Electron spin resonance (ESR) tooth dosimetry is an established method of evaluating radiation dose by measuring stable unpaired electrons produced by radiation exposure in a tooth [1, 2]. Ionizing radiation generates radicals proportionate to absorbed dose depending on radiation quality and energy. These radicals react rapidly and disappear in the most part of our body due to water, but in organized matrices without water, they can persist indefinitely. This phenomenon has been recognized in bones and teeth and has been shown to be a promising method for retrospective dosimetry, based on free radicals being stabilized in the hydroxyl apatite matrix in tooth enamel (Fig. 13.1). It has been confirmed that mean detection limit is 205 mGy, ranging from 56 to 649 mGy at the fourth International Comparison [3]. Using extracted teeth, X-band ESR tooth dosimetry has been applied to retrospective dose assessment of atomic bomb survivors of Hiroshima and Nagasaki (Hibakusha) [4, 5], the Chernobyl Nuclear Power Plant (CNPP) accident [6], nuclear bomb tests [7], and radiological accidents [8,9,10]. An “ideal” ESR spectrum with some parameters is shown in Fig. 13.2. The peak-to-peak amplitude “A” is the first derivative of the resonance microwave absorption. The peak-to-peak linewidth, “∆B,” is related to the modulation frequency that is important for controlling the signal-to-noise ratio. The resonance field “Br” indicates the natural resonance frequency of the radical species due to the Zeeman effect. The dose obtained with ESR dosimetry is well correlated with the standard method such as hematology-based assays [11]. Those dosimetry studies are carried out at conventional ESR frequencies called X-band of the electromagnetic spectrum (e.g., 9 GHz) so that the method is called X-band ESR tooth dosimetry. This X-band ESR has high sensitivity for dose estimation using the extracted or deciduous teeth. ESR spectra of tooth enamel after exposure to different doses are shown in Fig. 13.3.

Fig. 13.1
figure 1

Anatomy of a tooth and locations of tooth tissues

Fig. 13.2
figure 2

“Ideal” ESR spectrum with illustrations of some its parameters. (Original: International Atomic Energy Agency. Use of electron paramagnetic resonance dosimetry with tooth enamel for retrospective dose assessment. IAEA Vienna IAEA-TECDOC-1331 (2002). https://www-pub.iaea.org/MTCD/Publications/PDF/te_1331_web.pdf)

Fig. 13.3
figure 3

ESR spectra of tooth enamel after 5Gy irradiation. The dosimetric component of the spectrum after irradiation with 5 Gy is shown in this chart. R is the peak-to-peak amplitude used for ESR dose reconstruction. (Original: International Atomic Energy Agency. Use of electron paramagnetic resonance dosimetry with tooth enamel for retrospective dose assessment. IAEA Vienna IAEA-TECDOC-1331 (2002). https://www-pub.iaea.org/MTCD/Publications/PDF/te_1331_web.pdf)

With a radiation emergency in mind, an L-band ESR apparatus has been developed for in situ measurement as triage using teeth in the mouth as indices [12]. This method is available for retrospective dose assessment and for triage when workers are involved in an accident that caused radiation exposure such as a nuclear disaster or a large-scale radiation exposure accident [13].

ESR tooth dosimetry is a method of evaluating the accumulated radiation dose and is thought to be applicable for animals affected by the FNPP accident, which is crucial for studying the biological effect of persistent low-dose-rate (LDR) radiation [14,15,16]. The FNPP accident affected animals and several environmental study groups started research activities on livestock. We have been analyzing the influence of radiation exposure on affected cattle since September 2012. We have carried out dose estimation using information of environmental monitoring and radioactivity deposition density on the ground. We have collected 50 cattle at Okuma Farm in Okuma Town, Fukushima Prefecture (Fig. 13.4). Research of the radiation effect on large animals is still a challenging topic. To obtain scientific results, it is essential to estimate radiation dose accurately for discussing the dose response. In order to validate ESR tooth dosimetry, we estimated dose of cattle affected by the FNPP accident using ESR and compared the dose with that estimated using monitoring tendency data.

