon Treatment Planning Collaborative Working Group. 1991. Three-dimensional display in planning radiation therapy: A clinical perspective. Int J Radiat Oncol Biol Phys 21: 79–89.Google Scholar
  3. 3.
    Rabinowitz I, Broomberg J, Goitein M, McCarthy K, Leong J. 1985. Accuracy of radiation field alignment in clinical practice. Int J Radiat Oncol Biol Phys 11:1857–1867.PubMedGoogle Scholar
  4. 4.
    Byhardt RW, Cox JD, Hornburg A, Lierman G. 1978. Weekly localization films and detection of field placement errors. Int J Radiat Oncol Biol Phys 4:881–887PubMedGoogle Scholar
  5. 5.
    Marks JE, Haus AG, Sutton HH, Griem ML. 1974. Localization error in the radiotherapy of Hodgkin’s disease and malignant lymphoma with extended mantle fields. Cancer 34:83–90.PubMedGoogle Scholar
  6. 6.
    Kinzie JJ, Hanks HE, Maclean CE, Kramer S. 1983. Patterns of care study: Hodgkin’s disease relapse rate and adequacy of portals. Cancer 52: 2223–2226.PubMedGoogle Scholar
  7. 7.
    Taylor BW, Mendenhall NP, Million RR. 1990. Reproducibility of mantle irradiation with daily imaging films. Int J Radiat Oncol Biol Phys 19:149–151.PubMedGoogle Scholar
  8. 8.
    Doss LL. 1979. Localization error and local recurrence in upper airway carcinoma. In Reinstein LE, ed. Proceedings of the Workshop on Quality Control in Radiotherapy Departments. New York:Z CALGB.Google Scholar
  9. 9.
    Niederer J, Hawkins NV, Rider WD, Till JE. 1976. Failure of radical radiotherapy of supraglottic laryngeal carcinoma. Int J Radiat Oncol Biol Phys 2:612–629.Google Scholar
  10. 10.
    White JE, Chen T. 1982. The influence of radiotherapy quality control on survival, response and sites of relapse in oat cell carcinoma of lung: Preliminary report of the Southwestern Oncology Group study. Cancer 50:1084–1090.PubMedGoogle Scholar
  11. 11.
    Leong J, Shimm D. 1985. A method for consistent precision radiation therapy. Radiother Oncol 3:89–92.PubMedGoogle Scholar
  12. 12.
    Marks JE, Haus AG, Sutton HG, Griem ML. 1976. The value of frequent treatment verification films in reducing localization error in the irradiation of complex fields. Cancer 37:2755–2761.PubMedGoogle Scholar
  13. 13.
    Ezz A, Munro P, Porter A, Battista J, Jaffray D, Fenster A, Osborne S. 1992. Daily monitoring and correction of radiation field placement using a video based portal imaging system: A pilot study. Int J Radiat Oncol Biol Phys 22:159–165.PubMedGoogle Scholar
  14. 14.
    De Neve W, Van den Heuvel F, DeBeukeleer M, Coghe M, Thon L, DeRoover P, Van Lancker M, Storme G. 1992. Routine clinical on-line portal imaging followed by immediate field adjustment using a tele-controlled patient couch. Radiother Oncol 24:45–54.PubMedGoogle Scholar
  15. 15.
    Van den Heuvel F, De Neve W, Verellen D, Coghe M, Coen V, Storme G. 1995. Clinical implementation for intra-treatment correction using electronic portal imaging, Radiother Oncol 35:232–239.PubMedGoogle Scholar
  16. 16.
    Cionini L, Bucciolini M. 1993. Role of portal imaging in clinical radiotherapy: Florence experience. Radiother Oncol 29:230–236.PubMedGoogle Scholar
  17. 17.
    Urie MM, Goitein M, Doppke K, Kutcher JG, LoSasso T, Mohan R, Munzenrider JE, Sontag M, Wong JW. 1991. The role of uncertainty analysis in treatment planning. Int J Radiat Oncol Biol Phys 21:91–107.PubMedGoogle Scholar
  18. 18.
    Kesteloot K, Dutreix A, van der Schueren E. 1993. A model for calculating the costs of in vivo dosimetry and portal imaging in radiotherapy departments. Radiother Oncol 28: 108–117.PubMedGoogle Scholar
  19. 19.
    Shalev S. 1996. Megavoltage portal imaging. In Palta JR, Mackie TR, eds. Teletherapy Physics, Present and Future. Advanced Medical Publishing.Google Scholar
  20. 20.
    Herman MG, Abrams RA, Mayer RR. 1993. Clinical use of on-line portal imaging for daily patient treatment verification. Int J Radiat Oncol Biol Phys 28:1017–1023.Google Scholar
  21. 21.
