Comparison between silicon photomultiplier-based and conventional PET/CT in patients with suspected lung cancer—a pilot study

  • Johan Economou LundebergEmail author
  • Jenny Oddstig
  • Ulrika Bitzén
  • Elin Trägårdh
Open Access
Original research

Abstract

Background

Lung cancer is one of the most common cancers in the world. Early detection and correct staging are fundamental for treatment and prognosis. Positron emission tomography with computed tomography (PET/CT) is recommended clinically. Silicon (Si) photomultiplier (PM)-based PET technology and new reconstruction algorithms are hoped to increase the detection of small lesions and enable earlier detection of pathologies including metastatic spread. The aim of this study was to compare the diagnostic performance of a SiPM-based PET/CT (including a new block-sequential regularization expectation maximization (BSREM) reconstruction algorithm) with a conventional PM-based PET/CT including a conventional ordered subset expectation maximization (OSEM) reconstruction algorithm. The focus was patients admitted for 18F-fluorodeoxyglucose (FDG) PET/CT for initial diagnosis and staging of suspected lung cancer. Patients were scanned on both a SiPM-based PET/CT (Discovery MI; GE Healthcare, Milwaukee, MI, USA) and a PM-based PET/CT (Discovery 690; GE Healthcare, Milwaukee, MI, USA). Standardized uptake values (SUV) and image interpretation were compared between the two systems. Image interpretations were further compared with histopathology when available.

Results

Seventeen patients referred for suspected lung cancer were included in our single injection, dual imaging study. No statically significant differences in SUVmax of suspected malignant primary tumours were found between the two PET/CT systems. SUVmax in suspected malignant intrathoracic lymph nodes was 10% higher on the SiPM-based system (p = 0.026). Good consistency (14/17 cases) between the PET/CT systems were found when comparing simplified TNM staging. The available histology results did not find any obvious differences between the systems.

Conclusion

In a clinical setting, the new SiPM-based PET/CT system with a new BSREM reconstruction algorithm provided a higher SUVmax for suspected lymph node metastases compared to the PM-based system. However, no improvement in lung cancer detection was seen.

Keywords

Lung cancer Digital PET Analogue PET TNM stage BSREM 

Abbreviations

18F-FDG

18F-fluorodeoxyglucose

BSREM

Block-sequential regularization expectation maximization

CT

Computed tomography

D690

Discovery 690

DMI

Discovery MI

EBUS-TBNA

Endobronchial ultrasound with real-time guided transbronchial needle aspiration

OSEM

Ordered subset expectation maximization

PET

Positron emission tomography

ROI

Region of interest

SiPM

Silicon photomultiplier

SNR

Signal-to-noise ratio

SUV

Standardized uptake value

SUVmax

Maximum standardized uptake value

SUVmean

Mean standardized uptake value

Background

Lung cancer is one of the most common cancers in the world with about two million new cases each year [1]. Early detection and correct staging are fundamental for optimal treatment and prognosis [2]. TNM staging in lung cancer is based on the location and extent of the primary tumour (T); the presence of mediastinal, intrapulmonary or hilar lymph node metastases (N); and the presence of distant metastases (M) [3, 4]. Positron emission tomography with computed tomography (PET/CT) with 18F-fluorodeoxyglucose (FDG) is recommended in the initial diagnosis and staging of patients with suspected lung cancer and can help to discriminate between malignant and non-malignant lymph nodes [5]. PET/CT can also evaluate the prognosis of small lung cancer tumours (< 3 cm) using standardized uptake values (SUV) [6].

Recently, a new generation of PET systems was introduced based on silicon (Si) photomultiplier (PM) technology. SiPM-based PET technology is hoped to increase the diagnostic accuracy via a higher spatial resolution and higher sensitivity [7, 8, 9]. Concurrently, a new block-sequential regularization expectation maximization (BSREM) reconstruction algorithm was introduced that allows for fully convergent iterative reconstruction leading to higher image contrast with limited noise [10, 11, 12, 13, 14, 15]. Thus, this combination of new hardware and software could lead to early detection of pathologies including metastatic spread. In 2016, Knopp et al. [16] compared how a digital and an analogue PET/CT system perform looking at malignant and metastatic lung lesions in 20 patients. However, to the best of our knowledge, a comparison including lymph node lesions has not been previously published in patients with suspected lung cancer.

