3D nongadolinium-enhanced mra using flow-sensitive dephasing (fsd) prepared balanced ssfp: identification of the optimal first-order gradient moment

  • Zhaoyang Fan
  • Xiaoming Bi
  • Xiangzhi Zhou
  • Sven Zuehlsdorff
  • James Carr
  • Debiao Li
Open Access
Oral presentation
  • 825 Downloads

Keywords

Readout Direction Vessel Wall Imaging Venous Contamination Balance Ssfp Volunteer Extremity 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

Introduction

A noncontrast MRA technique using ECG-triggered 3D bSSFP with FSD preparation has recently been developed (Fig. 1) [1]. The angiographic quality relies on the blood signal suppression by FSD preparation, which is in turn determined by FSD's first-order gradient moment, m1. A suboptimal m1 may result in venous contamination or incomplete arterial delineation. This work aimed to develop an m1-scout approach to rapidly identify the optimal m1-value.
Figure 1

Schematic of FSD-prepared MRA method (a) and sequence diagrmas (b and c).

Materials and methods

Theory

For a FSD-based 3D isotropic resolution MRA with major flows along the readout direction, we hypothesized that the same imaging sequence can be switched to a 2D mode to rapidly identify the optimal m1 as applied in the readout direction during the 3D dark-artery measurement. This requires the 2D imaging plane be perpendicular to the major vessel of interest, the FSD gradient pulses applied in the slice-select direction, and the in-plane resolution identical to that of 3D MRA.

Imaging

(1) Flow phantom study (Gd-doped water 0.25 mM). Six flow rates (15, 20, 30, . . 60 cm/s) were tested. A 2D FSD-bSSFP scan, as m1-scout, acquired 11 images with incremental m1 values (0.9-mm in-plane and 5-mm slice-thickness) (Fig. 2). 3D FSD-bSSFP imaging was repeated with 6 selected m1-values (0.9-mm isotropic). (2) Volunteer study (3 M, 2 F). Left and right thighs were scanned separately. 2D m1-scout imaging employed m1 = 0, 5, . . 50 mT·ms2/m, respectively. 3D FSD-bSSFP imaging was repeated with 7 selected m1-values.
Figure 2

Schematic of the 2D m 1 -scout approach. A total of 11 images were collected within 1 min. The first uses m1 = 0, while the latter uses incremental m1 values (user specified)

Results

The 2D scout took <1 min as opposed to 4-5 min required for 3D MRA. In the phantom study, the lumen signal intensity from the 2D and 3D images were significantly correlated at all the velocities tested (Pearson correlation = 0.988 ± 0.011, p < 0.001) (Fig. 3). Similar results were observed in the volunteer study (Fig. 4a, Table 1).
Figure 3

2D scout vs. 3D reformatted images (flow phantom).

Figure 4

2D scout vs. 3D reformatted images of superficial femoral arteries from a volunteer (a) and MRA using the optimal m 1 -value (15 mT ms 2 /m). There is a discrepancy at m1 = 25 mT ms2/m (dashed ellipse), likely due to the altered flow pattern or ECG timing

Table 1

Signal intensity Pearson correlation in 5 healthy volunteers

Volunteer (Left/Right)

1

(L)

1

(R)

2

(L)

2

(R)

3

(L)

3

(R)

4

(L)

4

(R)

5

(L)

5

(R)

Pearson Correlation

0.941

0.955

0.974

0.984

0.990

0.992

0.947

0.960

0.924

0.950

p-value

0.002

0.001

<0.001

<0.001

<0.001

<0.001

0.001

0.001

0.003

0.001

Conclusion

The optimal m1 value determined by the 2D scout approach consistently offers high-quality MRA at the volunteer extremities (Fig. 4b, 5, 6). FSD-induced signal suppression is voxel size-dependent as its underlying mechanism is the intravoxel velocity variation [1, 2]. Despite the large pixel size in the 2D scout, this variation is equivalent between the 2D and 3D scans if the velocity is constant along the vessel course. This efficient approach is likely to be beneficial for FSD-based MRA and vessel wall imaging [3].
Figure 5

Superior arterial depiction in calf MRA.

Figure 6

Noncontrast hand MRA (a) has better arterial depiction than contrast MRA (b).

References

  1. 1.
    Fan Z, et al: MRM. 2009Google Scholar
  2. 2.
    Nguyen TD, et al: JMRI. 2008Google Scholar
  3. 3.
    Koktzoglou I, et al: JCMR. 2007Google Scholar

Copyright information

© Fan et al; licensee BioMed Central Ltd. 2010

This article is published under license to BioMed Central Ltd.

Authors and Affiliations

  • Zhaoyang Fan
    • 1
  • Xiaoming Bi
    • 2
  • Xiangzhi Zhou
    • 1
  • Sven Zuehlsdorff
    • 2
  • James Carr
    • 1
  • Debiao Li
    • 1
  1. 1.Northwestern UniversityChicagoUSA
  2. 2.Siemens Medical SolutionsChicagoUSA

Personalised recommendations