    Thomason C, MacGowan T. 1994. Comparison of three commercially available portal imagers. Med Phys 21:888.Google Scholar
  22. 22.
    Wade J, Nicholas D. 1991. Clinical application of an electronic imaging device for assisting patient set-up in radiotherapy. Br J Radiol 64:596–602.PubMedGoogle Scholar
  23. 23.
    Crooks I, Fallone BG. 1993. A novel algorithm for the edge detection and edge enhancement of medical images. Med Phys 20:993–998.PubMedGoogle Scholar
  24. 24.
    Wowk B, Shalev S, Radcliffe T. 1993. Grooved phosphor screens for on-line portal imaging. Med Phys 20:1641.PubMedGoogle Scholar
  25. 25.
    Jaffray DA, Battista JJ, Fenster A, Munro P. 1994. X-ray scatter in megavoltage transmission radiography: Physical characteristics and influence on image quality. Med Phys 21:45–60.PubMedGoogle Scholar
  26. 26.
    Munro P, Rawlinson JA. 1990. Therapy imaging: A signal-to-noise analysis of a fluoroscopic imaging system for radiotherapy localization. Med Phys 17:763–772.PubMedGoogle Scholar
  27. 27.
    Jaffray DA, Battista JJ, Fenster A, Munro P. 1995. Monte Carlo studies of x-ray energy absorption and quantum noise in megavoltage transmission radiography. Med Phys 22:1077–1086.PubMedGoogle Scholar
  28. 28.
    Bissonnette JP, Munro P. 1996. Evaluation of a high-density scintillating glass for portal imaging. Med Phys 23:401–406.PubMedGoogle Scholar
  29. 29.
    Swindell W, Evans PM. 1996. Scattered radiation in portal images: A Monte Carlo simulation and a simple physical model. Med Phys 23:63–73.PubMedGoogle Scholar
  30. 30.
    Leszczynski K, Shalev S, Ryder S. 1992. A study on the efficacy of digital enhancement of on line portal images. Med Phys 19:999–1005.PubMedGoogle Scholar
  31. 31.
    Griffiths SE, Pearcey RG. 1986. The daily reproducibility of large, complex shaped radiotherapy fields to the thorax, head and neck. Clin Radiol 37:39–41.PubMedGoogle Scholar
  32. 32.
    Pearcey RG, Griffiths SE. 1985. The impact of treatment errors on post operative radiotherapy for testicular tumors, Br J Radiol 58:1003–1005.PubMedGoogle Scholar
  33. 33.
    Hunt MA, Kutcher GJ, Burman C, Fass D, Harrison L, Leibel S, Fuks Z. 1993. The effect of setup uncertainties on the treatment of nasopharynx cancer. Int J Radiat Oncol Biol Phys 27:437–447.PubMedGoogle Scholar
  34. 34.
    Schewe JE, Balter JM, Lam KL, Ten Haken RK. 1996. Measurement of patient setup errors using port films and a computer-aided graphical alignment tool. Med Dosim 21:97–104.PubMedGoogle Scholar
  35. 35.
    Balter J, Pelizzari CA, Vijayakumar S, Chen GTY. 1990. Computer aided registration of simulation and portal images: An exploratory clinical study. Int J Radiat Oncol Biol Phys 19:86–87.Google Scholar
  36. 36.
    Bel A, van Herk M, Bartelink H, Lebesque JL. 1993. A verification procedure to improve patient set-up accuracy using portal images. Radiother Oncol 29:253–260.PubMedGoogle Scholar
  37. 37.
    Creutzberg CL, Althof BGM, Huizenga H, Visser AG, Levendag P. 1993. Quality assurance using portal imaging: The accuracy of patient positioning in irradiation in breast cancer. Int J Radiat Oncol Biol Phys 25:529–539.PubMedGoogle Scholar
  38. 38.
    De Neve, Van den Heuvel F, Coghe M, Verellen D, deBeukeleer M, Roelstraete A, de Roover P, Thon L, Storme G. 1993. Interactive use on on-line portal imaging in pelvic radiation. Int J Radiat Oncol Biol Phys 25:517–524.Google Scholar
  39. 39.
    El-Gayed A, Bel A, Vijibrief R, Barteling H, Lebesque J. 1993. Time trend of patient setup deviation during pelvic irradiation using electronic portal imaging. Radiother Oncol 26: 162–171.PubMedGoogle Scholar
  40. 40.
    De Neve W, Van den Heuvel F, De Beukeleer M, Coghe M, Thon L, De Roover P, Van Lancker M, Storme G. 1992. Routine clinical on-line portal imaging followed by immediate field adjustment using a telecontrolled patient couch. Radiother Oncol 24:45–54.PubMedGoogle Scholar
  41. 41.
