Introduction

The prevalence of diabetes is high in emerging countries or regions such as Brazil, India, China, the Middle East, and Southeast Asia and is expected to increase at least twofold in the next 20 years [1]. Most people with diabetes have type 2 diabetes mellitus (T2DM) and are initially treated with lifestyle changes and oral antidiabetic drugs [2]. However, because of the progressive nature of the disorder, many will need insulin to achieve glycemic control and thus, the use of insulin-based regimens is expected to continue to increase.

Some of the barriers to insulin use stem from the perceived discomfort and fear of insulin injections; these have been addressed by the introduction of insulin pens as a delivery mechanism [3]. Insulin pen design has focused on a combination of patient convenience/satisfaction and affordability. This is especially important in emerging countries where the need for affordable, convenient insulin delivery is readily apparent. As patient awareness of insulin pens increases in emerging countries, it is important to identify those insulin pen characteristics that are most vital to optimizing patient acceptance and adherence. The aim of this market research survey is provide insight from people with diabetes in emerging countries into the usage characteristics of three to four different reusable insulin pens available in their respective countries, and to observe their pen usage-based preferences.

Methods

Patients

Individuals with type 1 diabetes mellitus (T1DM) or T2DM were recruited for a market research survey in Brazil, China, Egypt, India, and Malaysia by various methods, including street intercept, snowball sampling (i.e., identifying one or more participants in the desired population and using them to find further participants), and recommendation by a healthcare professional. A mixture of gender, occupations, and insulin-naïve or insulin-experienced individuals were recruited. If patients were insulin naïve, they must have discussed starting insulin with their physician and must be willing to do so in the future, otherwise they were excluded from the study. People were also excluded if they had participated in diabetes market research in the past 3 months.

Patient Interviews

Face-to-face interviews were conducted in the local language in either the participant’s home or a central location. Interviews were conducted from June to August 2013 by experienced moderators fully briefed on the research survey’s objectives and materials (pens, cartridges, needles, instruction manuals), and the interview was recorded. After the interviewer confirmed information about the participant, each individual was given a shuffled deck of 12 cards, each showing a feature that could relate to a device for injecting a medicine for diabetes. The 12 key features were confirmed in a pilot phase consisting of 2 days of centrally viewed interviews in each market conducted earlier in May 2013. The participants were asked to put the 12 cards in order of importance to them. Each rank was assigned its corresponding numerical value (i.e., 1.0 for most important, 2.0 for second most important, and so on). For the entire group, the rankings were then averaged to determine the most important characteristics, with importance ranked from lowest to highest value.

The moderator then demonstrated the use of one of the devices, and the participants were asked to prepare the device and inject a 10-unit dose into a pad. This was repeated for each available device, and a manual in the local language was provided for each pen for reference. For each pen, participants were evaluated on their ability to (1) remove the cap, unscrew the cartridge holder and insert the cartridge; (2) screw/lock the cartridge holder; (3) attach the needle; (4) prime/safety test the device; (5) dial the dose; and (6) deliver the dose. The AllStar® (AS; Sanofi, Mumbai, India), NovoPen 4® (NP4; Novo Nordisk, Bagsværd, Denmark), and HumaPen Ergo II® (HE2; Eli Lilly, Indianapolis, USA) were tested in all markets; in addition, one local pen was tested in India (INSUPen®; Biocon, Bangalore, India), Brazil (HumaPen® Luxura™; Eli Lilly, Sao Paulo, Brazil), and China (Xuilin Pen®; Gan & Lee, Beijing, China). The order of presentation of the devices was randomized via three-way or four-way randomization (based on number of pens tested) so that a particular pen was not always first and two pens did not always follow one another. In all countries, participants were not tested on a device that they had used or were currently using, except in Malaysia. NP4 was tested by a sub-sample of current users in Malaysia (n = 52), where the pen is provided free of charge by the government.

The participants were next asked to compare each device across 12 key features by answering a series of questions (Table 1). They were asked to choose which pen was the best and worst for each of those features. Finally, they were asked which device was easiest to use overall, and which device was their last choice. For each question, participants in Brazil, China, and India were also asked to name their second choice, after naming their first choice, before naming their last choice. The results were tabulated and data were tested at the 95% significance level, using z tests and t tests (as appropriate) to establish significance between means and proportions, respectively. Data were analyzed using QPSMR CL (version 2011.2, NetMR Ltd., Cupar, Fife, Scotland, UK) and DigitabXL (version 12.0, Digitab, London, UK).