The objective of this study was to compare a SiPM-based time-of-flight system (Discovery MI; GE Healthcare, Milwaukee, WI, USA) using the BSREM reconstruction algorithm (Q.clear; GE Healthcare, Milwaukee, WI, USA) with a conventional PM-based time-of-flight system (Discovery 690; GE Healthcare, Milwaukee, WI, USA) using the ordered maximum likelihood expectation maximization (OSEM) with regard to simplified metabolic TNM staging and lesion SUVmax in patients undergoing 18F-FDG PET/CT due to suspected lung cancer.

Method

Study population

Patients with symptoms that might be due to lung cancer (such as bloody coughs or new dyspnea in a smoker or former smoker of at least 40 years of age) are referred according to a standardized care plan in Sweden. We included patients referred to Skåne University Hospital for PET/CT according to this standardized care plan. Patients were included when double examinations were possible and a previous diagnostic CT was available. Patients below 18 years of age, pregnant, or with a previous history of lung cancer were excluded. The study was approved by the Regional Ethical Review Board in Lund (#2016/417) and the Radiation Protection committee at Skåne University Hospital Sweden (#SSFo2016-018). All patients provided written informed consent.

Patient preparation

Patients fasted for 4 h prior to the examination. The B-glucose levels were measured before injection of 4 MBq/kg 18F-FDG. After injection, the patients rested in a heated room for 60 min.

PET/ CT scanning

Image acquisition was performed on the SiPM-based time-of-flight scanner GE Discovery MI (DMI) and the conventional PM-based time-of-flight scanner GE Discovery 690 (D690). Half of the patients were first examined on the DMI system, and the other half were first examined on the D690 system. A low-dose CT (120 kVp, 30–160 mA, noise index of 45, and slice thicknesses of 2.5 and 3.75 mm for the DMI and D690, respectively) was used without administration of oral or intravenous contrast agent on both systems for attenuation correction and anatomic correlation before the PET acquisition. An adaptive statistical iterative CT-reconstruction technique was used for the DMI, and a filtered back projection reconstruction was used for the D690.

The DMI and D690 have an axial field of view of 20 cm and 15.7 cm, respectively. An overlap of 24% between bed positions was used. PET data from the inguinal region to the base of the skull were acquired. For the D690, the PET acquisition time was 120 s per bed position; the OSEM reconstruction algorithm was 3 iterations and 12 subsets with a 5-mm Gaussian post filter and a 192 × 192 matrix (pixel size 3.6 × 3.6 mm2, slice thickness 3.3 mm). For the DMI, the PET acquisition time was 90 s per bed position, and the BSREM reconstruction algorithm (Q.clear; GE Healthcare, Milwaukee, WI, USA) had a β value of 550 with a 256 × 256 matrix (pixel size 2.7 × 2.7 mm2, slice thickness 2.8 mm). During installation of the DMI, signal-to-noise ratio (SNR) in the liver was used to select the acquisition time and reconstruction parameters in order to yield an equal SNR between the DMI and the D690. Due to different sensitivity for the systems (6.9 cps/kBq and 14 cps/kBq for the D690 and DMI, respectively), the same SNR was obtained at different acquisition times. All reconstructions included time-of-flight and point spread function corrections. Respiratory gating was not used. Both PET-CTs are cross-calibrated to the same dose calibrator, and the calibration is validated monthly in a SUV control with phantoms.