    Michalski JM, Wong JW, Gerber RL, Yan D, Chen A, Graham MV, Renna MA, Sawyer PJ, Perez CA. 1993. The use of on line image verification to estimate the variation in radiation therapy dose delivery. Int J Radiat Oncol Biol Phys 27:707–716.PubMedGoogle Scholar
  42. 42.
    Leong JC, Stracher MA. 1987. Visualization of internal motion within a treatment portal during a radiation therapy treatment. Radiother Oncol 9:153–156.PubMedGoogle Scholar
  43. 43.
    De Neve W, Van den Heuvel F, Coghe M, Verellen D, De Beukeleer M, Roelstraet A, DeRoover P, Thon L, Storme G. 1993. Interactive use of on line portal imaging in pelvic radiation. Int J Radiat Oncol Biol Phys 25: 517–524.PubMedGoogle Scholar
  44. 44.
    Gildersleve J, Dearnaley DP, Evans PM, Law M, Rawlings C, Swindell W. 1994. A randomized trial of patient repositioning during radiotherapy using a megavoltage imaging system. Radiother Oncol 31:161–168.PubMedGoogle Scholar
  45. 45.
    Balter JM, Chen GT, Pelizzari CA, Krishansamy S, Rubin S, Vijayakumar S. 1993. Online repositioning during treatment of the prostate: A study of potential limits and gains. Int J Radiat Oncol Biol Phys 27:137–143.PubMedGoogle Scholar
  46. 46.
    Bijhold J, Lebesque JV, Hart AA, Vijibrief RE. 1992. Maximizing setup accuracy using portal images as applied to conformai boost techniques for prostatic cancer. Radiother Oncol 24:261–271.PubMedGoogle Scholar
  47. 47.
    Verellen D, De Neve W, Van den Heuvel F, Coghe M, Louis O, Storme G. 1993. On-line portal imaging: Image quality defining parameters for pelvic fields — a clinical evaluation. Int J Radiat Oncol Phys 27:945–952.Google Scholar
  48. 48.
    Dunscombe P, Loose S, Leszczynski K. 1993. Sizes and sources of field placement error in routine irradiation for prostate cancer. Radiother Oncol 26:174–176.PubMedGoogle Scholar
  49. 49.
    Halverson K, Leung T, Pellet J, Gerber R, Weinhous M, Wong J. 1991. Study of treatment variation in the radiotherapy of head and neck tumors using a fiber-optic on-line radiotherapy imaging system. Int J Radiat Oncol Biol Phys 21:1327–1335.PubMedGoogle Scholar
  50. 50.
    Huizenga H, Levendag PC, De Porre PMZR, Visser AG. 1988. Accuracy in radiation field alignment in head and neck cancer: A prospective study. Radiother Oncol 11:181–187.PubMedGoogle Scholar
  51. 51.
    Bissett R, Leszczynski K, Loose S, Boyko S, Dunscombe P. 1996. Quantitative versus subjective portal verification using digital portal images. Int J Radiat Oncol Biol Phys 34:489–495.PubMedGoogle Scholar
  52. 52.
    Creutzberg CL, Visser AG, DePorre PMZR, Meerwaldt JH, Althof VGM, Levendag PC. 1992. Accuracy of patient positioning in mantle field irradiation. Radiother Oncol 23:257–264.PubMedGoogle Scholar
  53. 53.
    Baily NA, Horn RA, Kampp TD. 1980. Fluo-roscopic visualization of megavoltage therapeutic x-ray beams. Int J Radiat Oncol Biol Phys 6:935–939.PubMedGoogle Scholar
  54. 54.
    Boyer AL, Antonuk L, Fenster A, van Herk M, Meertens H, Munro P, Reinstein LE, Wong J. 1992. A review of electronic portal imaging devices (EPIDs). Med Phys 19:1–16.PubMedGoogle Scholar
  55. 55.
    Leong J. 1986. Use of digital fluoroscopy as an on-line verification device in radiation therapy. Phys Med Biol 31:985–992.PubMedGoogle Scholar
  56. 56.
    InfiMed, Inc., Liverpool, NY 13088.Google Scholar
  57. 57.
    Elekta Oncology Systems, Crawley, England.Google Scholar
  58. 58.
    Siemens Medical Systems, Inc., Concord, CA.Google Scholar
  59. 59.
    Munro P, Rawlinson JA, Fenster A. 1990. A digital fluoroscopic imaging device for radiotherapy localization. Int J Radiat Oncol Biol Phys 18:641–649.PubMedGoogle Scholar
  60. 60.
    Heijmen BJM, Stroom JS, Hiuzenga H, Visser AG. 1993. Application of a fluoroscopic portal imaging system with a CCD camera for accurate in vivo dosimetry. Med Phys 20:879.Google Scholar
  61. 61.
