Introduction

Globally, the incidence and prevalence of both types 1 and 2 diabetes mellitus (DM) have risen dramatically over the past two decades and they are expected to keep rising in the future. Diabetes is associated with various chronic complications that result in increased morbidity and mortality [1, 2].

In addition to persons with diabetes receiving appropriate and timely medical management, they also require assessment of their glycemic control. The monitoring of glycemic status is considered to be a cornerstone of diabetes care. Analysis of the glucose data provides an assessment of the efficacy of therapy and guides adjustments in lifestyle and medications to achieve the best possible blood glucose control in a safe manner. Primary techniques to assess the effectiveness of glycemic control include patient self-monitoring of blood glucose (SMBG) and measurement of hemoglobin A1c (HbA1c) [3].

The results of the Diabetes Control and Complications Trial (DCCT) and other studies provide reliable evidence on the health benefits of normal or near-normal blood glucose levels, especially in patients receiving insulin treatment [4]. SMBG enables assessment of an individual’s response to therapy. In a study of insulin-naive patients with suboptimal initial glycemic control, structured SMBG performed on 3 consecutive days reduced HbA1c by 0.3% compared with the control group [5]. Results from meta-analyses suggest that SMBG can reduce HbA1c by 0.25–0.3%. However, SMBG alone does not lower blood glucose levels. To be useful, the information must be integrated into clinical and self-management programs [3]. There are additional limitations. SMBG shows a single “point-in-time” measurement and does not provide any data on the direction or rate of change of glucose levels. In addition, obtaining adequate data is dependent upon the patient’s adherence to self-monitoring. Consequently, SMBG often fails to detect nocturnal and/or asymptomatic hypoglycemia [6, 7].

HbA1c represents the percentage of circulating hemoglobin that is glycated. Glycation is a non-enzymatic process and is a measure of glucose levels over time [8]. As a biomarker, it reflects the average plasma glucose over the previous 8–12 weeks [9]. It is currently used for both the diagnosis and management of diabetes [8] and is recommended as a gold standard in the assessment of diabetes-related outcomes [9, 10]. Historically, the elevated level of HbA1c in diabetic patients was reported first by Rahbar and associates in 1968, and over the next decades it became arguably the most important indicator of blood glucose control. It is widely used to judge the adequacy of diabetes treatment [11]. Nevertheless, for a given HbA1c, there is a wide range of mean glucose concentration values, and for any given mean glucose value there is a wide range of HbA1c values [12].

The test is easy to administer, cost-effective, and relatively inexpensive [13]. Furthermore, there is a curvilinear relationship between HbA1c and microvascular complications. Three landmark trials in patients with type 2 DM (ACCORD, ADVANCE, and VADT) showed that lower HbA1c levels are associated with reduced onset or progression of some microvascular complications [14,15,16]. To the contrary, elevated HbA1c is also regarded as an independent risk factor for coronary heart disease and stroke. One of the major limitations of HbA1c is that it does not reflect glycemic excursions, i.e., intra-day and inter-day glycemic changes, which have been linked to both microvascular and macrovascular complications [17]. Furthermore, it is an unreliable measure in patients with anemia, certain hemoglobinopathies, liver disease, and iron deficiency. Moreover, it does not provide detailed information on how to adjust the treatment regimen.

The literature suggests that ethnic and racial differences exist in the degrees of glycation that affect the accuracy of HbA1c measurements. For example, HbA1c concentration is higher (by 0.2–0.4%) in US African Americans and Hispanic populations than in Caucasians. The impact of age and race on HbA1c is currently under discussion. Some studies show that the HbA1c concentration increases by approximately 1 mmol/mol (0.1%) per decade of life and that this age effect results from decreasing red blood cell count with age [18, 19]. Despite these limitations, the diabetes research community has made considerable progress in agreeing upon core glycemic outcomes beyond HbA1c to overcome the above limitations [10].

In this review, we address the latest evidence for the use of alternative metrics of glucose control in patients with diabetes. We also discuss the impact of novel metrics on diabetes-related complications, the usefulness of CGM in the treatment of type 2 DM, and the impact of newer glucose-lowering agents on alternative glycemic metrics beyond HbA1c.

This article is based on previously conducted studies and does not contain any studies with human participants or animals performed by any of the authors.

Continuous Glucose Monitoring

Continuous glucose monitoring (CGM) does not measure blood sugar concentrations, rather it measures interstitial glucose levels. CGM data for a period of 10–14 days provide a good estimate of CGM metrics for a 3-month period. HbA1c can be estimated (eA1c) if 70% of the CGM data are available. Real-time CGM (rtCGM) data can help individuals learn how dramatically their blood glucose level can rise after certain meals or how it is affected by stress or exercise [20]. Therefore, CGM seems to be the best example of precision medicine in diabetes [12]. However, interstitial glucose readings have a time lag of approximately 15 min compared to blood glucose readings; consequently, CGM results do not always match finger stick blood glucose readings.

