Beam Optimization for Digital Mammography – II

  • Conference paper
Digital Mammography (IWDM 2006)

Abstract

Optimization of acquisition technique factors (target, filter, and kVp) in digital mammography is required for maximization of the image SNR, while minimizing patient dose. The goal of this study is to compare, for each of the major commercially available FFDM systems, the effect of various technique factors on image SNR and radiation dose for a range of breast thickness and tissue types. This phantom study follows the approach of an earlier investigation [1], and includes measurements on recent versions of two of the FFDM systems discussed in that paper, as well as on three FFDM systems not available at that time. The five commercial FFDM systems tested are located at five different university test sites and include all FFDM systems that are currently FDA approved. Performance was assessed using 9 different phantom types (three compressed thicknesses, and three tissue composition types) using all available x-ray target and filter combinations. The figure of merit (FOM) used to compare technique factors is the ratio of the square of the image SNR to the mean glandular dose (MGD). This FOM has been used previously by others in mammographic beam optimization studies [2],[3]. For selected examples, data are presented describing the change in SNR, MGD, and FOM with changing kVp, as well as with changing target and/or filter type. For all nine breast types the target/filter/kVp combination resulting in the highest FOM value is presented. Our results suggest that in general, technique combinations resulting in higher energy beams resulted in higher FOM values, for nearly all breast types.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 84.99
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 109.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. Williams, M.B., More, M.J., Venkatakrishnan, V., Niklason, L., Yaffe, M.J., Mawdsley, G., Bloomquist, A., Maidment, A.D.A., Chakraborty, D., Kimme-Smith, C., Fajardo, L.L.: Beam optimization for digital mammography. In: Martin, J.Y. (ed.) IWDM 2000: 5th International Workshop on Digital Mammography, pp. 108–119. Medical Physics Publishing, Madison, Wisconsin (2001)

    Google Scholar 

  2. Boone, J., Shaber, G., Tecotzky, M.: Dual energy mammography: A detector analysis. Med. Phys. 17, 665–675 (1990)

    Article  Google Scholar 

  3. Jennings, R.L., Quinn, P.W., Gagne, R.M., Fewell, T.R.: Evaluation of x-ray sources for mammography. In: Proc. SPIE, vol. 1896, pp. 259–268 (1993)

    Google Scholar 

  4. Huda, W., Sajewicz, A.M., Ogden, K.M., Dance, D.R.: Experimental investigation of the dose and image quality characteristics of a digital mammography imaging system. Med. Phys. 30(3), 442–448 (2003)

    Article  Google Scholar 

  5. Boone, J.: Glandular breast dose for monoenergetic and high-energy x-ray beams: Monte Carlo assessment. Radiology 213, 23–37 (1999)

    Google Scholar 

  6. Sobol, W.T., Wu, X.: Parameterization of mammography normalized average glandular dose tables. Medical Physics 24(4), 547–555 (1997)

    Article  Google Scholar 

  7. Stanton, L., Villafana, T., Day, J., Lightfoot, D.: Dosage evaluation in mammography. Radiology 150, 577–584 (1984)

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2006 Springer-Verlag Berlin Heidelberg

About this paper

Cite this paper

Williams, M.B. et al. (2006). Beam Optimization for Digital Mammography – II. In: Astley, S.M., Brady, M., Rose, C., Zwiggelaar, R. (eds) Digital Mammography. IWDM 2006. Lecture Notes in Computer Science, vol 4046. Springer, Berlin, Heidelberg. https://doi.org/10.1007/11783237_38

Download citation

  • DOI: https://doi.org/10.1007/11783237_38

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-540-35625-7

  • Online ISBN: 978-3-540-35627-1

  • eBook Packages: Computer ScienceComputer Science (R0)

Publish with us

Policies and ethics