Image interpretation

Images were interpreted by two nuclear medicine physicians with 6 and more than 10 years of experience with PET/CT in lung cancer, respectively. All suspected malignant lesions were recorded to determine a simplified metabolic TNM stage. The metabolic TNM stages were defined as: a primary tumour (T+ or T−), suspected malignant intrathoracic lymph nodes (N+ or N−), or distant metastases (M+ or M−). The reviewers evaluated the scans in a blinded fashion without knowledge of the PET/CT scanner, clinical history, sex, age, and patient outcome. The latest diagnostic CT or referral information was available on request for all patients when the findings were difficult to interpret on the PET examination, for example to better determine if lesions were malignant or inflammatory. Information was given in 7 cases. All PET/CT images were evaluated on an Extended Brilliance Workstation version V4.5.3.40140 (Philips Healthcare, Cleveland, OH, USA).

The histology results from endobronchial ultrasound with real-time-guided transbronchial needle aspiration (EBUS-TBNA), surgery, and biopsies were collected from the patients’ medical records when available and compared to the results from the simplified TNM staging (9 patients).

Quantitative analysis

>SUVmax was measured in the primary tumour and a maximum of 3 intrathoracic suspected malignant lymph nodes with variable sizes. Lymph nodes from different intrathoracic regions were preferred, if available, to allow for representative measurements. Subgroup analysis of SUVmax in lymph nodes smaller than 10 mm was performed (short-axis measured on an axial CT slice).

In addition, the SUVmean was measured in a circular region-of interest (ROI) in the blood pool in the left atrium (ROI 900–1000 mm2) and in the mid-region of the right liver lobe (ROI 3000 mm2), respectively. If liver metastases were present, the ROI was moved to an area with no detectable metastases. The SUVratio was defined as lesion SUVmax/physiological SUVmean and was calculated using blood pool SUVmean and liver SUVmean as physiological references. The SNR was calculated by dividing liver SUVmean by SUVstandard deviation of the liver ROI.

Statistical analysis

Differences in SUV between the SiPM-based and the PM-based PET/CT systems were compared using the Wilcoxon signed-rank test. Continuous variables are presented as mean ± SD. A P value of less than 0.05 was considered statistically significant. Bland-Altman analyses were used to assess the agreement between the two systems. The statistical software GraphPad Prism7 version 7.00 was used.

Results

A total of 17 patients were included (5 women, 12 men; 69 ± 5 years) (see Table 1 for patient information). The injected 18F-FDG dose was 3.99 ± 0.15 MBq/kg. The accumulation time for the first scan was 58 ± 2 min, and the delay to the second examination was 31 ± 6 min. Accumulation time for scans at the DMI system was 75 ± 15 min and for the D690 system was 72 ± 18 min. The B-glucose level before 18F-FDG injection was 114 ± 27 mg/dL.
Table 1

Patient characteristics and scanning setup

Case

Age (decade)

Sex

Weight (kg)

Height (cm)

B-glucose (mg/dL)

18F-FDG-dose (MBq)

Accumulation time DMI (min)

Accumulation time D690 (min)

1

6th

M

101

176

169

399

57

87

2

7th

M

94

171

101

375

98

56

3

7th

M

54

169

114

212

57

81

4

7th

F

70

158

115

281

84

56

5

6th

M

65

178

97

236

59

86

6

6th

F

64

162

123

250

59

85

7

6th

M

94

182

97

368

57

95

8

6th

M

67

173

85

277

58

85

9

6th

M

73

176

132

296

88

55

10

7th

F

49

165

92

194

61

87

11

6th

M

89

183

103

353

57

88

12

7th

M

83

178

87

333

55

85

13

6th

F

37

168

128

164

88

61

14

5th

M

62

176

97

249

107

57

15

7th

M

71

169

180

282

89

62

16

7th

M

74

168

101

294

87

57

17

6th

F

80

165

96

318

91

59

Quantitative image analysis

No statistically significant difference was found between the PET/CT systems regarding SNR (Table 2). The SUVmean in the liver was significantly different, but only slightly lower for the DMI compared to the D690; no statistical difference was found in the blood pool.
Table 2

Physiological, signal-to-noise ratio, and all lesion SUV parameters

SUV parameters (mean ± SD)

