Objective. An algorithm to improve the response time of a clinical respiratory multigas analyser is presented. Methods. The algorithm involves the application of a second order differential equation to the analyser gas output signals in real-time. The adjusted analyser output signals are compared with those of a quadrupole respiratory mass spectrometer sampling and analysing simultaneously. Results. Our results show a close correlation between the adjusted clinical gas analyser and the mass spectrometer signals. Lung volumes derived from a non-invasive sinusoidal inert gas forcing technique, in a model test lung, using the adjusted clinical gas analyser and the mass spectrometer signals demonstrated comparable results. Conclusions. The algorithm provides an improvement on the relatively slow response times of the clinical gas analyser for breath-by-breath time-dependent applications. The same algorithm can also be applied to other instruments which have slow response times.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
REFERENCES
Sykes MK. Gas and vapour analysis. In: Sykes MK, Vickers MD, Hull CJ, eds. Principles of Measurement and Monitoring in Anaesthesia and Intensive Care. Oxford: Blackwell Scientific Publications 1985; 237-240.
Williams EM, Gavaghan DJ, Oakley PA, Sainsbury MC, Xiong L, Black AM, Hahn CEW. Measurement of dead-space in a model lung using an oscillating argon signal. Acta Anaesthesiol Scand 1994; 38: 126-129.
Williams EM, Aspel JB, Burrough SM, Ryder WA, Sainsbury MC, Sutton L, Xiong L, Black AM, Hahn CEW. Assessment of cardiorespiratory function using oscillating inert gas forcing signals. J Appl Physiol 1994; 76: 2130-2139.
Williams EM, Hahn CEW. Measurement of cardio-respiratory function using single frequency inspiratory gas concentration forcing signals. Adv Exp Med Biol 1994; 361: 187-195.
Gavaghan DJ, Hahn CEW. A mathematical evaluation of the alveolar amplitude response technique. Resp Physiol 1995; 102: 105-120.
Williams EM, Gale LB, Oakley PA, Sainsbury MC, Hahn CEW. Development of a concentric water-displacement model lung. J Biomed Eng 1993; 15: 420-424.
Mitchell RR. Incorporating the gas analyser response time in gas exchange computations. J Appl Physiol 1979; 47: 1118-1122.
Noguchi H, Ogushi Y, Yoshiya I, Itakura N, Yamabayashi H. Breath-by-breath VCO2 and VO2 require compensation for transport delay and dynamic response. J Appl Physiol 1982; 52: 79-84.
Arieli R, Van Liew HD. Corrections for the response time and delay of mass spectrometers. J Appl Physiol 1981; 51: 1417-1422.
Bates JHT, Prisk GK, Tanner TE, McKinnon AE. Correcting for the dynamic response of a respiratory mass spectrometer. J Appl Physiol 1983; 55: 1015-1022.
Tavener SJ, Withy SJ, Harris EA. Response characteristics of a mass spectrometer. Med Biol Eng Comp 1984; 22: 493-498.
Barton SA, Sutton L, Hahn CEW, Black A. A digital system for generating dynamic sinusoidal gas concentration signals. J Appl Physiol 1988; 65: 945-949.
Gavaghan DJ, Hahn CEW. A tidal breathing model of the forced inspired inert gas sinewave technique. Respir Physiol 1996; 106: 209-221.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Wong, L., Hamilton, R., Palayiwa, E. et al. A Real-Time Algorithm to Improve the Response Time of a Clinical Multigas Analyser. J Clin Monit Comput 14, 441–446 (1998). https://doi.org/10.1023/A:1009941900141
Issue Date:
DOI: https://doi.org/10.1023/A:1009941900141