Table 1 Questionnaire: comparison of devices

As a market research survey, analyses herein do not fall under the auspices of the ethical standards on human experimentation (institutional and national) and the Helsinki Declaration of 1975 (revised 2000, 2008). Consent was gained from all patients for their participation in the demonstration of injection devices (following instructions from the moderators and presentation of instruction manuals) for the data to be reported at an aggregate level; patients were also informed about adverse events reporting procedures.

Results

Patients

A total of 503 participants from India, Malaysia, Brazil, Egypt, and China were interviewed (Table 2). The participants were equally divided between those who were either insulin naïve (49%) or insulin users (51%). More of the insulin users had T2DM (58%) than had T1DM (42%) due to only a few of the insulin users in Malaysia having T1DM (13%). In the other four countries, insulin users were equally divided between those with T1DM and those with T2DM.

Table 2 Distribution of participants by country and type of diabetes

Ease of Use of Pens

Participants completed the preparation and dose injection process with AS, NP4, or HE2 in an average of 2–3 min (Fig. 1), with no significant differences between insulin-naïve participants and insulin users. There were differences between the pens for individual steps. For removing the cap, unscrewing the cartridge holder, and inserting the cartridge (Step 1), 20% of participants struggled to remove the cartridge holder from NP4 due to its stiffness or to trying to pull the cartridge holder out rather than unscrew it; with HE2, 5% of participants tried to twist the cap off and 2% had difficulty inserting the cartridge; with AS, 2% had difficulty inserting the cartridge into the pen/holder (e.g., inserting cartridge upside down). Three percent of participants using AS and 6% using NP4 struggled to position the cartridge holder correctly (Step 2), while 5% with HE2 had some difficulty with this step because of the stiff cartridge holder. Help to attach the needle (Step 3) was needed for 2% of participants with AS, 4% with NP4, and 6% with HE2. In addition, 3% of insulin-naïve participants struggled with the needle safety cap with AS and 5% struggled with NP4, both due to unfamiliarity with the cap.

Fig. 1
figure 1

Ease of use of AllStar, NovoPen 4, and HumaPen Ergo II pens—by steps. AS AllStar, HE2 HumaPen Ergo II, NP4 NovoPen 4

There were 7–12% of participants who failed to prime the devices (Step 4); participants skipping the priming/safety test ranged from 3% (China) to 11% (Egypt) for AS, 4% (Brazil) to 10% (Egypt) for NP4, and 2% (China) to 16% (Egypt) for HE2. A total of 424 participants (88%) primed AS, while 397 (93%) primed NP4, and 407 (90%) primed HE2. Of those who did prime the devices, 28%, 40%, and 35% had difficulty priming with AS, NP4, and HE2, respectively. More than 90% of all problems were due to misunderstanding of the methodology required to perform a step, and were not due to mechanical issues with the pen. The correct dose (Step 5) was dialed by 90–93% of participants; 4% of participants dialed an incorrect dose with each pen, while the remainder either was unable to dial a dose (either due to dose button jam or forgetting to twist/pull out the dose button) or asked which dose to dial. Most participants delivered the entire dose (Step 6) with AS (97%), NP4 (93%) or HE2 (95%), with significantly fewer participants failing to hold down the button for the required time with AS (6%), compared with NP4 (22%) and HE2 (18%). Most participants knew an entire dose was delivered because the dose window read zero (all pens) and/or the dose button could not be depressed further (AS, NP4, and HE2) or the clicking sound stopped (NP4 and HE2).

Most Important Features of an Insulin Pen

Participants ranked ease of injecting oneself, ease of dialing the correct dose, how easy the device was to use overall, and ease of reading the dose display as the most important ease-of-use features of any new insulin pen (Table 3). Being discreet and easy to carry was ranked as the least important of the 12 ease-of-use features. There were no significant differences between insulin-naïve and current insulin users in the top four characteristics.

Table 3 Most to least important features of a new insulin pen: by participant preference

Participants ranked AS highest on nine of 12 key features (Fig. 2), including three of the four most important features: easiest to use overall (52%), easiest to read the dose (42%), and easiest to self-inject (39%). HE2 was ranked highest for easiest to dial back the dose (47%) and to feel/hear dial clicks (39%). It was also ranked highest for easiest to dial the right dose (36%) but not significantly higher than AS (35%). Local pens available in Brazil, China, and India were ranked the lowest for all 12 features. When asked which pen they would choose, if they were thinking specifically about how the pen feels to hold when they inject, 54% chose AS compared with 15% for NP4, 23% for HE2, and 8% for the local pen (P < 0.05 for AS vs. all other pens).