Discovery 690

Discovery MI

P value

Change (%)

Signal-to-noise ratio

10.9 ± 2.6

10.2 ± 2.2

0.53

− 6

Liver SUVmean

2.4 ± 0.5

2.2 ± 0.3

*0.01

− 10

Blood pool SUVmean

1.8 ± 0.3

1.7 ± 0.3

0.22

− 7

Primary lesions SUVmax

10.9 ± 8.2

10.3 ± 7.1

0.29

− 6

Primary lesions to blood pool SUVratio

6.6 ± 5.2

6.3 ± 4.6

0.84

− 5

Primary lesions to liver SUVratio

5.0 ± 3.9

4.9 ± 3.5

0.84

− 2

Lymph node SUVmax

6.4 ± 5.5

7.0 ± 5.2

*0.026

10

Lymph node to blood pool SUVratio

3.8 ± 3.5

4.2 ± 3.0

0.11

12

Lymph node to liver SUVratio

2.8 ± 2.6

3.2 ± 2.3

*0.021

17

*Indicating a statistically significant difference

Of the 17 patients, 13 had a suspected malignant primary lesion (seen on either of the systems) with a mean long axis at 45 ± 25 mm and short axis 33 ± 20 mm measured on a trans-axial CT-slice. For the primary lesions, no statistically significant difference was found between the PET/CT systems regarding SUVmax, lesion to blood SUVratio, and lesion to liver SUVratio.

A total of 22 metabolically suspected malignant intrathoracic lymph nodes (seen on either of the systems) with a mean long axis at 12.3 ± 4.2 mm and a short axis at 8.2 ± 2.7 mm were examined. The lesion SUVmax and lesion to liver SUVratio were significantly higher on the DMI compared to the D690 (Table 2); the lesion to blood pool SUVratio did not differ significantly. Figure 1 shows a representative example of differences in SUVmax between the DMI and the D690 for a mediastinal lymph node.
Fig. 1

Axial PET and CT images. An example of axial PET and CT images of a mediastinal lymph node with a SUVmax of 2.7 on the Discovery 690 and 3.6 on the Discovery MI

Furthermore, subgroup analysis was done on 18 of the 22 metabolically suspected malignant intrathoracic lymph nodes with a short axis < 10 mm (mean long axis at 10.9 ± 3.1 mm and short axis 7.1 ± 1.4 mm). SUVmax in these lymph nodes was significantly higher on the DMI compared with the D690 (Table 3). The extra-thoracic metastases were not studied with SUV measurements due to a very low number of suspect lesions (6 lesions in 2 patients). Bland-Altman plots of primary lesion SUVmax and lymph node metastases for the two systems are shown in Fig. 2. The difference between the two PET/CT systems (expressed as absolute difference and difference in percent) was higher for low SUVmax.
Table 3

SUV parameters in lymph nodes < 10 mm

SUV parameters (mean ± SD)

Discovery 690

Discovery MI

P value

Change (%)

Lymph node lesion SUVmax

5.8 ± 5.5

6.3 ± 4.9

0.026*

9

Lymph node lesion to blood pool SUVratio

3.4 ± 3.5

3.8 ± 2.9

0.17

11

Lymph node lesion to liver SUVratio

2.5 ± 2.7

2.9 ± 2.2

0.054

14

*Indicating a statistically significant difference

Fig. 2

Bland-Altman plots comparing SUVmax between the Discovery 690 (D690) and Discovery MI (DMI). Absolute differences (left column) and differences in percent (right column) between SUVmax on D690 and DMI (y-axis) vs. the average of D690 and DMI (x-axis). White circles indicate lesions from patients first examined on the D690 and black circles first examined on the DMI. The DMI produced higher values for SUVmax in all lymph nodes (b) and lymph nodes less than 10 mm on short axis (c) but not in primary lesions (a)

Image interpretation

Overall there was a good consistency between the cameras comparing the simplified TNM-stage, but a different interpretation was seen in three cases (Table 4).
Table 4

