Advertisement

DTA/EGA Using a Specific Detector

  • S. B. Warrington
  • P. A. Barnes

Abstract

A highly sensitive, continuous-reading electrolytic hygrometer capable of measuring water concentrations in a gas stream down to ppm levels has been coupled to a Stanton-Redcroft DTA 671B Differential Thermal Analyser. The method of coupling and operating precautions are described. Integration of the EGA peaks obtained shows the combination to be capable of measuring sub-milligram quantities of evolved water to an accuracy of better than 4% Applications discussed include hydrate decomposition and degradation of plant acids used in the food industry.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    H. A. Keidel Anal. Chem. 31, 2043 (1959)CrossRefGoogle Scholar
  2. 2.
    J. E. Still and H. J. Cluley SAC Conf. (1965) P.405Google Scholar
  3. 3.
    M. I. Pope and D. I. Sutton Thermochim. Acta 23, 188 (1978)CrossRefGoogle Scholar
  4. 4.
    B. Lorant Mitt.Geb. Lebensmittelunters u. Hyg. 57, 231 (1966)Google Scholar

Copyright information

© Springer Basel AG 1980

Authors and Affiliations

  • S. B. Warrington
    • 1
  • P. A. Barnes
    • 1
  1. 1.School of Health and Applied SciencesThe PolytechnicLeeds 1England