Different definitions are currently used for CGM systems. Intermittently viewed CGM (iCGM), known comercially as flash glucose monitoring (FGM), shows continuous glucose measurements retrospectively. This type of monitoring may be described as a separate entity from CGM, occupying a position somewhat between a traditional blood glucose meter and a CGM system. On the other hand, real-time CGM (rtCGM) provides real-time data on glucose trends, direction, and rate of change [17]. Although both systems provide the means to move beyond the HbA1c measurement as the sole marker of glycemic control, utilization of CGM as a biomarker of blood glucose levels has remained fairly limited to date [21].

Riddlesworth et al. [22], and the “Beyond A1c Writing Group” [10] recently declared that HbA1c does not accurately reflect the glucose pattern at the individual level. Furthermore, the Steering Committee of decision-making for the “Type 1 Diabetes Outcomes Program” recommended the use of defined clinically meaningful outcomes beyond HbA1c in research, development, and evaluation of therapies for type 1 DM. The novel CGM outcomes include hypoglycemia, hyperglycemia, time in glycemia range, diabetic ketoacidosis, and patient-reported outcomes [23]. A summary of the new definitions of clinically meaningful outcomes is shown in Table 1.  

Table 1 New definitions of hypoglycemia, hyperglycemia, and time in glycemic range

Apart from providing the mean glucose concentration, CGM profiles provide additional details on the patterns of glycemic excursions, as well as potentially dangerous high or low glucose concentrations that are often missed with SMBG [21, 22, 24]. Examples of some profiles with the same HbA1c are shown in Fig. 1.

Fig. 1
figure 1

Different patterns of glycemic variability (GV) in two patients with same hemoglobin A1C (HbA1c). 15-day glucose traces of two patients who had identical HbA1c of 8.0% but different degrees of GV. High GV in patient 1 was reflected by numerous episodes of both hypo- and hyperglycemia (a), whereas low GV in subject 2 resulted in no such episodes (b). Patient 1 (a) had visibly higher glucose fluctuations than patient 2 (b) that resulted in seven episodes of moderate hypoglycemia (≤ 50 mg/dL) and eight episodes of moderate hyperglycemia (≥ 350 mg/dL)

Advantages and Limitations of CGM

Continuous glucose monitoring provides valuable information unattainable by SMBG, including coverage of the individual data throughout the glucose monitoring, detection of glycemic variability (GV) and the rate of change over time, as well as time spent in hypo- and hyperglycemia. CGM can be considered for patients with severe or nocturnal hypoglycemia, especially those with hypoglycemia unawareness [21]. Also, the GOLD [25] and DIAMOND [26] trials showed that among patients with type 1 DM treated with multiple daily insulin injections, the use of CGM resulted in better glycemic control than did conventional treatment. Moreover, several studies have shown that the use of CGM can improve the mean amplitude of glycemic excursion (MAGE) and result in better glycemic control in persons with type 2 DM [27,28,29,30]. A recent meta-analysis of randomized controlled trials conducted on the utility of real-time and retrospective CGM in patients with type 2 DM showed that CGM in these patients could reduce HbA1c levels and time spent with hypoglycemia [31]. Finally, CGM data can be used as a valuable tool for patient education [21].

Despite these advantages, CGM has a number of limitations, namely reimbursement issues, the periodic replacement of sensors for long-duration implantable CGM devices, and the need for development of clinical guidelines on the role of CGM in the management of people with type 2 DM [21, 32]. However, recent advancements in technology and relevant evidence from the literature have addressed several of these issues. A new generation of insulin pumps with automated suspension of insulin infusion in response to observed or predicted hypoglycemia as well as the development of closed-loop insulin delivery systems are expected to dramatically increase the clinical utility and impact of CGM [21].

Time in Range and Hypoglycemia

Both rtCGM and iCGM facilitate monitoring of time spent in the target blood glucose range, referred to as the “time in range.” However, only rtCGM can warn users if the blood glucose level is trending toward hypoglycemia or hyperglycemia. Using iCGM data, the trends can only be viewed retrospectively [17]. In addition, retrospective analysis of CGM data quantifies patterns of hypo- and hyperglycemia and GV. In short, CGM helps individuals with diabetes and clinicians to personalize their management strategies [22]. The time in range target is becoming the new standard for patients and healthcare providers [23]. The use of CGM is associated with increased time in range and reduced incidence of severe hypoglycemic events [33].