Metabolic TNM stage based on the two PET examinations

Case

Staging DMI

Staging D690

Change in staging

Histology results

1

T+N−M−

T+N+M−

YES

Epithelial cancer (S)

2

T+N+M−

T+N+M−

NO

Colon cancer (B)

3

T+N−M−

T−N−M−

YES

Not done

4

T−N−M−

T−N−M−

NO

Adenocarcinoma (S, E)

5

T+N−M−

T+N−M−

NO

Suspect malignant cells (B)

6

T+N+M−

T+N+M−

NO

Non-small cell lung cancer (E)

7

T+N+M+

T+N+M+

NO

Adenocarcinoma (B)

8

T+N+M−

T+N+M−

NO

No malignant cells (Br)

9

T+N+M−

T+N+M−

NO

Non-small cell lung cancer (Br)

10

T+N−M−

T+N−M−

NO

Non-small cell lung cancer (S,E)

11

T+N−M+

T+N−M+

NO

Non-small cell lung cancer (Br)

12

T−N−M−

T+N+M−

YES

No malignant cells (E,B)

13

T+N+M−

T+N+M−

NO

Small cell lung cancer (E)

14

T−N−M−

T−N−M−

NO

Not done

15

T+N+M−

T+N+M−

NO

Epithelial cancer (Br)

16

T−N−M−

T−N−M−

NO

Not done

17

T−N−M−

T−N−M−

NO

Not done

In case of malignant and non-malignant findings, only the malignant histology results are shown above

S surgery, B biopsy, E EBUS-TBNA, Br bronchoscopy with brush/fluid sample

In the first case, the physicians identified an additional lymph node on the D690 that was not seen on the DMI leading to up-scaling of the TNM-stage. However, the histology results showed no malignant lymph nodes indicating a false-positive result on the D690. The second case showed a suspect primary tumour on the DMI that was not seen on the D690. No histology results were available because the treating physician chose to monitor the patient clinically. In the last case, the physicians identified a suspect primary tumour and two lymph nodes on the D690; neither was seen on the DMI. Histology results from EBUS-TBNA showed no malignant cells indicating a false positive on the D690; however, a few reactive cells were seen. All three patients were first examined on the DMI. Neither PET/CT system detected a malignant primary lesion and malignant lymph nodes (adenocarcinoma) in one patient resulting in a false-negative result. Of the 4 cases where no histology was performed, 3 patients were monitored clinically with control CTs within 1–2 years and 1 patient was deemed not needing further follow-ups after the PET/CT examination.

The PET results (both suspected malignant and none malignant findings) were compared to histology results from the EBUS-TBNA, surgery, and biopsies as reference methods (Tables 5 and 6). The histology results from bronchoscopy with brush/fluid samples are not shown due to the uncertain origin of these cells. Similar findings were shown regarding lymph nodes, but two additional false-positive lymph nodes were seen on the Discovery D690.
Table 5

Primary lesion—histology compared to metabolic TNM stage

Discovery 690

Discovery MI

True positive

False positive

True positive

False positive

4

1

4

0

False negative

True negative

False negative

True negative

1

0

1

1

The histology results from the EBUS-TBNA, surgery, and biopsy’s as gold standard compared with PET results

Table 6

Lymph node—histology compared to metabolic TNM stage

Discovery 690

Discovery MI

True positive

False positive

True positive

False positive

2

2

2

0

False negative

True negative

False negative

True negative

2

20

2

22

The histology results from the EBUS-TBNA, surgery, and biopsy’s as gold standard compared with PET results

Discussion

To evaluate the clinical value, we compared the whole PET/CT systems including both hardware and software. Several studies have analysed how the BSREM algorithm affects the images with better image quality, tumour conspicuousness, and image sharpness compared to OSEM [10] as well as a lower background noise and a higher SUVmax. Teoh et al. [17] showed that SUVmax was higher in lung lesions smaller than 10 mm using BSREM but it did not improve the differentiation between malignant and benign lesions. Other studies have compared the hardware of the SiPM-based and the conventional PM-based PET/CT systems in phantom studies using the same reconstruction algorithms [18, 19]. Wagatsuma et al. [19] showed that the Discovery MI had a better contrast compared to the Discovery 710 when the same reconstruction algorithms were used. In 2016, Knopp et al. [16] compared how a digital PET/CT system and an analogue perform looking at malignant and metastatic lung lesions. However, to the best of our knowledge, a comparison including lymph node lesions has not been previously investigated in patients with suspected lung cancer.