Fig. 13.4
figure 4

Radiation dose-rate map obtained by airborne monitoring results and the location of Okuma Farm as of May 31, 2012. This map was quoted from the website provided by Nuclear Regulation Authority, Japan. Dose rate shows μSv/h as a dose equivalent rate including background radiation. (http://ramap.jaea.go.jp/map/)

2 Material and Methods

2.1 Sample Collection

Teeth were collected from cattle that had continued to stay in Okuma Farm, that is located in the ex-evacuation zone of the FNPP accident (Fig. 13.4). Dissection of cattle was performed on May 14, 2017. Table 13.1 shows information on cattle whose radiation dose was measured in this study.

Table 13.1 Basic characteristics of each cattle and estimated radiation dose

2.2 Nondestructive Measurement Using L-Band ESR Spectroscopy

For the comparison of the L-band ESR response among different energy photons, we used two standard materials. One was a standard tooth irradiated by the cesium-137 (137Cs) source provided from the ESR Center for the Study of Viable Systems at Dartmouth, NH, USA. The other was a tooth sample exposed to 20 Gy of X-rays in an X-ray irradiator (MBR-1505R2, Hitachi Medical, Tokyo, Japan) at 150 kV and 1 mA, with 0.1-mm copper plus 0.3-mm aluminum filtering. This X-ray irradiation gives a response about 4 times as large as γ-rays from 137Cs.

In order to adjust for variations in the measured radiation-induced signal (RIS) amplitude that result from instrumental instability and/or external environmental factors, the ratio of differentiated voltage as a radiation-induced signal to the differentiated voltage of 4-oxo-2,2,6,6,-tetramethylpiperidine-d16–1-15 N-1-oxyl (15N-PDT) for each measurement was calculated and normalized to the same ratio in the standard tooth exposed to 20 Gy of 150 kV X-ray as described above.

After extracting from the jawbone portion of each cattle, a maxillary incisor tooth was washed with tap water. Then with a design knife, as much soft tissue as possible was removed and the labial side was used for the measurement surface by L-band ESR tooth dosimetry. The ESR spectrum was acquired using standard measurement parameters such as 20 scans, including a scan range of 2.5 mT, a scan time of 3 sec, and a modulation amplitude of 0.4 mT [17]. This process was repeated 3 times. A plastic tube containing a solution of 15N-PDT was fixed in a position in close proximity to the surface loop and used as a reference of positive standard. Spectra from each of the collected datasets were analyzed using non-linear least-squares fitting to estimate the peak-to-peak signal amplitudes of the radiation-induced signals and of 15N-PDT. RIS and signals of 15N-PDT were then averaged to provide the mean amplitude for each tooth and at each dose.

2.3 Destructive Measurement by X-Band ESR Spectroscopy

The enamel portion on the labial side of the cattle teeth was taken out by a nipper about 3 mm in size, taking care not to contain dentin, and was used as a sample for X-band ESR spectroscopy after grinding in an agate mortar to a size of 1 mm in diameter so as to enter a sample tube. ESR spectra were acquired using parameters of 40 scans, including a scan range of 5 mT, a scan time of 30 sec and a modulation amplitude of 0.2 mT power of 2 mW, and a time constant of 0.03 s. This process was repeated for a total of five datasets at each dose.

2.4 Measurement of Radioactivity Concentration

Radioactive concentrations of Cs and strontium-90 (90Sr) in the molar of Cattle A-3 were measured. For radioactive Cs, the ashed sample was analyzed by a germanium semiconductor detector (CFG-SV-76, ORTEC, TN, USA). Sr-90 was extracted by the ion exchange method and analyzed by a low-background beta-ray measuring device after achieving radioactive equilibrium with yttrium-90 (Y-90). In principle, measurement was performed for 3,600 sec, and the radioactivity concentration was calculated.