    Visser AG, Huizenga H, Althof VGM, Swanenburg. 1990. Performance of a prototype fluoroscopic radiotherapy imaging system. Int J Radiat Oncol Biol Phys 18:43–50.PubMedGoogle Scholar
  62. 62.
    Shalev S, Lee T, Leszczynski K, Cosby S, Chu T, Reinstein L, Meek A. 1989. Video techniques for on-line portal imaging. Comp Med Imag Graphics 13:217–226.Google Scholar
  63. 63.
    Varian Oncology Systems, Palo Alto, CA 94304.Google Scholar
  64. 64.
    Meertens H, van Herk M, Weeda J. 1985. A liquid ionisation detector for digital radiography of therapeutic megavoltage photon beams. Phys Med Biol 30:313–321.PubMedGoogle Scholar
  65. 65.
    van Herk M, Meertens H. 1988. A matrix ionization chamber imaging device for on line patient setup verification during radiotherapy. Radiother Oncol 11:369–378.PubMedGoogle Scholar
  66. 66.
    van Herk M. 1991. Physical aspects of a liquid filled ionization chamber with pulsed polarizing voltage. Med Phys 18:692–702.PubMedGoogle Scholar
  67. 67.
    Swindell W. 1991. The lens coupling efficiency in megavoltage imaging. Med Phys 18:1152–1153.PubMedGoogle Scholar
  68. 68.
    Shalev S. 1996. On-line portal imaging: Contributions and limitations in clinical practice. In Meyer JL, Purdy JA, eds. Three-dimensional conformai therapy. Frontiers in Radiation Therapy and Oncology. Basel: Karger, pp. 156–167.Google Scholar
  69. 69.
    Lutz WR, Bjarngard BE. 1985. A test object for resolution of portal films. Int J Radiat Oncol Biol Phys 11:631–634.PubMedGoogle Scholar
  70. 70.
    Shalev S. 1995. Treatment verification using digital imaging. In Smith A, ed. Radiation Therapy Physics. Berlin: Springer-Verlag, pp. 155–173.Google Scholar
  71. 71.
    Shalev S, McGee K. 1992. Evaluation of image processing techniques by ROC analysis. Med Phys 19:822.Google Scholar
  72. 72.
    Cumberlin R, Rodgers J, Fahey F. 1989. Digital image processing of radiation therapy portal films. Comput Med Imaging Graph 13:227–233.PubMedGoogle Scholar
  73. 73.
    Low DA, Klein EE, Maag DK, Umfleet WE, Purdy JA. 1996. Commissioning and periodic quality assurance of a clinical electronic portal imaging device. Int J Radiat Oncol Biol Phys 34:117–123.PubMedGoogle Scholar
  74. 74.
    Amols HI, Reinstein LE, Lagueux B. 1986. A quantitative assessment of portal film contrast as a function of beam energy. Med Phys 13:711–716.PubMedGoogle Scholar
  75. 75.
    Santos A, Perez J, Liso F, Carmona V, Baumann G, Bermisz M, Tormo A, Petschen I. 1996. On the use of a Philips radiotherapy imaging system in a radiotherapy department. Proceedings of the 4th International Workshop on Electronic Portal Imaging 9.Google Scholar
  76. 76.
    Rajapakshe R, Radcliffe TJ, Shalev S. 1993. Quality control for on-line portal imaging systems. Proc. SPIE 1897:121–127.Google Scholar
  77. 77.
    Dong L, Boyer AL. 1994. An objective method for evaluation electronic portal imaging devices. Med Phys 21:755–760.PubMedGoogle Scholar
  78. 78.
    Harrison, RM, Lambert GD, Cahapple CL. 1993. Spectral estimation and contrast calculation in the design of contrast-detail test objects for radiotherapy portal imaging. Phys Med Biol 38:545–556.Google Scholar
  79. 79.
    Rajapakshe R, Luchka K, Shalev S. 1996. A quality control test for electronic portal imaging devices. Med Phys 23:1237–1244.PubMedGoogle Scholar
  80. 80.
    Task Group No. 58, Electronic Portal Imaging Devices, JW Wong, Chair. American Association of Physicists in Medicine.Google Scholar
  81. 81.
    Shalev S. 1989. The Las Vegas Phantom design developed for the AAPM spring seminar, Las Vegas, Nevada.Google Scholar
  82. 82.
    J Wong, personal communication.Google Scholar
  83. 83.
    IFF™, Aldus Corporation, public domain format.Google Scholar
  84. 84.
    Microsoft PowerPoint™, Microsoft Corporation.Google Scholar
  85. 85.
    Corel Draw™, Corel Systems Corporation.Google Scholar
  86. 86.
    LView Pro l.B, Leonardo Haddad Loureiro, distributed by Media Research Corporation.Google Scholar
  87. 87.
    IMPAC Medical Systems, Inc., Mountain View, CA 94041-1203.Google Scholar
  88. 88.
