Introduction

As the interest in the esthetic aspect of implant prosthesis increases, the importance of the esthetic element has been emphasized. It has been reported that ready-made titanium abutments leave a gray shade if the soft tissue is thin or if the gingival sulcus is shallow, and a non-esthetic outcome may be caused by the exposure of titanium in the gingival margin when a gingival recession occurs [1]. Alumina ceramic abutments have been developed to overcome the non-esthetic problem, but failed due to fracture caused by the brittleness of the alumina ceramic during functioning was reported. Since then, zirconium with high biological stability, an esthetic feature, and high mechanical stability has been used as an abutment [1]. According to the 2011 MFDS (Ministry of Food and Drug Safety, Republic of Korea) report, esthetic prosthetic treatment using zirconium prosthesis and implant abutment is on the spotlight, and over 12% of the high annual average growth rate has been shown in South Korea.

Zirconium abutments are widely used in South Korea, but it is not easy to find evidence of their clinical stability for long-term use by the Koreans. Glauser et al. reported that there was no fracture in the zirconium abutments in their study for an average of 49.2 months in single-tooth restoration [2]. Zembic et al. reported the long-term stability of zirconium abutments, showing a 96.3% success rate [3]. Additionally, the studies conducted by Canullo et al., Sailer et al., Zembic et al., and Nothdurft et al. reported a 100% success rate in zirconium abutment clinical practices [4,5,6,7]. All these studies, however, were conducted in Caucasian countries. Michael et al. reported that the maximum bite force was 725 N, and the average masticatory force was 262 N in natural teeth [8]. Ferrario et al. reported that the maximum bite force was around 700 N in healthy adult males and females [9]. In Korean research, however, Cho reported that the maximum bite force was 923.8 N on average [10], and Yoon reported that the average bite force was 744.5 N [11]. Koreans who favor hard foods such as kimchi, gakdugi, squid, and ribs deliver a strong masticatory force to the molar area [12]. Kim et al. examined the subjective food intake ability of the maximum occlusal force among Korean adults [13]. Considering the difference in diet and bite force between the Koreans and Caucasians, the stability of zirconium needs to be evaluated in the Koreans and the evidence level must be established.

According to the Straumann report 2013, Korea is the country where dental implant treatment is most performed [14]. The Korean National Health Insurance Corporation announced that it would expand the coverage of dental implants in patients aged 70 and older beginning in 2015 and those aged 65 and older in 2016. Only patients older than 65 are eligible, but Korea is the second country to add implant treatment to the national health insurance after Sweden. However, the zirconium abutment was not included in the health insurance. We have reviewed the validity of zirconium abutment through Korean and English literature and wanted to provide a basis for its use. This study intended to help increase the treatment stability by offering evidence to the dentist’s choice through the stability assessment of zirconium abutments, with a focus on evidence-based dentistry. The goal of this study is to review the status of clinical studies on the stability of zirconium abutments in papers in Korea and abroad and to refer to them for application in clinical medicine.

Methods

A key question was set. The significant element in determining the key question included the patient/problem (P), intervention (I), comparison (C), and outcome (O) (PICO). The combination of these four major elements is “PICO asking questions.” According to PICO asking questions, “Does the zirconium abutment (P and I) have mechanical, biological, and esthetic stability (O) compared to the titanium abutment (C)?” was set as a key question.

Search strategy

Literature was searched, using the keywords “zirconia abutment,” or “zirconium abutment,” in KoreaMed, KMbase, and National Assembly Electronic Library, which were recommended as domestic literature search sites by 「Evidence Literature Utilization Guidelines」 published by the Health Insurance Review and Assessment Service of Republic of Korea (2013a). Literature was searched using keywords including “zirconia abutment” and “zirconium abutment” in PubMed to compare the evidence levels of the stability of zirconium abutments in the domestic and international literature. The searched literature was selected by determining the inclusion and exclusion criteria based on the concept of evidence-based dentistry. The domestic literature included studies with Korean subjects and literature written in Korean. The scope of this study was clinical literature. Studies with international subjects, animal experiments, laboratory studies, expert commentaries, and scholarly literature were excluded.

Stability assessment criteria and classification of the evidence level

Stability assessment criteria

In this literature, mechanical stability means that there are no problems of zirconium abutment fracture, abutment screw fracture, etc. To evaluate the biological stability, the conditions of the tissues around the zirconium abutments including the height of the bones, the plaque accumulation level in the soft tissue, the bleeding on probing, and the depth of the periodontal pocket were compared with the conditions of the tissues around the titanium abutments or around the healthy natural teeth. The esthetic stability was evaluated to be stable when the color and shape of the soft tissue surrounding the zirconium abutments were similar to those of the natural teeth.

Classification of the study type according to the evidence level

In this study, the grade criteria for clinical research literature suggested by the Health Insurance Review and Assessment Service were used. After the confirmation of the study type and contents that were selected based on Table 1, the evidence level was classified. According to the grade criteria for clinical research literature, the evidence level of systematic literature review targeting randomized controlled trials (RCTs) is the highest, and the evidence level of cross-sectional study, case series, before/after study, case report, and non-analytic study is the lowest. As such, the lower the grade is, the higher the level of evidence.