Fig. 2
figure 2

Percentage of participants ranking each device in first place (globally)

Discussion

The observations from this market research survey revealed, in a non-clinical setting, how people with diabetes in the emerging countries of Brazil, China, Egypt, India, and Malaysia used the reusable insulin pens available in their country. Three pens—AS, NP4, and HE2—were available in all five countries, while an additional local pen was available in India, Brazil, and China. Most participants found the AS, NP4, and HE2 pens easy to use overall, but with some differences between pens in the individual steps used to inject a dose. The AS pen was ranked highest for nine of 12 ease-of-use features, while HE2 was ranked highest for two features.

In the first part of the interviews, participants performed the sequential steps used to inject a dose of insulin once per pen. For insulin-naïve participants, this task had the possibility of being the easiest for the last pen tried as they learned from the preceding pens. To reduce this bias, a Latin-squares design was used to balance the order in which the pens were introduced. The tasks were followed by a question-and-answer session to determine the pen preferences of each participant. This design has been used in clinical studies involving insulin pens [46] and allows for a comparison of pen features without any influence from the insulin therapy. This may mimic clinical practice for many people with diabetes, where a diabetes educator or nurse would offer the person the opportunity to test and choose from a selection of devices appropriate for the prescribed insulin regimen. Clinical studies have evaluated pen devices over an extended period of actual use with insulin [711], and evaluation of pen features in many of these studies may have been influenced by the therapeutic outcome with the pen.

Most people with diabetes who require insulin self-administer it by subcutaneous injection, using either a vial and syringe or an insulin pen device. Worldwide, pen devices are used by approximately 60% of insulin users, although usage varies from country to country [12]. Among the countries involved in the current study, 95% of people with diabetes in China are pen users, whereas approximately 60% and 80% use vial and syringe in Brazil and India, respectively. Compared with vial and syringe, insulin pens provide better dosing accuracy, easier dosing and administration, convenience, increased patient acceptance and satisfaction; they are also discreet and easily portable, which lessens social embarrassment [1317]. Pens are considered less painful than syringes and are associated with less needle fear [17]. Their use is also linked with improved adherence to insulin therapy and reduced costs compared with vial and syringe [18], and may reduce the resistance to starting insulin therapy.

The results provided herein give important clinical insights into the preferences and usage capabilities of individuals in regions of the world where reusable insulin pen availability is extensive and/or increasing. From a clinical perspective, the ability of an individual to feel confident and comfortable with the use of such a device is an essential factor in enhancing patient adherence to diabetes management. As insulin injections are a daily requirement, recognizing the attributes most desirable to the patient and providing a device that best fits those preferences help to ensure seamless integration of insulin delivery to a patient’s everyday routine. Additionally, remaining cognizant of those steps that potentially pose difficulty with each pen is also important, so that the diabetes management team (whether physician, nurse, or caregiver) can focus their instruction on those steps which may be most problematic. This holds true for both newly initiated insulin users (where pen characteristics may factor in treatment choice) as well as experienced insulin pen users (for their current insulin treatment, or if a change in treatment is considered).

Potential limitations to this study include first-impression preferences by participants without long-term, learned pen use, as well as the inability to gauge comfort and execution of actual self-injection, particularly for those who have never utilized insulin. As such, areas of further research and discussion if data were available include: (1) influence of age: younger versus older participants; (2) insulin-naïve participants versus insulin users; and (3) T1DM vs. T2DM participants.

Conclusions

As a result of interviews with individuals with T1DM or T2DM, it was identified that new and existing users of insulin pens seek ease of injection, overall ease of use, and correct dose delivery as key characteristics for an insulin pen device. Through hands-on use of these different pens, priming the reusable insulin pens was the most difficult aspect of administering a dose; however, each pen showed slight variation in the steps that posed difficulty with administration. The AS pen was easiest to use overall compared with other reusable pens tested, and ranked highest by uses in most of the characteristics identified as most preferred for a reusable insulin pen. Selection of an appropriate reusable insulin pen may provide benefit and comfort for patients starting or continuing insulin therapy; identifying those characteristics that are most preferred by patients may assist in overcoming barriers to appropriate dose delivery and overall adherence with treatment.