2.5 Data and Statistical Analysis

Data analysis was carried out using code developed by Ivannikov [18]. Statistical analysis was performed using R3.3.1 [19]. Radiation dose of each sample was determined by the additive dose-response method assuming a linear dose-response of the ESR signals to additional irradiated doses of each sample. This method employs step-by-step irradiation to each sample additionally. Absorbed dose is estimated by a linear regression analysis [2].

3 Results

3.1 Dose Estimation by Dose-Rate Monitoring Using Survey Meters

Cumulative radiation dose based on ambient monitoring using an air chamber survey meter during 6 years since March 2011 was estimated to be 270 mGy [20].

3.2 Dose Estimation by Radiation Dose Monitoring Using Individual Dosimeters

Radiation dose measured by an individual dosimeter attached to representative cattle from December 2013 to December 1, 2017 was 57 mGy (Fig. 13.5). Considering weathering, total dose from the day the FNPP accident happened to the start of monitoring was calculated from deposition density of iodine-131 (131I), tellurium-132 (132Te), 132I, 134Cs and 137Cs on the ground immediately after the accident [21]. Estimated total dose of Cattle A-1 and A-2 from March 2011 was 160 mGy. Similarly total dose of Cattle A-2 calculated was 67 mGy after adjusting deposition density of radionuclides to its birth month.

Fig. 13.5
figure 5

Cumulative radiation doses to cattle monitored by using glass dosimeters since December 2013

3.3 Dose Estimation by ESR Tooth Dosimetry

Positive RIS were not detectable in any cattle sample examined by nondestructive measurement using L-band ESR dosimetry (Fig. 13.6). For comparison, an exposed human tooth is shown in Fig. 13.7. In this spectrum, RIS were detected between 0.7 and 1.1 mT.

Fig. 13.6
figure 6

Representative ESR spectrum of a sample of incisors from Cattle A-2 using L-band spectrometry. Measurement condition: 20 scans, scan range of 2.5 mT, a scan time of 30 sec, a modulation amplitude of 0.4 mT. The signal around 0.8–1.2 mT shows a positive reference control of 15N-PDT

Fig. 13.7
figure 7

Example of ex vivo ESR spectrum of human tooth using the L-band. This tooth is exposed to 1Gy by X-ray. Measurement condition: 20 scans, scan range of 2.5 mT, a scan time of 3 sec, a modulation amplitude of 0.4 mT. The signal from a positive control is observed around 0.6–0.9 mT. Radiation induced signal is observable at 0.5–1 mT

A representative X-band ESR tooth spectroscopy of Cattle A-1 is shown in Fig. 13.8. The additive-dose-response for radiation dose estimation is shown in Fig. 13.9. The dose of this sample is the value of negative x-intercept value, 67 mGy. The uncertainty of dose estimation was calculated using the Monte Carlo method by picking up randomly generated parameters 100 times, adapting the model obtained by the additive-dose method. Estimated 95% confidence interval was between 1.3 and 119 mGy. Total dose estimated by X-band ESR tooth dosimetry was compared with other methods, that is, calculated from ambient dose-rate and from individual monitors after adding the initial estimated dose before December 2013 (Table 13.2). Both estimations using ground-deposited radioactive Cs and individual monitor were in good agreement. Although the dose range was large, doses estimated by X-band ESR tooth dosimetry of Cattle A-1 and A-3 were not much different from those calculated by other two methods. In Cattle A-2, the dose assessed by X-band ESR tooth dosimetry was eight times the dose assessed by other two methods.