Differences in SNR in the liver were not expected since the acquisition time and reconstruction parameters for the DMI were selected to give the same SNR as for the D690 at the installation.

Even though statistically significant, the difference between the PET/CT systems regarding liver SUVmean was relatively small. It cannot be explained by the difference in accumulation time [20] and might be due to a random effect. In the blood pool, we saw only a marginal decrease in SUV that was not statistically significant which might indicate that blood SUVmean is more reliable than liver SUVmean as suggested by Hofheinz et al. [21].

No significant difference in SUVmax was seen for the primary tumours. This is most likely because the primary tumours were relatively large leading to less impact on the different reconstruction algorithms. The higher SUVmax in suspected lymph node metastases (mean of 12 × 8 mm) for the DMI is most likely due to the BSREM algorithm that leads to an increased SUV in small lesions [17].

The new SiPM-based PET/CT systems have a higher spatial resolution and a higher sensitivity and thus may increase the detectability of malignancy in PET/CT [7, 17]. Although the SiPM technology together with BSREM reconstruction yielded higher SUV values for metabolically malignancy suspected lymph nodes, we could not verify that the new system will detect more malignant lesions. Of our three cases with differences in the metabolic TNM stage, two were an up-stage due to findings on the PM-based system and one was up-staged on the SiPM-based system. These three patients were first examined on the DMI so our findings are not explained by the order in which the PET/CT systems were used. We found only very minor differences in a further analysis using histology results from EBUS-TBNA, surgery, and biopsies. This was contrary to our initial beliefs since previous studies indicated an increased sensitivity and decreased specificity [7, 17]. Thus, we have not verified that the new digital PET/CT system will improve the ability to detect malignancy in a clinical setting where both systems had been optimized individually regarding image quality and time efficiency. This might be because the D690 is a relatively new system including time-of-flight and point-spread-function. A larger difference in image quality and SUV is to be expected when comparing SiPM-based PET/CT to older generations of PET-scanners without time-of-flight and point-spread-function correcting capabilities. The greater axial FOV (20 vs 15.7 cm for the DMI and the D690, respectively) and the high sensitivity permit faster image acquisition (90 vs 120 s/bed for the DMI and the D690, respectively). Generally, our experience is that the total PET acquisition time is almost half that for the D690. A short acquisition time potentially reduces problems with patient motion and bladder filling and would allow more patients to be examined per day.

The difference in 18F-FDG accumulation time may have resulted in differences between the two PET scans in each patient. The randomized scanning order for the systems yielded a similar mean accumulation time (75 ± 15 min for the DMI and 72 ± 18 min for the D690 system) implying that this probably did not affect the results for the whole group. Optimization of the reconstruction algorithm settings can be further improved as more studies and clinical experiences are gathered. Furthermore, the small sample size, lack of respiratory gating, and few patients with biopsy-proven disease limits the strength of our study.

Conclusion

This clinical study found that a SiPM-based PET/CT system together with a BSREM reconstruction algorithm yielded higher SUVmax for suspected malignant lymph nodes compared to the conventional PM-based system. However, no improved ability to detect lung cancer was seen. Further larger studies are thus warranted to investigate if the new SiPM-based PET/CT systems as well as improved reconstruction algorithms will impact diagnosis, clinical management, and patient outcome. It was possible to use a shorter acquisition time for the Discovery MI with preserved diagnostic result.

Notes

Acknowledgements

The authors thank the staff at the Department of Clinical Physiology and Nuclear Medicine, Skåne University Hospital, for their help with data collection. We especially thank Annmarie Svensson and Mohammed El-Khatib for the help with including patients in our study.

