Abstract
Measurement data (MD) converters with negative and positive feedback from series-connected aperiodic and pulse circuits are designed. Their static errors and correction for multiplicative and additive errors are studied. Stability of MD converters is investigated and a method of enhancing it by a proper choice of operation mode for the pulse circuit is described. Different variants of the structural circuit of the MD converter are described. Results of an experiment with a pilot MD converter are presented.
Similar content being viewed by others
REFERENCES
Popov, V.S. and Yampol'skii, Yu.S., Instrumentation Amplifier, Patent 253235, Byul. Izobr., 1970, no. 30.
Popov, V.S. and Yampol'skii, Yu.S., A Highly Stable Instrumentation Amplifier, Pribory Sist. Upravlen., 1972, no. 2.
Tuz, Yu.M., Strukturnye metody povysheniya tochnosti izmeritel'nykh ustroistv (Structural Methods of Enhancement of Accuracy of Measuring Devices), Kiev: Vishcha Shkola, 1976.
Popov, E.P., Avtomaticheskoe regulirovanie (Automatic Adjustment), Moscow: GITTL, 1957.
Gutnikov, V.S., Integral'naya elektronika v izmeritel'nykh ustroistvakh (Integral Electronics in Measuring Devices), Leningrad: Energoatomizdat, 1988.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Popov, V.S., Smirnov, K.I. Error Compensation for Measurement Data Converters by Combined Feedback. Automation and Remote Control 65, 671–686 (2004). https://doi.org/10.1023/B:AURC.0000023544.15050.f5
Issue Date:
DOI: https://doi.org/10.1023/B:AURC.0000023544.15050.f5