Small retained foreign bodies: what is the limit of detection using current emergency ultrasound equipment?

  • Daniel Jafari
  • KJ Cody
  • NL Panebianco
  • FS Shofer
  • BS Ku
  • A Au
  • AJ Dean
Open Access
Meeting abstract

Keywords

Visual Analog Scale Detection Rate Foreign Body Interventional Radiology Analog Scale 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Background

Previous studies of small foreign bodies (FB) have shown a wide range of accuracies of FB detection using animal models, with high accuracy rates for FB > 10 mm and variable accuracy rates for 4 to 5 mm FB.

Objective

To determine the lower limit of sonographic detection of FB using current emergency ultrasound equipment in a soft tissue model.

Methods

FBs made of metal, glass, wood, and plastic (3 of each) 1 x 1 x 3 mm in size were placed at a depth of 0.5-2.0 cm in 12 pork feet. 8 feet were punctured without FB placement. Pork feet were submerged during this process to minimize air in tissue. 7 ED sonologists with > 2 years experience were blinded to overall number, type and depth of FB, but not to size. FB sites were scanned by each sonologist using either a hockey stick or traditional linear array transducer in a randomized pre-assigned order. Sonologist confidence in the diagnosis was reported using a visual analog scale for each site. Sensitivity, specificity, positive and negative predictive values (PPV and NPV) with 95% confidence intervals were calculated. To determine if sonologist confidence differed by perceived presence or absence of a foreign body, paired t-test was used.

Results

140 ultrasound scans were performed which reported sensitivity, specificity, PPV and NPV as 50% (95% CI: 39%-61%), 50% (37%-61%), 60% (48%-72%), and 40% (28%-52%) respectively. There was little agreement among the sonologists (only 2 sites with 100% agreement). Sensitivity ranged from 25% to 75%, specificity 37% to 62%, PPV 42% to 75%, and NPV 25% to 57% for each sonologist. Sonologists were more confident reporting a positive result (81% vs 51%, p<0.0001), irrespective of the actual presence of FB. The difference between detection rates of 4 types of FB did not reach statistical significance.

Conclusion

Current emergency ultrasound equipment utilized by ED sonologists is unreliable in detection of 3 mm FB in a human extremity soft tissue model. Future studies may further delineate accuracy rates among different sizes and materials of FB.

Copyright information

© Jafari et al; licensee Springer. 2012

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Authors and Affiliations

  • Daniel Jafari
    • 1
  • KJ Cody
    • 1
  • NL Panebianco
    • 1
  • FS Shofer
    • 1
  • BS Ku
    • 2
  • A Au
    • 2
  • AJ Dean
    • 1
  1. 1.Department of Emergency MedicineHospital of the University of PennsylvaniaPhiladelphiaUSA
  2. 2.Department of Emergency MedicineThomas Jefferson University HospitalPhiladelphiaUSA

Personalised recommendations