Fig. 13.8
figure 8

Example of spectrum obtained by X-band ESR tooth dosimetry. In the middle, tooth signals are shown including radiation-induced signal. At both sides, there are standard signals from a Mn marker. The right side is third signal of Mn and the left side is fourth signal of Mn

Fig. 13.9
figure 9

Estimated radiation dose of a sample from Cattle A-1 by the additive-dose method. (a) Irradiated dose was converted from X-ray to 137Cs equivalent considering the energy dependency of this ESR tooth dosimetry. Dose is the value of negative x-intercept value, 67 mGy. (b) The uncertainty of this dose estimation. The top of the box represents the third quartile and the bottom of the box represents the first quartile. The upper whisker and the lower beard show 1.5 interquartile range. Outliers are plotted as individual points. In this measurement, estimated dose was 67 ± 30 mGy showing lower value for this cattle

Table 13.2 Estimated radiation dose [mGy]

3.4 Radioactivity Concentration in the Molar Tooth of Cattle A-3

Radionuclides and their concentrations detected in the ashed molar tooth of Cattle A-3 were 90Sr (46 ± 2.3 Bq kg), 134Cs (60.7 ± 9.9 Bq/kg), and 137Cs (60.6 ± 1.7 Bq/kg).

4 Discussion

4.1 Comparisons of Estimated Radiation Dose

Total dose using X-band ESR tooth dosimetry was 37–1230 mGy. These values are consistent with the estimated radiation dose using monitoring and individual dosimeters, considering that Cattle A-2, who showed had the highest exposure dose, was moved from the higher dose area to Okuma Farm during the 6 years.

4.2 Limitations of ESR Tooth Dosimetry in Regard to Variance and Energy Dependency

This study tried to measure the actual dose of cattle affected by the FNPP accident using both X-band and L-band ESR tooth dosimetries. Due to its low sensitivity, L-band ESR tooth dosimetry revealed to be not practical for dose assessments associated with the FNPP accident and that the X-band dosimetry can be available with some improvements. Dose of each sample was measured by the additive-dose method. Measurements were performed on each enamel sample individually, taking into account the sample properties and mass. This is because, even with the same tooth, the degree of stable radical generation is affected by the chemical structure of the enamel sample. Response characteristics of thermoluminescence dosimeter (TLD) depend on the mass energy-absorption coefficient of photons and the mass stopping power of charged particles. In dose measurements, energy dependence of ESR dosimetry is relatively similar to TLD. When measuring with animal samples by the additive dose method, the dose response may vary from sample to sample. Therefore, it is more desirable to measure separately for each sample. In fact, there is significant variation in dose response of ESR dosimetry among samples of the same boar, especially in older boars [22].

4.3 Effect of External Exposure Due to ß-Particles from Deposited Radionuclides

Assuming that the deposited density of 137Cs was 1 MBq/m2, external dose-rate of the skin in contact with the ground was estimated to be around 0.14 mGy/h using coefficient of ICRU report 56 [23]. Assuming 8 h a day of contact with the ground, the total skin dose from β-particles for 6 years after the FNPP accident was 1.9 Gy with consideration of weathering. It is a large value in the dose calculation and cannot be ignored. However, this estimated dose was not detectable by ESR tooth dosimetry, indicating that β-particle exposure to the teeth is negligible unless getting a tooth close to the land.

4.4 Effect of Internal Exposure

Other concern was a potential impact of internal exposure by radioactive substances inside a tooth and other organs. However, self-absorption dose, that is, radiation dose to a tooth from radionuclides in the same tooth, is limited. It was because of the low concentration of the total radionuclides in tooth, and taking the radioactive strontium as an example, the tooth dose was as low as 2 mGy.

4.5 Limitations of ESR Tooth Dosimetry in Regard to the Variance of Background Signals

The background signal in enamel could be induced by ultraviolet from sunlight [24, 25] or mechanical stimulation during sample preparation [26]. Therefore, further confirmation on the added radiation dose due to the nuclear accident is needed to clarify the attribution of possible causes.

5 Conclusion

Radiation dose measured by ESR tooth dosimetry reflects the radiation history of each subject. In this study, the dose measured by ESR tooth dosimetry was consistent with the monitored radiation dose. These indicate that teeth can be used to assess radiation dose from a radiation accident. However, L-band ESR dosimetry was not sensitive enough to detect the radiation exposure to cattle in Okuma Town.