Escape of a Leaflet from a ST. JUDE MEDICAL® Prosthesis in the Mitral Position

  • Chapter
Cardiac Valve Replacement
  • 49 Accesses

Abstract

Twenty-three months after mitral valve replacement for mitral stenosis with a ST. JUDE MEDICAL® heart valve prosthesis size 31, a 29-year-old man suddenly went into a profound cardiogenic shock and pulmonary edema due to escape of one of the leaflets from the prosthesis. At emergency operation the valve was replaced with another type of prosthesis. The escaped leaflet was retrieved from the abdominal aorta and was found to be intact. The cause of the escape was found to be a fracture in the pivot area. Further examination with optic and scanning electron microscopy (SEM) revealed an old fracture in the base of the pivot, covered with organic material, and another fracture at the top of the pivot area, not covered with organic material, and therefore probably very recent and responsible for the break-off of part of the area and subsequent escape of the leaflet. The cause of the primary fracture is unknown.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 129.00
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 169.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info
Hardcover Book
USD 169.99
Price excludes VAT (USA)
  • Durable hardcover edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Bowen TE, Tri TB, Wortham DS: Thrombosis of a St. Jude Medical tricuspid prosthesis. J Thorac Cardiovasc Surg 1981; 82: 257.

    PubMed  CAS  Google Scholar 

  2. Emery RW, Nicoloff DM: St. Jude Medical cardiac valve prosthesis. J Thorac Cardiovasc Surg 1979; 78: 269.

    PubMed  CAS  Google Scholar 

  3. Hehrlein RW, Gottwik M, Fraedrich G, et al: First clinical experience with a new all-pyrolytic carbon bileaflet heart valve prosthesis. J Thorac Cardiovasc Surg 1980; 79: 632.

    PubMed  CAS  Google Scholar 

  4. Nicoloff DM, Emery RW, Aron KV, et al: Clinical and hemodynamic results with St. Jude Medical cardiac valve prosthesis. J Thorac Cardiovasc Surg 1981; 82: 674.

    PubMed  CAS  Google Scholar 

  5. Palmquist WE, Nicoloff DM, Emery RW, et al: St. Jude Medical all-pyrolytic carbon heart valve. Presented at the 13th Annual Meeting of the Association for Advancement of Medical Instrumentation, 1978.

    Google Scholar 

  6. Ziemer G, Luhmer I, Oelert H, et al: Malfunction of a St. Jude Medical heart valve in mitral position. Ann Thorac Surg 1982; 33: 391.

    Article  PubMed  CAS  Google Scholar 

Download references

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 1985 Martinus Nijhoff Publishing, Boston

About this chapter

Cite this chapter

Hjelms, E. (1985). Escape of a Leaflet from a ST. JUDE MEDICAL® Prosthesis in the Mitral Position. In: Matloff, J.M. (eds) Cardiac Valve Replacement. Springer, Boston, MA. https://doi.org/10.1007/978-1-4613-2601-4_33

Download citation

  • DOI: https://doi.org/10.1007/978-1-4613-2601-4_33

  • Publisher Name: Springer, Boston, MA

  • Print ISBN: 978-1-4612-9629-4

  • Online ISBN: 978-1-4613-2601-4

  • eBook Packages: Springer Book Archive

Publish with us

Policies and ethics