Table 1 Study type according to evidence level

Main text

As a result of the domestic literature search, a total of 102 kinds of literature were searched, including 28 kinds of literature at KoreaMed, 35 kinds of literature at KMbase, and 39 kinds of literature at the National Assembly Library. Of these 102 kinds of literature, 48 duplicate kinds of literature, 35 kinds of literature that fell under the exclusion criteria (including animal experiments and laboratory studies) based on a review of titles and abstracts, and three kinds of literature whose scope was not related to implant zirconium abutment were excluded. After a review of the full texts of the literature, six expert commentaries and one Master’s thesis that was published in an international journal with a different title were excluded. Finally, nine literature were selected (Fig. 1).

Fig. 1
figure 1

The flow chart of the search strategy of the Korean literature

International literature was searched at the Refomax Electronic Library of Hallym University Medical Center by linking with PubMed using keywords including “zirconium abutment” and “zirconia abutment.” The literature type was set to review, clinical trial, and case report, and a total of 92 kinds of literature were selected: 13 reviews, 37 clinical trials, and 42 case reports. Based on the titles and abstracts, animal experiments, laboratory studies, and studies not related to implant zirconium abutment were excluded, and only the literature whose full texts could be obtained from the electronic library of the Hallym University Medical Center were selected. Finally, five reviews, eight clinical trials, and six case reports were selected (Fig. 2).

Fig. 2
figure 2

The flow chart of the search strategy of the international literature

After the confirmation of the study types of the domestic and international literature, the evidence levels were classified. The classification results are presented in Table 2. For the domestic literature, there was no systematic literature review targeting RCTs, RCTs, and systematic literature reviews, which fall under evidence levels 1 and 2, respectively. There were three cohort studies with evidence level 3, and 6 case reports with evidence level 4.

Table 2 Status of the evidence levels of domestic and international studies

In the international literature search process, there was a function for setting a literature type; thus, it was easy to classify the evidence level compared to the domestic literature search. About the evidence levels of international literature, there were five systematic literature reviews with evidence level 2, four cohort studies, and three case-control studies with evidence level 3, and 1 case series and 6 case reports with evidence level 4.

Domestic literature

Evidence level 3

A summary of three domestic kinds of literature with evidence level 3 is presented in Table 3. In the studies conducted by Kim et al., regarding the stability of zirconium abutments, the survival rate was high regardless of the implant location and prosthesis type, and the number of prosthesis units and prosthesis types appeared to be significantly associated with the complication rate of zirconium abutments [15]. Also, the zirconium abutments were mechanically stable within the 5-year follow-up period. Bae et al. reported that the zirconium abutments in their study were biologically stable [16].

Table 3 Summary of domestic studies with evidence level 3

Evidence level 4

A summary of six domestic kinds of literature with evidence level 4 is presented in Table 4. Evidence for the esthetic stability of zirconium abutments was shown in five kinds of literature out of six case reports with evidence level 4.

Table 4 Summary of domestic studies with evidence level 4

International Literature

Evidence level 2

A summary of the outcomes of five international literature with evidence level 2 is presented in Table 5. The stability evidence of zirconium abutments derived from the international literature with evidence level 2 can be roughly classified into two kinds. First, zirconium abutments are mechanically, biologically, and esthetically stable in general. In 16 literature evaluated by Medeiros et al. using the systematic literature review method [17], four literature evaluated by Guess et al. [18, 19], five literature evaluated by Gomes et al. [19], eight literature evaluated by Nakamura et al. [20], and one literature evaluated by Linkevicius et al. [21], the mechanical, biological, and esthetic stability of zirconium abutments was shown. The stability of zirconium abutments for long-term use, however, cannot be ensured. Medeiros et al. summarized 16 literature and revealed that the outcomes about the biological stability of zirconium abutments were varied [17]. Second, the stability of zirconium appeared to be positive within the short-term (4-year) clinical observation period. Four review literature, all reported that zirconium abutments are stable in the short term (within 5 years), but it is impossible to conclude the long-term stability of zirconium abutments due to the lack of long-term clinical studies.

Table 5 Summary of international studies with evidence level 2

Evidence level 3

In Table 6, the outcomes of seven international literature with evidence level 3 are summarized. The representative literature falling under evidence level 3 include cohort studies conducted by Zembic and Sailer [3, 6]. The studies conducted by Zembic and Sailer reported the survival rate of the single crown using 20 titanium abutments and 20 zirconium abutments in 40 patients [3, 6]. Additionally, the probing pocket depth, plaque control record, bleeding on probing, and bone level on radiography were compared with those of the opposite tooth. The stability evidence of zirconium abutments derived from international literature with evidence level 3 is roughly classified into three kinds.