    Reinstein LE, Pai S, Meek AG. 1992. Assessment of geometric treatment accouracy using time-lapse display of electronic portal images. Int J Radiat Oncol Biol Phys 22:1139–1146.PubMedGoogle Scholar
  89. 89.
    Pouliot J, Lirette A, Vigneault W. 1995. Dynamical aspects of real-time portal imaging. Med Phys 22:661.Google Scholar
  90. 90.
    Weinhous M. 1990. Treatment verification using a computer workstation. Int J Radiat Oncol Biol Phys 19:1549–1554.PubMedGoogle Scholar
  91. 91.
    Bijhold J, van Herk M, Vijlbrief R, Lebesque J. 1991. Fast evaluation of patient set-up during radiotherapy by aligning features in portal and simulator images. Phys Med Biol 36:1665–1679.PubMedGoogle Scholar
  92. 92.
    Graham M, Cheng A, Geer L, Binns W, Vannier M, Wong J. 1991. A method to analyze 2-dimensional daily radiotherapy portal images from an on-line fiber-optic imaging system. Int J Radiat Oncol Biol Phys 20:613–619.PubMedGoogle Scholar
  93. 93.
    Evans P, Gildersleve J, Morton E, Swindell W, Coles R, Ferraro M, Rawlings C, Xiao Z, Dyner J. 1992. Image comparison techniques for use with megavoltage imaging systems. Br J Radiol 65:701–709.PubMedGoogle Scholar
  94. 94.
    Lam W, Herman M, Lam K, Lee D. 1991. Online portal imaging: Computer assisted error measurements. Radiology 179:871–873.PubMedGoogle Scholar
  95. 95.
    Meertens H, Bijhold J, Strackee J. 1990. A method for the measurement of field placement errors in digital portal images. Phys Med Biol 35:299–323.PubMedGoogle Scholar
  96. 96.
    Boesecke R, Bruckner T, Ende G. 1990. Landmark based correlation of medical images. Phys Med Biol 35:121–126.Google Scholar
  97. 97.
    Kessler M, Pitluck S, Petti P, Castro J. 1991. Integration of multimodality imaging data for radiotherapy treatment planning. Int J Radiat Oncol Biol Phys 21:1652–1667.Google Scholar
  98. 98.
    McParland BJ. 1993. Uncertainty analysis of field placement error measurements uring digital portal and simulation image correlations. Med Phys 20:679–685.PubMedGoogle Scholar
  99. 99.
    Lam KL, Ten Haken RK, McShan DL, Thornton AF. 1993. Automated determination of patient setup errors in radiation therapy using spherical radio-opaque markers. Med Phys 20:1145–1152.PubMedGoogle Scholar
  100. 100.
    Thomason C. 1993. Limitations and usability of comparison algorithms found on commercially available electronic portal imaging devices. Med Phys 20:922.Google Scholar
  101. 101.
    Gilhuijs KGA, van Herk M. 1993. Automatic on line inspection of patient setup in radiation therapy using digital portal images. Med Phys 20:667–677.PubMedGoogle Scholar
  102. 102.
    Balter J, Pelizzari C, Chen T. 1992. Correlation of projection radiographs in radiation therapy using open curve segments and points. Med Phys 19:329–334.PubMedGoogle Scholar
  103. 103.
    Jones S, Boyer A. 1991. Investigation of an FFT-based correlation technique for verification of treatment set-up. Med Phys 19:1116–1125.Google Scholar
  104. 104.
    Munro P, Towner J, Jaffray D, Battista J, Fenster A. 1991. A semi-automated method of aligning portal images. Med Phys 18: 850.Google Scholar
  105. 105.
    Boyer A, Dong G, Starkschall G, Hogstrom K. 1992. A portal image correlation procedure. Med Phys 19:802.Google Scholar
  106. 106.
    Moseley J, Munro P. 1994. A semiautomatic method for registration of portal images. Med Phys 21:551–558.PubMedGoogle Scholar
  107. 107.
    Wang H, Fallone BG. 1994. A robust morphological algorithm for automatic radiation field extraction and correlation of portal images. Med Phys 21:237–244.PubMedGoogle Scholar
  108. 108.
    Borgefors G. 1988. Heirarchial chamfer matching: A parameter edge matching algorithm. IEEE Trans PAMI 10:849–865.Google Scholar
  109. 109.
    Gilhuijs KGA, El-Gayed AAH, van Herk M, Vijlbrief RE. 1993. An algorithm for automatic analysis of portal images: Clinical evaluation for prostate treatments. Radiother Oncol 29:261–268.PubMedGoogle Scholar
  110. 110.
