Skip to main content

The Effect of Basic Assumptions on the Tissue Oxygen Saturation Value of Near Infrared Spectroscopy

  • Conference paper
  • First Online:
Oxygen Transport to Tissue XXXIV

Abstract

Tissue oxygen saturation (StO2), a potentially important parameter in clinical practice, can be measured by near infrared spectroscopy (NIRS). Various devices use the multi-distance approach based on the diffusion approximation of the radiative transport equation [1, 2]. When determining the absorption coefficient (μ a) by the slope over multiple distances a common assumption is to neglect μ a in the diffusion constant, or to assume the scattering coefficient \( ({\mu }_{\text{s}}{}^{\prime })\) to be constant over the wavelength. Also the water influence can be modeled by simply subtracting a water term from the absorption. This gives five approaches A1–A5. The aim was to test how these different methods influence the StO2 values. One data set of 30 newborn infants measured on the head and another of eight adults measured on the nondominant forearm were analyzed. The calculated average StO2 values measured on the head were (mean ± SD): A1: 79.99 ± 4.47%, A2: 81.44 ± 4.08%, A3: 84.77 ± 4.87%, A4: 85.69 ± 4.38%, and A5: 72.85 ± 4.81%. The StO2 values for the adult forearms are: A1: 58.14 ± 5.69%, A2: 73.85 ± 4.77%, A3: 58.99 ± 5.67%, A4: 74.21 ± 4.76%, and A5: 63.49 ± 5.11%. Our results indicate that StO2 depends strongly on the assumptions. Since StO2 is an absolute value, comparability between different studies is reduced if the assumptions of the algorithms are not published.

Andreas Jaakko Metz is the member of the Ph.D. Program imMed

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 169.00
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 219.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info
Hardcover Book
USD 249.99
Price excludes VAT (USA)
  • Durable hardcover edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

References

  1. Hueber DM, Fantini S, Cerussi AE et al (1999) New optical probe designs for absolute (self-calibrating) NIR tissue hemoglobin measurements. Proc SPIE 3597:618–631

    Article  CAS  Google Scholar 

  2. Matcher SJ, Kirkpatrick P, Nahid K et al (1995) Absolute quantification methods in tissue near infrared spectroscopy. Proc SPIE 2359:486–495

    Article  Google Scholar 

  3. van Bel F, Lemmers P, Naulaers G (2008) Monitoring neonatal regional cerebral oxygen saturation in clinical practice: value and pitfalls. Neonatology 94(4):237–244

    Article  Google Scholar 

  4. Wolf M, Greisen G (2009) Advances in near-infrared spectroscopy to study the brain of the preterm and term neonate. Clin Perinatol 36(4):807–834

    Article  Google Scholar 

  5. Suzuki S, Takasaki S, Ozaki T et al (1999) A tissue oxygenation monitor using NIR spatially resolved spectroscopy. Proc SPIE 3597:582–592

    Article  CAS  Google Scholar 

  6. McKeating EG, Monjardino JR, Signorini DF et al (1997) A comparison of the Invos 3100 and the Critikon 2020 near-infrared spectrophotometers as monitors of cerebral oxygenation. Anaesthesia 52(2):136–140

    Article  CAS  PubMed Central  Google Scholar 

  7. Thavasothy M, Broadhead M, Elwell C et al (2002) A comparison of cerebral oxygenation as measured by the NIRO 300 and the INVOS 5100 Near-Infrared Spectrophotometers. Anaesthesia 57(10):999–1006

    Article  CAS  PubMed Central  Google Scholar 

  8. Yoshitani K, Kawaguchi M, Tatsumi K et al (2002) A comparison of the INVOS 4100 and the NIRO 300 near-infrared spectrophotometers. Anesth Analg 94(3):586–590

    Article  CAS  PubMed Central  Google Scholar 

  9. Jenny C, Biallas M, Trajkovic I et al (2011) Reproducibility of cerebral tissue oxygenation saturation measurements by near infrared spectroscopy in newborn infants. J Biomed Opt 16(9):097004

    Article  PubMed Central  Google Scholar 

  10. Haensse D, Szabo P, Brown D et al (2005) New multichannel near infrared spectrophotometry system for functional studies of the brain in adults and neonates. Opt Express 13(12):4525–4538

    Article  PubMed Central  Google Scholar 

  11. Muehlemann T, Haensse D, Wolf M (2008) Wireless miniaturized in-vivo near infrared imaging. Opt Express 16(14):10323–10330

    Article  CAS  Google Scholar 

  12. Matcher SJ, Elwell CE, Cooper CE et al (1995) Performance comparison of several published tissue near-infrared spectroscopy algorithms. Anal Biochem 227(1):54–68

    Article  CAS  PubMed Central  Google Scholar 

  13. Matcher SJ, Cope M, Delpy DT (1997) In vivo measurements of the wavelength dependence of tissue-scattering coefficients between 760 and 900 nm measured with time-resolved spectroscopy. Appl Opt 36(1):386–396

    Article  CAS  PubMed Central  Google Scholar 

  14. Arri SJ, Muehlemann T, Biallas M et al (2011) Precision of cerebral oxygenation and hemoglobin concentration measurements in neonates measured by near-infrared spectroscopy. J Biomed Opt 16(4):047005

    Article  PubMed Central  Google Scholar 

  15. Wolf M, Keel M, Dietz V et al (1999) The influence of a clear layer on near-infrared spectrophotometry measurements using a liquid neonatal head phantom. Phys Med Biol 44(7):1743–1753

    Article  CAS  PubMed Central  Google Scholar 

Download references

Acknowledgments

This work was financially supported by the Zurich Center of Integrative Human Physiology (ZIHP), University of Zurich, Switzerland. The authors would like to thank Raphael Zimmermann for very helpful discussions.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Andreas Jaakko Metz .

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2013 Springer Science+Business Media New York

About this paper

Cite this paper

Metz, A.J., Biallas, M., Jenny, C., Muehlemann, T., Wolf, M. (2013). The Effect of Basic Assumptions on the Tissue Oxygen Saturation Value of Near Infrared Spectroscopy. In: Welch, W.J., Palm, F., Bruley, D.F., Harrison, D.K. (eds) Oxygen Transport to Tissue XXXIV. Advances in Experimental Medicine and Biology, vol 765. Springer, New York, NY. https://doi.org/10.1007/978-1-4614-4989-8_24

Download citation

Publish with us

Policies and ethics