Abstract

A broad band and a tuned pulsed X-band ENDOR system with a high field ENDOR probehead are presented. With the tuned version, pulsed radio frequency (rf) fields up to 20 mT (rotating frame) may be generated. The performance of the probehead is demonstrated by pulsed ENDOR experiments on solids and liquids.

References

  1. [1]
    J. Forrer, S. Pfenninger, J. Eisenegger and A. Schweiger, Rev. Sci. Instrum., submitted.Google Scholar
  2. [2]
    S. Pfenninger, J. Forrer, A. Schweiger and Th. Weiland, Rev. Sci. Instrum. 59, 752 (1988).ADSCrossRefGoogle Scholar
  3. [3]
    C. Gemperle, A. Schweiger and R.R. Ernst, Chem. Phys. Lett. 145, 1 (1988).ADSCrossRefGoogle Scholar
  4. [4]
    S. Pfenninger, J. Forrer, A. Schweiger and Th. Weiland, Physica Medica 5, in press.Google Scholar
  5. [5]
    D. Schmalbein, A. Witte, R. Röder and G. Laukien, Rev. Sci. Instrum. 43, 1664 (1972).CrossRefGoogle Scholar
  6. [6]
    J. Forrer, A. Schweiger and Hs. H. Günthard, J. Phys. E 10, 470 (1977).ADSCrossRefGoogle Scholar

Copyright information

© Springer-Verlag Berlin Heidelberg 1990

Authors and Affiliations

  • Jörg Forrer
    • 1
  • Susanne Pfenninger
    • 1
  • Josef Eisenegger
    • 1
  • Arthur Schweiger
    • 1
  1. 1.Laboratorium für Physikalische ChemieEidgenössische Technische HochschuleZürichSwitzerland