eA1C and Glucose Management Indicator

eA1c and Measured Laboratory HbA1c

With the increasing use of CGM to facilitate safe and effective diabetes management, it is important to understand how CGM metrics, namely eA1c, correlates with and can be used as a metric instead of or in addition to HbA1c. eA1c for every single patient is calculated based on CGM readings from a population of individuals. In some patients, the results of a laboratory-measured HbA1c and eA1c are approximately the same, but in others, the eA1c may be either higher or lower than the measured HbA1c. This issue is discussed in a recent publication by Beck et al. [12] who suggested that a laboratory-measured HbA1c of 8.0% can be associated with a range of eA1c of 7–8.5%. The discordance can be confusing for both patients and clinicians. For this reason, the US Food and Drug Administration determined that the nomenclature of eA1c needs to be changed, leading to the generation of a new index: the glucose management indicator (GMI). The results from the HypoDE study and from a prior study were used to generate a pooled equation to convert CGM-derived mean glucose to the GMI [22, 34]:

$${\text{GMI}}\, (\%) = 3.31 + 0.02392 \times {\text{mean}}\;{\text{glucose}}\;{\text{in}}\;{\text{mg}}/{\text{dL}} .$$

Based on this formula, an increment of 25 mg/dL in mean glucose corresponds to a 0.6% increase in GMI [22].

GMI Versus Laboratory-Measured HbA1c

The difference between the GMI and laboratory-measured HbA1c in 528 individuals with diabetes who had both values measured concurrently was evaluated by Bergenstal et al. and showed in 19% of the measurements, the GMI, calculated from CGM-derived mean glucose, and laboratory-measured HbA1c have identical values; however, in 28% of the measurements, the difference between the two values is > 0.5%. Therefore, the introduction of GMI as a new glucose index could be an important step in the direction of a more personalized diabetes management program. If a person has a GMI that is always considerably lower than a measured HbA1c, the healthcare professional has to be careful not to set the therapeutic goal too low to overcome the risk of hypoglycemia. On the other hand, if the GMI is higher, it may be safe to set the HbA1c target slightly lower to minimize excessive hyperglycemia [20, 33]. It should be noted that current evidence indicates that the difference in laboratory-measured HbA1c and GMI remains relatively stable for each individual over time.

Glycemic Variability

Hypoglycemia and glucose variability are considered to be the major challenges during optimization of glycemic control [35, 36]. Individualization of glycemic targets, namely HbA1c, may not always translate into improved clinical outcomes. Thus, glycemia metrics beyond HbA1c may be used to predict the risk of diabetes-related complications. In this context, CGM metrics can be associated with micro- and macrovascular complications in diabetic patients. For example, GV is associated with an increased risk of adverse cardiovascular outcomes due primarily to hypoglycemia [37]. Furthermore, it has been reported that patients with more advanced diabetic retinopathy spend significantly less time in the target range and have higher measures of GV [38].

A simple definition of GV is the degree by which a patient’s blood glucose level fluctuates between high and low levels [39]. The percentage coefficient of variation (%CV) is defined as: %CV = [(standard deviation of glucose)/(mean glucose)] × 100. It is characterized by the amplitude, frequency, and duration of fluctuations in blood glucose [40, 41], with a larger magnitude of GV being associated with a higher incidence of hypoglycemia [42]. To define the threshold for GV, Monnier et al. [41] showed that a percentage coefficient of variation for glucose of 36% appears to be a suitable threshold to distinguish between stable and unstable glycemia in diabetes. It appears that beyond the 36% limit, the frequency of hypoglycemia is significantly increased, especially in insulin-treated subjects [17]. CGM profile and mean glucose and glycemic variability are showed in Fig. 2

Fig. 2
figure 2

The electronic CGM profile

Many factors affect GV in patients, including lifestyle [43], diet, the presence of comorbidities, diabetes treatment [44], and even insulin injection technique [45]. The best method by which to assess GV is still a matter of debate.

Types of GV in terms of duration vary from the short term to the long term. In routine clinical practice, GV is often reported according as follows. Short-term GV is defined as approximately GV of 24–72 h duration; intermediate-term GV, as GV of 3 days to 1 month duration; and long-term GV, as 1 month to years, especially referring to variations in HbA1c determined at repeated clinic visits.

GV may contribute to diabetes-related complications. Several studies have reported that long-term GV, i.e., longitudinal variations in HbA1c, are related to micro- and macrovascular complications. However, the association of varying duration and magnitude of GV with diabetes-related outcomes is still unclear. Further clinical trials that focus on the mechanism of GV are required to evaluate whether GV can be a valuable therapeutic target in patients with type 2 DM [43].

The Atherosclerosis Risk in Communities (ARIC) study [46] explored the association between short-term GV and diabetes-related complications by evaluating the associations of 1, 5-anhydroglucitol (1, 5-AG) with risk of diabetes and microvascular disease. A blood glucose level of > 180 mg/dL is associated with lower 1, 5-AG levels. The authors of this study showed that low levels of l, 5-AG were associated with an increased risk of retinopathy and incident chronic kidney disease. Furthermore, short-term GV was associated with a range of complications, such as hypoglycemia, intensive care unit mortality, cognitive impairment, reduced quality of life, and negative mood [46,47,48,49].