    Fritsch DL, Chaney EL. 1992. Computer-automated registration of on line radiotherapy portal images. Med Phys 19:1134.Google Scholar
  111. 111.
    van Herk M, Gilhuijs K. 1992. A comprehensive system for the interactive and automatic analysis of digital portal images. Med Phys 19:802.Google Scholar
  112. 112.
    Lesczynski K, Shalev S, Crosby S. 1992. The enhancement of radiotherapy verification images by an automated edge detection technique. Med Phys 19:611–622.Google Scholar
  113. 113.
    Gilhuijs K, Touw A, van Herk M, Vijlbrief RE. 1995. Optimization of automatic portal image analysis. Med Phys 22:1089–1099.PubMedGoogle Scholar
  114. 114.
    Bijhold J, Gilhuijs K, van Herk M. 1992. Automatic verification of radiation field shape using digital portal images. Med Phys 19: 1007–1014.PubMedGoogle Scholar
  115. 115.
    Eilertsen K, Skretting A, Tennvassas TL. 1994. Methods for fully automated verification of patient set-up in external beam radiotherapy with polygon shaped fields. Phys Med Biol 39:993–1012.PubMedGoogle Scholar
  116. 116.
    Fritsch DS, Chane EL, Boxwala A, McAuliffe A, Raghavan S, Thall A, Earnhard JRD. 1995. Core-based portal image registration for automatic radiotherapy treatment verification. Int J Radiat Oncol Biol Phys 33:1287–1300.PubMedGoogle Scholar
  117. 117.
    van Herk M, Gilhuijs K, Williams P, Cinquin P, Cionini L, Swindell W. 1993. Role of electronic portal imaging in high dose/high prescision radiotherapy. Radiother Oncol 29: 269–270.Google Scholar
  118. 118.
    P Dunscombe, Fox K. 1992. Portal Radiography practice in Canada. Int J Radiat Oncol Biol Phys 24:177–179.PubMedGoogle Scholar
  119. 119.
    Denham JW, Dally MJ, Hunter K, Wheat J, Fahey PP, Hamilton CS. 1993. Objective decision making following a portal film: The results of a pilot study. Int J Radiat Oncol Biol Phys 26:869–876.PubMedGoogle Scholar
  120. 120.
    Mitine C, Dutreix A, van der Schueren E. 1993. Black and white in accuracy assessment of megavoltage images: The medical decision is often grey. Radiother Oncol 28:31–36.PubMedGoogle Scholar
  121. 121.
    Dong L, Boyer AL. 1996. A portal image alignment and patient setup verification procedure using moments and correlation techniques. Phys Med Biol 41:697–723.PubMedGoogle Scholar
  122. 122.
    Hristov DH, Fallone BG. 1996. A grey-level image alignment algorithm for registration of portal images and digitally reconstructed radiographs. Med Phys 23:75–84.PubMedGoogle Scholar
  123. 123.
    McParland BJ, Kumaradas JC. 1995. Digital portal image registration by sequential anatomical matchpoint and image correlations for real time continuous field alignment verification. Med Phys 22:1063–1075.PubMedGoogle Scholar
  124. 124.
    Dunscombe PB, Fox, K, Loose S, Leszczynski. 1993. The investigation and rectification of field placement errors in the delivery of complex head and neck fields. Int J Radiat Oncol BiolPhys 26:155–161.Google Scholar
  125. 125.
    Leong JC, Stacher MA. 1987. Visualization of internal motion within a treatment portal during a radiation therapy treatment. Radiother Oncol 9:153–156.PubMedGoogle Scholar
  126. 126.
    Gladstone DJ, van Herk M, Chin LM. 1993. Verification of lung attenuator positioning before total body irradiation using an electronic portal imaging device. Int J Radiat Oncol Biol Phys 27:449–454.PubMedGoogle Scholar
  127. 127.
    Holmberg O, Huizenga H, Idzes MHM, Lebesque JV, Vijlbrief RE, Mijnherr BJ. 1994. In vivo determination of the accuracy of field matching in breast cancer irradiation using an electronic portal imaging device. Radioth Oncol 33:157–166.Google Scholar
  128. 128.
    Zhou A-M, Verhy LJ. 1994. A robust way for multileaf collimator (MLC) leaf position verification. Med Phys 922.Google Scholar
  129. 129.
    Alford B. InfiMed, Inc., personal communication.Google Scholar
  130. 130.
    Shalev S. 1994. Progress in the evaluation of electronic portal imaging — taking one step at a time. Int J Radiat Oncol Biol Phys 28:1043–1045.PubMedGoogle Scholar
  131. 131.
    Haus AG, Strubier K, Marks JE. 1972. A film technique for determining the absorbed dose at exit from the patient during radiation exposure. Radiology 104:197–200.PubMedGoogle Scholar
  132. 132.
    Haus AG, Marks JE. 1976. Film techniques in radiotherapy for treatment verification, determination of patient exit dose, and detection of localization error. J Appl Photogr Eng 2:11–16.Google Scholar
  133. 133.