Table 6 Summary of international studies with evidence level 3

First, zirconium abutments are mechanically stable within the 3- and 5-year follow-up periods. This study emphasized that the average 5-year follow-up period is longer than the follow-up periods of other studies. Second, zirconium abutments are stable for the restoration of a single tooth in the anterior and premolar regions. In the study of Glauser et al., single crowns using 54 zirconium abutments were observed for 4 years, and mechanical problems such as abutment fracture and abutment screw fracture, and marginal bone loss, were examined. During the average 49.2-month follow-up period, abutment fracture did not occur, and the mean marginal bone loss was reported to be 1.2 mm. Third, the esthetic stability of zirconium abutments cannot be determined [2]. Sailer et al. reported that there was no significant difference between zirconium and titanium abutments regarding their esthetic features. Fourth, zirconium abutments are biologically stable, but there is no significant difference between zirconium and titanium abutments regarding biological stability [6]. In the study of Van Brakel et al., in both zirconium and titanium, similar amounts of bacteria were detected, and the gingival health conditions were reported to be similar [22, 23]. After designing an experiment as described above, Van Brakel performed the histological examination. In the 3-month histological examination, no statistically significant differences in the vascular density and the inflammation level were found [23].

Evidence level 4

In Table 7, the outcomes of seven international literature with evidence level 4 are summarized. The study object of all the literature with evidence level 4 is the maxillary anterior region.

Table 7 Summary of international studies with evidence level 4

The following three results were derived from international literature with evidence level 4. First, zirconium abutments are mechanically stable. In the study conducted by Lee et al., when nine patients were observed for 52 weeks after early loading, no abutment fracture or abutment screw loosening occurred [24]. Second, zirconium abutments are biologically stable. Aydin et al. reported that there was no difference in the probing pocket depth, plaque control record, bleeding on probing, and bone level on radiography at the soft tissue around the zirconium implant 6 months after the prosthesis installation [25]. Third, zirconium abutments are generally esthetically stable. Kalman et al., Wadhwani et al., Mahn et al., and Schneider et al. reported the esthetic features in the maxillary anterior region [26,27,28,29]. However, Tan et al. reported that cast metal abutments have an excellent gingival contour and that zirconium abutment has an excellent gingival color when cast metal abutments are compared with ready-made zirconium abutments, indicating that zirconium abutments are not always superior to titanium abutments in terms of both the gingival contour and the color [30].

In this study, the domestic and international literature on the stability of zirconium abutments were analyzed, with a focus on evidence-based dentistry. As a result, the following shortcomings were shown.

First, the studies on the stability of zirconium abutments conducted for longer than 5 years are not sufficient. Considering that complications of zirconium prosthesis occur more than 5 years after, long-term studies lasting for more than 5 years are required. The analysis results of the domestic literature showed that zirconium abutments are mechanically, biologically, and esthetically stable, but most of the said literature have evidence level 4, meaning that the level of scientific evidence presented by them is low. Additionally, the stability of zirconium abutments is limited to the 5-year follow-up period. The evidence levels of international literature range from level 2 to level 4, and the scientific-evidence levels are higher and more varied compared to those of domestic literature. Like domestic literature, however, international literature reports the mechanical stability of zirconium abutments to last for 4 years; thus, the scientific evidence for the long-term stability of zirconium abutments is not sufficient. Further studies on the long-term survival rate of zirconium abutments are also required in other countries, and biological measurement values and esthetic outcomes need to be suggested using objective figures in long-term clinical practice.

Second, the domestic and international literature supporting the esthetic stability of zirconium abutments are case reports with a low evidence level, and most of them report the analysis results in a subjective language and have a short follow-up period or do not report the follow-up period. Among the domestic literature, three literature (written by Byeon et al. [31], Lee et al. [32], and Kim et al. [33], respectively) reported the follow-up period, and among the international literature, only two literature (written by Lee et al. [24] and Aydin et al. [25], respectively) reported the follow-up period. The rest of the literature did not mention the follow-up period. For the literature that reported the follow-up period, the maximum follow-up period was 52 weeks, and the average follow-up period was 6 months; thus, it is difficult to ensure esthetic stability. The outcomes were expressed in subjective language, only as “It is esthetic” or “It is stable,” and no statistically significant objective data were provided. Therefore, the conclusion that zirconium abutments are preferred to titanium abutments for an esthetically superior outcome does not have sufficient scientific evidence. To provide a scientific basis for the esthetic stability of zirconium abutments, the gingival color change, etc. need to be expressed objectively through the standardization of figures, using a tool such as a spectrophotometer. The esthetic stability must be evaluated after a more than 5-year-long observation after prosthesis installation.

Conclusions

Zirconium abutments have a high survival rate regardless of implant location and prosthesis type within the 5-year follow-up period. Zirconium abutments are biologically stable. The evidence regarding the esthetic stability of zirconium abutments is not sufficient. Zirconium abutments are mechanically stable within the 5-year follow-up period. The long-term stability of zirconium abutments after 5 years cannot be determined. Therefore, clinical studies that investigate the long-term stability of zirconium abutments are required.