In addition, intra-day and inter-day GV are significantly associated with the risk of hypoglycemia in insulin-treated patients with type 2 DM, even after adjusting for mean blood glucose value and HbA1c. The intra-day GV prior to initiating insulin therapy is significantly associated with the risk of hypoglycemia during insulin treatment [17]. On the other hand, Feng-fei Li et al. showed that higher HbA1c values are associated with a higher 24-h mean amplitude of glycemic excursions compared with those found at lower and moderate HbA1c values. These results led these authors to suggest that patients with higher HbA1c values should receive special therapy aimed at reducing the larger GV [50].

Large fluctuations in blood glucose are dangerous and should be considered to be an important treatment target. Achieving this treatment targets will require the use of newer glucose-lowering agents (such as incretin hormones or sodium–glucose cotransporter 2 [SGLT2]), more rapid-acting prandial insulins, stable (“flat”) long-acting insulins, and more routine use of CGM in patients with type 1 and type 2 DM [36].

Average glucose, GMI, time spent in hyperglycemia, and time spent in the target range are general indicators of overall glucose management. However, additional metrics, such as time spent in hypoglycemia and GV, supply the patient and healthcare provider with critical information beyond the A1C value, by providing objective data on various safety concerns that need to be addressed in the diabetes management plan [22].

Improvement of GV and Time Spent in Glycemic Target

Recent evidence supports the use of certain glucose-lowering agents, namely, novel long-acting basal insulins and dual sodium–glucose co-transporter-1 and -2 inhibitors employed in treating type 1 DM, to improve GV and the time spent in target range. Recently, Bergenstal et al. compared glucose control in participants with type 1 DM receiving insulin glargine 300 units/mL (Gla-300) or glargine 100 units/mL (Gla-100). They showed less increase in CGM-based glucose levels in the last 4 h of the 24-h injection interval, smoother average 24-h glucose profiles with less GV irrespective of injection time, and reduced nocturnal hypoglycemia with Gla-300 compared to Gla-100 [51].

According to the DEVOTE trial, insulin degludec, as an ultra-long-acting, once-daily formulation of basal insulin, was similar to insulin glargine in improving glycemic control, but with a lower risk of severe hypoglycemia [52]. Insulin degludec has also been reported to be associated with lower day-to-day variation in glucose level than insulin glargine [53]. Also, a lower incidence of global and nocturnal hypoglycemic events has been reported in patients with high GV being treated with insulin degludec compared with those receiving insulin glargine [54]. Furthermore, the authors of the VARIATION study reported the lowest GV and lowest hypoglycemia in patients with type 2 DM using the combination of basal insulin with a glucagon-like peptide-1 receptor agonist (GLP-1RA) [55]. In another study, treatment with exenatide led to an improvement in glycemic excursions in patients with type 2 DM [56].

In a pooled analysis, combination treatment with sotagliflozin and insulin significantly increased the time in range and reduced post-prandial glucose levels, meeting type 1 DM glycemic control endpoints beyond a reduction in HbA1C levels [57]. Meanwhile, in type 2 DM, a randomized controlled trial comparing dapagliflozin with placebo showed an improvement of glycemic control and variability, using a 7-day assessment of the glycemic profile obtained from CGM [58]. In addition, dapagliflozin may reduce GV and oxidative stress in newly diagnosed type 2 DM [59].

Conclusion

Hemoglobin A1c is the current gold standard for assessing glycemic control. However, this index only provides an average measure of glycemic status over a period of 2–3 months. In addition, HbA1c does not address GV and hypoglycemia. In order to personalize the treatment decision, healthcare providers need to standardize the glycemic data, view the data, and use the data to achieve increased efficacy and safety for patients.

The CGM metrics provide a more personalized approach to diabetes management and resolve most of the limitations of HbA1c. CGM detects within-day and day-to-day GV, time in glycemic target, and time in hypoglycemia. These metrics may enhance a patient’s self-management of diabetes.

The degree of GV is potentially associated with the frequency, duration, and severity of hypoglycemic events. Moreover, GV may be linked to the pathogenesis of diabetes complications and could impact patient management and quality of life.

Newer glucose-lowering agents (SGLT inhibitors or GLP-1RAs) and novel long-acting insulin analogs with smoother pharmacodynamic profiles, and ultra-fast-acting prandial insulins could reduce the degree of GV. Randomized trials are needed to examine the relationship between new glycemic metrics and hard endpoints, such as retinopathy, nephropathy, or cardiovascular outcomes.