The Effect of the Position of Compression Screw in Plate-Screws Method Used in Fracture Treatment on Fatigue Strength of Joint

  • R. Cagri Kutlubay
  • Tezcan SekerciogluEmail author
  • A. Cagdas Yorukoglu
Chapter
Part of the Advanced Structured Materials book series (STRUCTMAT, volume 98)

Abstract

A number of combination studies on the use of plate-screws, which has become a traditional method in fracture treatment, have been conducted to obtain a more rigid joint. These studies in the literature were reviewed and it was seen that the effect of the position of the compression screw, used with the aim of making the fracture line come closer to each other during fracture fixation, on the fatigue strength of the joint hasn’t been investigated. Within the scope of this study, a screw combination study by including Limited Contact Dynamic Compression Plate (LC-DCP) plates that possess combined hole properties was carried out. Three different combinations were created during the study. In these combinations, the unlocked screw that was located on the part where the force was applied and that was used for compression was placed in different holes and fatigue strengths of the joints were investigated under dynamic bending force. As a result of the study, it was observed that as the compression screw moved away from the fracture line, fatigue strength of the joint consisting of plate, screw and bone decreased. In addition, it was found that all the joints in the 3rd combination got damaged as a result of the fracture of the bone.

Keywords

Anatomic plate Plate screw Limited contact dynamic compression plate (LCP) Fatigue Biomechanics 

Notes

Acknowledgements

This study was performed in the framework of The Scientific Projects Unit of Pamukkale University, Project No: 2016-FEBE-038

References

  1. 1.
    Törnkvist, H., Hearn, T.C., Schatzker, J.: The strength of plate fixation in relation to the number and spacing of bone screw. J. Orthop. Trauma 10, 204–208 (1996)CrossRefGoogle Scholar
  2. 2.
    Field, J.R., Törnkvist, H., Hearn, T.C., Summer-Smith, G., Woodside, T.D.: The influence of screw omission on construction stiffness and bone surface strain in the application of bone plates to cadaveric bone. Int. J. Care Injured 30, 591–598 (1999)CrossRefGoogle Scholar
  3. 3.
    ElMaraghy, A.W., ElMaraghy, M.W., Nousiainen, M., Richards, R.R., Schemitsch, E.H.: Influence of the number of cortices on the stiffness of plate fixation of diaphyseal fractures. J. Orthop. Trauma 15, 186–191 (2001)CrossRefGoogle Scholar
  4. 4.
    Sanders, R., Haidukewych, G.J., Milne, T., Dennis, J., Latta, L.L.: Minimal versus maximal plate fixation techniques of the ulna: the biomechanical effect of number of screws and plate length. J. Orthop. Trauma 16, 166–171 (2002)CrossRefGoogle Scholar
  5. 5.
    Stoffel, K., Dieter, U., Stachowiak, G., Gächter, A., Kuster, M.S.: Biomechanical testing of the LCP-how can stability in locked internal fixators be controlled. Int. J. Care Injured 34, B11–B19 (2003)CrossRefGoogle Scholar
  6. 6.
    Stoffel, K., Stachowiak, G., Forster, T., Gächter, A., Kuster, M.: Oblique screws at the plate ends increase the fixation strength in synthetic bone test medium. J. Orthop. Trauma 18, 611–616 (2004)CrossRefGoogle Scholar
  7. 7.
    Bekler, H., Bulut, G., Usta, M., Gökçe, A., Okyar, F., Beyzadeoğlu, T.: Osteoporotik kemikte kilitli plak ve açılı vida kullanımının stabilizasyonun dayanıklılığına katkısı: Deneysel çalışma. Acta Orthop. Traumato. 42, 125–129 (2008)Google Scholar
  8. 8.
    Freeman, A.L., Tornetta, P., Schmidt, A., Bechtold, J., Ricci, W., Fleming, M.: How much do locked screw add to the fixation of “hybrid” plate construct in osteoporotic bone? J. Orthop. Trauma 24, 163–169 (2010)CrossRefGoogle Scholar
  9. 9.
    Hak, D.J., Althausen, P., Hazelwood, S.J.: Locked plate fixation of osteoporotic humeral shaft fractures: are two locking screw per segment enough? J. Orthop. Trauma 24, 207–211 (2010)CrossRefGoogle Scholar
  10. 10.
    Grawe, B., Le, T., Williamson, S., Archdeacon, A., Zardiackas, L.: Fracture fixation with two locking screws versus three non-locking screws, a biomechanical comparison in a normal and an osteoporotic bone model. Bone Joint Res. 1, 118–124 (2012)CrossRefGoogle Scholar
  11. 11.
    Lee, C.H., Shih, K.S., Hsu, C.C., Cho, T.: Simulation-based particle swarm optimization and mechanical validation of screw position and number for the fixation stability of a femoral locking compression plate. Med. Eng. Phys. 36, 57–64 (2014)CrossRefGoogle Scholar
  12. 12.
    Overturf, S.J., Morris, R.P., Gugala, Z., Lindsey, R.W.: Biomechanical comparison unicortal far-cortex-abutting locking screw-plate fixation for comminuted radial shaft fractures. J. Hand Surg. Am. 39, 1907–1913 (2014)Google Scholar

Copyright information

© Springer Nature Switzerland AG 2019

Authors and Affiliations

  • R. Cagri Kutlubay
    • 1
  • Tezcan Sekercioglu
    • 2
  • A. Cagdas Yorukoglu
    • 3
  1. 1.Technology Faculty, Department of Biomedical EngineeringPamukkale UniversityDenizliTurkey
  2. 2.Engineering Faculty, Department of Mechanical EngineeringPamukkale UniversityDenizliTurkey
  3. 3.Faculty of Medicine, Department of Orthopaedics and TraumatologyPamukkale UniversityDenizliTurkey

Personalised recommendations