Funding

This study was funded by generous grants from the Knut and Alice Wallenberg Foundation, governmental funding of the clinical research within the National Health Services, and Skåne University Hospital Cancer Foundation.

Availability of data and materials

The datasets used and analysed during the current study are available from the corresponding author on reasonable request.

Authors’ contributions

All authors made substantial contributions to the conception and design of the work, data analysis, and interpretation of data. UB and ET interpreted the data. JEL performed the statistical analyses and drafted the manuscript. ET, JO, and UB reviewed and revised the manuscript. All authors provided final approval of the version submitted for publication.

Ethics approval and consent to participate

All procedures performed in studies involving human participants were in accordance with the ethical standards of the Institutional Research Committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. Informed consent was obtained from all individual participants included in the study. The study was approved by the Regional Ethical Review Board at Lund University (#2016/417) and the Radiation Protection Committee at Skåne University Hospital (#SSFo2016-018).

Consent for publication

All patients agreed to participate in this study and provided written informed consent.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  1. 1.
    Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2018;68(6):394–424.CrossRefGoogle Scholar
  2. 2.
    Allemani C, Weir HK, Carreira H, Harewood R, Spika D, Wang XS, et al. Global surveillance of cancer survival 1995-2009: analysis of individual data for 25,676,887 patients from 279 population-based registries in 67 countries (CONCORD-2). Lancet. 2015;385(9972):977–1010.CrossRefGoogle Scholar
  3. 3.
    Rami-Porta R, Crowley JJ, Goldstraw P. The revised TNM staging system for lung cancer. Ann Thorac Cardiovasc Surg. 2009;15(1):4–9.PubMedGoogle Scholar
  4. 4.
    Compton CC, American Joint Committee on Cancer. AJCC cancer staging atlas: a companion to the seventh editions of the AJCC cancer staging manual and handbook. New York: Springer; 2012.Google Scholar
  5. 5.
    Cronin P, Dwamena BA, Kelly AM, Carlos RC. Solitary pulmonary nodules: meta-analytic comparison of cross-sectional imaging modalities for diagnosis of malignancy. Radiology. 2008;246(3):772–82.CrossRefGoogle Scholar
  6. 6.
    Deng SM, Zhang W, Zhang B, Chen YY, Li JH, Wu YW. Correlation between the uptake of 18F-fluorodeoxyglucose (18F-FDG) and the expression of proliferation-associated antigen Ki-67 in cancer patients: a meta-analysis. PLoS One. 2015;10(6):e0129028.CrossRefGoogle Scholar
  7. 7.
    Nguyen NC, Vercher-Conejero JL, Sattar A, Miller MA, Maniawski PJ, Jordan DW, et al. Image quality and diagnostic performance of a digital PET prototype in patients with oncologic diseases: initial experience and comparison with analog PET. J Nucl Med. 2015;56(9):1378–85.CrossRefGoogle Scholar
  8. 8.
    Zhang J, Maniawski P, Knopp MV. Performance evaluation of the next generation solid-state digital photon counting PET/CT system. EJNMMI Res. 2018;8(1):97.CrossRefGoogle Scholar
  9. 9.
    Hsu DF, Ilan E, Peterson WT, Uribe J, Lubberink M, Levin CS. Studies of a next generation silicon-photomultiplier-based time-of-flight PET/CT system. J Nucl Med. 2017;58:1511.CrossRefGoogle Scholar
  10. 10.
