Dynamic Shunt Testing Applying Short Lasting Pressure Waves — Inertia of Shunt Systems

  • Conference paper
Intracranial Pressure and Brain Biochemical Monitoring

Part of the book series: Acta Neurochirurgica Supplements ((NEUROCHIRURGICA,volume 81))

  • 366 Accesses

Summary

Laboratory shunt testing often comprises only static pressure flow and flow pressure tests. We applied shorter acting pressure waves using a computerised shunt testing rig to investigate shunt behaviour under conditions that might occur in the clinical situation, e.g. during nocturnal vasogenic pressure waves or shorter rises in ICP at movements or exercise. Additionally the influence of a human-like compliance situation compared to a fixed pressure/volume relationship was investigated. Shunts behaved very differently than seen in static tests and demonstrated a marked inertia the shorter pressure waves acted. Although some valves opened at higher pressure, all valves showed marked hysteresis and none did close - if at all - at the specified pressure level. This behaviour might be a cause of shunt overdrainage not related to siphoning. The simulation of a human-like variable pressure-volume relationship with higher compliance at lower pressures had a positive effect on shunt function by decreasing the amount of drained volume. We therefore suggest to include dynamic pressure wave testing if hydrodynamic properties of shunts are to be evaluated.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 84.99
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 109.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Similar content being viewed by others

References

  1. Czosnyka M, Czosnyka Z, Whitehouse H, Pickard JD (1997) Hydrodynamic properties of hydrocephalus shunts: United Kingdom Shunt Evaluation Laboratory. J Neurol Neurosurg Psychiatry 62: 43-50

    Article  PubMed  CAS  Google Scholar 

  2. Kadowaki C, Hara M, Numoto M, Takeuchi K, Saito I (1995) CSF shunt physics: factors influencing inshunt CSF flow. Childs Nerv Syst 11: 203-206

    Article  PubMed  CAS  Google Scholar 

  3. Schuhmann MU, Engel M, Runge L, Samii M, Brinker T Application of clinically recorded ICP patterns - an extension of shunt testing. Accepted for publication at Childs Nerv Syst

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2002 Springer-Verlag

About this paper

Cite this paper

Schuhmann, M.U., Schneekloth, C.G., Klinge, P., Engel, M., Samii, M., Brinker, T. (2002). Dynamic Shunt Testing Applying Short Lasting Pressure Waves — Inertia of Shunt Systems. In: Czosnyka, M., Pickard, J.D., Kirkpatrick, P.J., Smielewski, P., Hutchinson, P. (eds) Intracranial Pressure and Brain Biochemical Monitoring. Acta Neurochirurgica Supplements, vol 81. Springer, Vienna. https://doi.org/10.1007/978-3-7091-6738-0_5

Download citation

  • DOI: https://doi.org/10.1007/978-3-7091-6738-0_5

  • Publisher Name: Springer, Vienna

  • Print ISBN: 978-3-7091-7397-8

  • Online ISBN: 978-3-7091-6738-0

  • eBook Packages: Springer Book Archive

Publish with us

Policies and ethics