Journal of Medical and Biological Engineering

, Volume 37, Issue 6, pp 800–809 | Cite as

Minimizing a Wireless Passive LC-Tank Sensor to Monitor Bladder Pressure: A Simulation Study

  • Jacob Melgaard
  • Johannes J. Struijk
  • Nico J. M. Rijkhoff
Original Article
  • 59 Downloads

Abstract

In this simulation study, a wireless passive LC-tank sensor system was characterized. Given the application of continuous bladder monitoring, a specific system was proposed in terms of coil geometries and electronic circuitry. Coupling coefficients were spatially mapped by simulation, as a function of both coil distance, and longitudinal and transverse translation of the sensor relative to the antenna. Further, two interrogation schemes were outlined. One was an auto-balancing bridge for computing the sensor-system impedance. In this case, the theoretical noise limit of the analogue part of the system was found by simulations. As the full system is not necessary for obtaining a pressure reading from the sensor, a simplified circuit more suited for an implantable system was deduced. For this system, both the analogue and digital parts were simulated. First, the required ADC resolution for operating the system at a given coupling was found by simulations in the noise-free case. Then, for one selected typical operational point, noise was added gradually, and through Monte-Carlo type simulations, the system performance was obtained. Combining these results, it was found that it at least is possible to operate the proposed system for distances up to 12 mm, or equivalently for coupling coefficients above 0.005. In this case a 14 bit ADC is required, and a carrier SNR of 27 dB can be tolerated.

Keywords

Urinary bladder Closed loop stimulation Pressure sensor Wireless sensor 

References

  1. 1.
    Hansen, J., Media, S., Nohr, M., Biering-Sørensen, F., Rijkhoff, N. J. M., & Sinkjaer, T. (2005). Treatment of neurogenic detrusor overactivity in spinal cord injured patients by conditional electrical stimulation. Journal of Urology, 173, 2035–2039.CrossRefGoogle Scholar
  2. 2.
    Melgaard, J., & Rijkhoff, N. J. (2011). Detecting the onset of urinary bladder contractions using an implantable pressure sensor. IEEE Transactions on Neural Systems and Rehabilitation Engineering, 19, 700–708.CrossRefGoogle Scholar
  3. 3.
    Takayama, K., Takei, M., Soejima, T., & Kumazawa, J. (1987). Continuous monitoring of bladder pressure in dogs in a completely physiological state. British Journal of Urology, 60, 428–432.CrossRefGoogle Scholar
  4. 4.
    Mills, I. W., Noble, J. G., & Brading, A. F. (2000). Radiotelemetered cystometry in pigs: Validation and comparison of natural filling versus diuresis cystometry. The Journal of Urology, 164, 1745–1750.CrossRefGoogle Scholar
  5. 5.
    Gerber, M. T. (2007). Transmembrane sensing device for sensing bladder condition. US2007/0027494 A1.Google Scholar
  6. 6.
    Shaw, G. L., Choong, S. K., & Fry, C. (2005). Encrustation of biomaterials in the urinary tract. Urological Research, 33, 17–22.CrossRefGoogle Scholar
  7. 7.
    Stickler, D. J. (2012). Surface coatings in urology. In M. Driver (Ed.), Coatings for biomedical applications (pp. 304–335). Cambridge: Woodhead Publishing.CrossRefGoogle Scholar
  8. 8.
    Koldewijn, E. L., van Kerrebroeck, P. E. V., Schaafsma, E., Wijkstra, H., Debruyne, F. M. J., & Brindley, G. (1994). Bladder pressure sensors in an animal model. Journal of Urology, 151, 1379–1384.CrossRefGoogle Scholar
  9. 9.
    Park, E. C., Yoon, J., & Yoon, E. (1998). Hermetically sealed inductor-capacitor (LC) resonator for remote pressure monitoring. Japanese Journal of Applied Physics, 37, 7124–7128.CrossRefGoogle Scholar
  10. 10.
    Puers, R., Vandevoorde, G., & de Bruyker, D. (2000). Electrodeposited copper inductors for intraocular pressure telemetry. Journal of Micromechanics and Microengineering, 10, 124–129.CrossRefGoogle Scholar
  11. 11.
    Akar, O., Akin, T., & Najafi, K. (2001). A wireless batch sealed absolute capacitive pressure sensor. Sensors and Actuators A, 95, 29–38.CrossRefGoogle Scholar
  12. 12.
    DeHennis, A., & Wise, K. D. (2002). A double-sided single-chip wireless pressure sensor. In The fifteenth IEEE international conference on micro electro mechanical systems (pp. 252–255).Google Scholar
  13. 13.
    Fonseca, M. A., English, J. M., von Arx, M., & Allen, M. G. (2002). Wireless micromachined ceramic pressure sensor for high-temperature applications. Journal of Microelectromechanical Systems, 11, 337–343.CrossRefGoogle Scholar
  14. 14.
    Chen, P. J., Rodger, D. C., Saati, S., Humayun, M. S., & Tai, Y. C. (2008). Implantable parylene-based wireless intraocular pressure sensor. In IEEE 21st international conference on micro electro mechanical systems (MEMS) (pp. 58–61).Google Scholar
  15. 15.
    Lotz, H. T., Remeijer, P., van Herk, M., Lebesque, J. V., de Bois, J. A., Zijp, L. J., et al. (2004). A model to predict bladder shapes from changes in bladder and rectal filling. Medical Physics, 31(6), 1415–1423. doi: 10.1118/1.1738961.CrossRefGoogle Scholar
  16. 16.
    Fokdal, L., Honoré, H., Høyer, M., Meldgaard, P., Fode, K., & von der Maase, H. (2004). Impact of changes in bladder and rectal filling volume on organ motion and dose distribution of the bladder in radiotherapy for urinary bladder cancer. International Journal of Radiation Oncology Biology Physics, 59, 436–444.CrossRefGoogle Scholar
  17. 17.
    Black, J. (1992). Biological performance of materials: Fundamentals of biocompatibility. New York: Dekker.Google Scholar
  18. 18.
    Lizon-Martinez, S., Giannetti, R., Rodriguez-Marrero, J. L., & Tellini, B. (2005). Design of a system for continuous intraocular pressure monitoring. IEEE Transactions on Instrumentation and Measurement, 54, 1534–1540.CrossRefGoogle Scholar
  19. 19.
    Collins, C. C. (1967). Miniature passive pressure transensor for implanting in the eye. In IEEE Transactions on Biomedical Engineering, vol. BME-14, pp. 74–83.Google Scholar
  20. 20.
    Hall, H. P. (1980). Method of and apparatus for automatic measurement of impedance or other parameters with microprocessor calculation techniques, 04/01.Google Scholar
  21. 21.
    Finkenzeller, K. (2010). RFID handbook: Fundamentals and applications in contactless smart cards, radio frequency identification and near-field communication. Chichester: John Wiley Sons Ltd.CrossRefGoogle Scholar
  22. 22.
    Mohan, S. S., Hershenson, M. D. M., Boyd, S. P., & Lee, T. H. (1999). Simple accurate expressions for planar spiral inductances. IEEE Journal of Solid-State Circuits, 34, 1419–1424.CrossRefGoogle Scholar
  23. 23.
    O’Rahilly, R., Müller, F., Carpenter, S., & Swenson, R. (2008). Basic human anatomy: A regional study of human structure. Retrieved from http://www.dartmouth.edu/~humananatomy/.

Copyright information

© Taiwanese Society of Biomedical Engineering 2017

Authors and Affiliations

  1. 1.Department of Health Science and Technology, The Faculty of MedicineAalborg UniversityAalborg ØDenmark

Personalised recommendations