    Ying XG, Geer LY, Wong JW. 1990. Portal dose images. II. Patient dose estimation. Int Radiat Oncol Biol Phys 18:1465–1475.Google Scholar
  134. 134.
    Zhu, Y, Jiang X-Q, Van Dyke J. 1995. Portal dosimetry using a liquid ion chamber matrix: Dose response studies. Med Phys 22:1101–1106.PubMedGoogle Scholar
  135. 135.
    Weltens C, van Dam J, Geunens G, Dutreix A, van der Schueren E. 1994. Reliability of clinical port films for measuring dose inhomogeneities in radiotherapy for head and neck tumors. Radiother Oncol 30:167–170.PubMedGoogle Scholar
  136. 136.
    Van Dam J, Vaerman C, Blanckaert N, Leunens G, Dutreix A, Van der Schueren E. 1992. Are port films reliable for in vivo exit does measurements? Radiother Oncol 25:67–72.PubMedGoogle Scholar
  137. 137.
    Fiorino C, del Veccchio A, Cattaneo GM, Fusca M, Longobardi B, Signorotto P, Calandrino R. 1993. Exit dose measurements by portal film dosimetry. Radiother Oncol 29:336–340.PubMedGoogle Scholar
  138. 138.
    Fiorino C, Cattaneo GM, Del vechio A, Fusca M, Longobardi B, Signorotto P, Uleri C, Calnadrino R. 1994. Estimation of midline does distributions by using portal film exit dosimetry data. Med Phys 21:964.Google Scholar
  139. 139.
    Heijmen BJM, Stroom JS, Huizenga H, Visser AG. 1993. Application of a fluoroscopic portal imaging system with a CCD camera for accurate in vivo dosimetry. Med Phys 20:870.Google Scholar
  140. 140.
    Roback DM, Gerbi BJ. 1995. Evaluation of electronic portal imaging device for missing tissue compensator design and verification. Med Phys 22:2029–2034.PubMedGoogle Scholar
  141. 141.
    Huyskens D, Goossens A, Mersserman B, Reymen R, Van Dam J, Dutreix A, van der Schueren E. 1996. About the estimation of exit doses using portal imaging: Redistribution of primary and scattered radiation behind the patient. Proceedings of the 4th International Workshop on Electronic Portal Imaging, 1996, p. 43.Google Scholar
  142. 142.
    Rajapakshe R, Parsaei H, Wolters J, El-Khatib. 1996. Exit dosimetry with electronic portal imaging devices. Proceedings of the 4th International Workshop on Electronic Portal Imaging, 1996, p. 49.Google Scholar
  143. 143.
    Yin F-F, Schell MC, Rubin P. 1994. Input/ output characteristics of a matrix ion-chamber electronic portal imaging device. Med Phys 21: 1447–1454.PubMedGoogle Scholar
  144. 144.
    Kirby MC, Williams PC. 1993. Measurement possibilities using an electronic portal imaging device. Radiother Oncol 29:237–243.PubMedGoogle Scholar
  145. 145.
    Wong J, Slessinger ED, Hermes RE, Offutt CJ, Roy T, Vannier MW. 1990. Portal dose images 1: Quantitative treatment plan verification. Int J Radiat Oncol Biol Phys 18:1455–1463.PubMedGoogle Scholar
  146. 146.
    Evans MDC, Schreiner LJ. 1992. A simple technique for film dosimetry. Radiother Oncol 23:265-167.Google Scholar
  147. 147.
    Herman MG, Nicholson M, Cong YS. 1996. Exit dose verification using an electronic portal imaging device. Proceedings of the 4th International Workshop on Electronic Portal Imaging, 1996, p. 50.Google Scholar
  148. 148.
    Renner WD, O’Conner TP, Bermudez NM. 1982. An electronic device for digitizing radio-therapy films for the construction of tissue compensators. Med Phys 9:910–916.PubMedGoogle Scholar
  149. 149.
    Dixon RL, Ekstrand KE, Ferree C. 1979. Compensating filter design using megavoltage radiography. Int J Radiat Oncol Biol Phys 5: 281–287.PubMedGoogle Scholar
  150. 150.
    Dirkx MLP, Kroonwijk M, de Boer JCJ, Heijmen BJM. 1995. Daily dosimetric quality control of the MM50 Racetrack Microtron using an electronic portal imaging device. Radio-ther Oncol 27:55–60.Google Scholar
  151. 151.
    Dong L, Shiu A, Tung S, Boyer A. 1995. Online verification of alignment using an electronic portal imaging device for linac-based stereotactic radiosurgery. Med Phys 22:978.Google Scholar
  152. 152.