    Messerli M, Stolzmann P, Egger-Sigg M, Trinckauf J, D'Aguanno S, Burger IA, et al. Impact of a Bayesian penalized likelihood reconstruction algorithm on image quality in novel digital PET/CT: clinical implications for the assessment of lung tumors. EJNMMI Phys. 2018;5(1):27.CrossRefGoogle Scholar
  11. 11.
    Ter Voert E, Muehlematter UJ, Delso G, Pizzuto DA, Muller J, Nagel HW, et al. Quantitative performance and optimal regularization parameter in block sequential regularized expectation maximization reconstructions in clinical (68)Ga-PSMA PET/MR. EJNMMI Res. 2018;8(1):70.CrossRefGoogle Scholar
  12. 12.
    Reynes-Llompart G, Gamez-Cenzano C, Vercher-Conejero JL, Sabate-Llobera A, Calvo-Malvar N, Marti-Climent JM. Phantom, clinical, and texture indices evaluation and optimization of a penalized-likelihood image reconstruction method (Q.Clear) on a BGO PET/CT scanner. Med Phys. 2018;45(7):3214–22.CrossRefGoogle Scholar
  13. 13.
    Teoh EJ, McGowan DR, Schuster DM, Tsakok MT, Gleeson FV, Bradley KM. Bayesian penalised likelihood reconstruction (Q.Clear) of (18)F-fluciclovine PET for imaging of recurrent prostate cancer: semi-quantitative and clinical evaluation. Br J Radiol. 2018;91(1085):20170727.CrossRefGoogle Scholar
  14. 14.
    Lindstrom E, Sundin A, Trampal C, Lindsjo L, Ilan E, Danfors T, et al. Evaluation of penalized-likelihood estimation reconstruction on a digital time-of-flight PET/CT scanner for (18)F-FDG whole-body examinations. J Nucl Med. 2018;59(7):1152–8.CrossRefGoogle Scholar
  15. 15.
    Teoh EJ, McGowan DR, Macpherson RE, Bradley KM, Gleeson FV. Phantom and clinical evaluation of the Bayesian penalized likelihood reconstruction algorithm Q.Clear on an LYSO PET/CT system. J Nucl Med. 2015;56(9):1447–52.CrossRefGoogle Scholar
  16. 16.
    Knopp M, Wright C, Binzel K, Zhang J, Saif T, Siva A, et al. Precision nuclear medicine enabled by ultra-high definition PET: improved lesion detectability using digital PET technology in patients with malignant/metastatic lung lesions. J Nucl Med. 2016;57(supplement 2):246.Google Scholar
  17. 17.
    Teoh EJ, McGowan DR, Bradley KM, Belcher E, Black E, Gleeson FV. Novel penalised likelihood reconstruction of PET in the assessment of histologically verified small pulmonary nodules. Eur Radiol. 2016;26(2):576–84.CrossRefGoogle Scholar
  18. 18.
    Nemoto R, Miwa K, Wagatsuma K, Yamao T, Kamitaka Y. Comparison between SiPM-based and PMT-based TOF-PET/CT to detect sub-centimeter spheres. J Nucl Med. 2018;59(supplement 1):1857.Google Scholar
  19. 19.
    Wagatsuma K, Miwa K, Sakata M, Oda K, Ono H, Kameyama M, et al. Comparison between new-generation SiPM-based and conventional PMT-based TOF-PET/CT. Phys Med. 2017;42:203–10.CrossRefGoogle Scholar
  20. 20.
    Chin BB, Green ED, Turkington TG, Hawk TC, Coleman RE. Increasing uptake time in FDG-PET: standardized uptake values in normal tissues at 1 versus 3 h. Mol Imaging Biol. 2009;11(2):118–22.CrossRefGoogle Scholar
  21. 21.
    Hofheinz F, Bütof R, Apostolova I, Zöphel K, Steffen IG, Amthauer H, et al. An investigation of the relation between tumor-to-liver ratio (TLR) and tumor-to-blood standard uptake ratio (SUR) in oncological FDG PET. EJNMMI Res. 2016;6(1):19.CrossRefGoogle Scholar

Copyright information

© The Author(s). 2019

Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Authors and Affiliations

  1. 1.Department of Clinical Physiology and Nuclear MedicineSkåne University HospitalLundSweden
  2. 2.Radiation PhysicsSkåne University HospitalLundSweden
  3. 3.Wallenberg Center for Molecular MedicineLund UniversityLundSweden

Personalised recommendations