    Luchka K, Chen D, Shalev S, Gluhchev G, Rajapakshe R. 1996. Assessing radiation and light field congruence with a video based electronic portal imaging device. Med Phys 23: 1245–1252.PubMedGoogle Scholar
  153. 153.
    Convery DJ, Rosenbloom ME. 1995. Treatment delivery accuracy in intensity-modulated conformai radiotherapy. Phys Med Biol 40: 979–999.PubMedGoogle Scholar
  154. 154.
    van Herk M, Boellaard R, Brugmans M, van Dalen A. 1996. Feasibility of the PortalVision system for verification of dynamic therapy. Proceedings of the 4th International Workshop on Electronic Portal Imaging, 1996, p. 53.Google Scholar
  155. 155.
    Michalski JM, Wong JW, Gerber R, Yan D, Cheng A, Graham MB, Renna MA, Sayer PJ, Perez CA. 1993. The use of on-line image verification to estimate the variation in radiation therapy dose delivery. Int J Radiat Oncol Biol Phys 27:707–716.PubMedGoogle Scholar
  156. 156.
    Gur D, Weiser JC, Deutsch M, Fuhrman CR, Gennari RC, Rosenthal MS. 1989. Verification of electron beam therapy with conventional and storage phosphor images: Preliminary results. Int J Radiat Oncol Biol Phys 17:1337–1340.PubMedGoogle Scholar
  157. 157.
    Keller BE. 1978. Electron-beam radiographs. Radiology 128:830–831.PubMedGoogle Scholar
  158. 158.
    Rajapakshe R, Luchka K, Shalev S. 1994. Online imaging of electron treatments. Proceedings of the 3rd International Workshop on Electronic Portal Imaging, Oct 7-8, San Francisco.Google Scholar
  159. 159.
    Lewis DG, Swindell W, Morton EJ, Evan PM, Xizo ZR. 1992. A megavoltage CT scanner for radiotherapy verification. Phys Med Biol 37:1985–1999.PubMedGoogle Scholar
  160. 160.
    Feldkamp LA, Davis LC, Dress JR. 1987. Practical cone beam algorithm. J Opt Soc Am 1:612–619.Google Scholar
  161. 161.
    Webb S, Sutcliffe J, Burkinshaw L, Horsham A. 1987. Tomographie reconstruction from experimentally obtained cone beam projections. IEEE Trans Med Imag 6:67–73.Google Scholar
  162. 162.
    Midgley SM, Dudson JF, Millar RM. 1996. A feasibility study for the use of megavoltage photons and a commercial electronic portal imaging area detector for beam geometry CT scanning to obtain 3D tomographic data sets of radiotherapy patients in the treatment position. Porceedings of the 4th International Workshop on Electronic Protal Imaging, 1996, p. 60.Google Scholar
  163. 163.
    Jaffray DA, Drake KG, Pan G, Wong JW. 1996. Conebeam CT using a clinical fluoroscopic portal imager. Proceedings of the 4th International Workshop on Electronic Portal Imaging, 1996, p. 59.Google Scholar
  164. 164.
    Antonuk LE, Boudry J, El-Mohri Y, Huang W, Siewerdsen J, Yorkston J, Street RA. 1994. A high resolution, high frame rate, flat panel TFT array for digital imaging. Proc SPIE 2163:118–128.Google Scholar
  165. 165.
    Antonuk LE, Boudry J, Huang W, McShan D, Morton EJ, Yorkston J, Longo MJ, Street RA. 1992. Demonstration of megavoltage and diagnostic x-ray imaging with hydrogenated amorphous silicon arrays. Med Phys 19:1455–1466.PubMedGoogle Scholar
  166. 166.
    Antonuk L, Yorkston J, Boudry J. 1991. Large area amorphous silicon photodiode arrays for radiotherapy and diagnostic imaging. Nucl Instr Methods Phys Res A310:450–464.Google Scholar
  167. 167.
    Antonuk LE, Boudry J, Yorkston J, Wild CF, Long MJ, Street RA. 1990. Radiation-damage studies of amorphous-silicon photodiode sensors for applications in radiotherapy x-ray imaging. Nucl Inst Methods A299:143–146.Google Scholar
  168. 168.
    Mackie TR, Holmes T, Swerdloff S, Reckwerdt P, Deasy JO, Yang J, Paliwal B, Kinsella T. 1993. Tomotherapy: A new concept for the delivery of dynamic conformai radiotherapy. Med Phys 20:1709–1719.PubMedGoogle Scholar
  169. 169.
    Wong JW, Yu CX, Yan D, Mazur A. 1993. On-line double exposure small field verification. Med Phys 20:871.Google Scholar

Copyright information

© Springer Science+Business Media New York 1998

Authors and Affiliations

  • Cynthia L. Thomason

There are no affiliations available

Personalised recommendations